Base neutrals - Base neutrals to EPA 8270

Applications | 2011 | Agilent TechnologiesInstrumentation
GC/MSD, GC columns, Consumables
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


Base neutrals are a key class of semi-volatile organic compounds regulated under EPA Method 8270. Their accurate separation and quantitation are critical for environmental monitoring of contaminated water, soil, and air matrices. Efficient analysis of these analytes supports compliance with regulatory limits and informs remediation efforts.

Objectives and Study Overview


This application note demonstrates a rapid GC/MS procedure using an Agilent FactorFour VF-200ms column to resolve 13 base neutrals specified by EPA Method 8270. The goal is to achieve complete separation within a 20-minute run while maintaining sensitivity and reproducibility.

Methodology


A split injection of 1 µL sample in methylene chloride (2 mg/mL) was introduced into a GC/MS system. The temperature program commenced at 45 °C (3 min hold), ramping at 10 °C/min to 325 °C. Helium served as carrier gas (1.0 mL/min) under a constant pressure of 60 kPa. Mass spectral detection enabled selective monitoring of target ions.

Used Instrumentation


  • Agilent 7890A Gas Chromatograph
  • Agilent 5975C Mass Selective Detector
  • Agilent FactorFour VF-200ms column (0.25 mm × 30 m, 1.0 µm film thickness)
  • Split/Splitless inlet operated in split mode (1:100)

Key Results and Discussion


  • Efficient baseline separation of all 13 base neutrals within 20 min.
  • Sharp peak shapes and minimal coelution achieved through optimized temperature programming and column selectivity.
  • Compounds resolved include: methyl methane sulfonate, ethyl methane sulfonate, acetophenone, nitrobenzene, isophorone, safrole, isosafrole, 1,4-naphthoquinone, m-dinitrobenzene, 2,6-dinitrotoluene, 2,4-dinitrotoluene, 1,3,5-trinitrobenzene, and pentachloronitrobenzene.

Benefits and Practical Applications


  • High throughput analysis for routine environmental testing laboratories.
  • Robust method suited for regulatory compliance and quality assurance/quality control programs.
  • Reduced analysis time lowers operational costs and increases sample throughput.

Future Trends and Potential Applications


Ongoing developments include coupling to high-resolution MS for enhanced specificity, miniaturized GC systems for field deployment, and automation of sample preparation to further accelerate workflows. Integration with data analytics platforms may provide deeper insights into contaminant source tracking and risk assessment.

Conclusion


The presented GC/MS method delivers rapid, reliable separation of EPA 8270 base neutrals using an Agilent FactorFour VF-200ms column. It meets regulatory requirements while improving laboratory efficiency.

References


  • Agilent Technologies, Inc. Base neutrals to EPA 8270. Application Note A02090, 31 October 2011.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Base neutrals
Base neutrals
2011|Agilent Technologies|Applications
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. Using GC/MS, an Agilent FactorFour VF-200ms column separates 14 base neutrals in 20 minutes. Conditions Technique : GC Column : Agilent FactorFour VF-200ms, 0.25mm x 30 m (df = 1.0…
Key words
neutrals, neutralscourtesy, courtesybase, baseprinted, printedinjector, injectorusa, usaauthors, authorsprogram, programenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationpressure, pressuresolvent
Base neutrals
Base neutrals
2011|Agilent Technologies|Applications
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC analysis of 14 base neutrals with an Agilent VF-200ms column achieves a separation in 28 minutes. Conditions Technique : GC Column : Agilent FactorFour VF-200ms, 0.25 mm x 30…
Key words
neutrals, neutralscourtesy, courtesybase, baseprinted, printedinjector, injectorusa, usaauthors, authorsprogram, programenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationpressure, pressuresolvent
Base neutrals
Base neutrals
2011|Agilent Technologies|Applications
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. Analysis of 14 halogenated base neutrals using Agilent FactorFour VF-1301ms in 22 minutes. Conditions Technique : GC-capillary Column : Agilent FactorFour VF-1301ms, 0.25 mm x 30 m fused silica (df…
Key words
neutrals, neutralscourtesy, courtesybase, baseprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentaltechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak, peaktemperature
Nitro- and chlorinated aromatics - Separation of nitro-and halogenated aromatic compounds (base neutrals)
Nitro- and chlorinated aromatics Separation of nitro-and halogenated aromatic compounds (base neutrals) Application Note Environmental Authors Introduction Agilent Technologies, Inc. Separation of 14 nitro-and halogenated aromatic compounds (base neutrals) by Agilent VF-1701ms in 20 minutes. Conditions Technique : GC-capillary Column…
Key words
nitro, nitroneutrals, neutralshalogenated, halogenatedaromatics, aromaticschlorinated, chlorinatedcourtesy, courtesyaromatic, aromaticbase, baseprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike