Nitro- and chlorinated aromatics - Separation of nitro-and halogenated aromatic compounds (base neutrals)
Applications | 2011 | Agilent TechnologiesInstrumentation
Halogenated and nitroaromatic compounds are common environmental pollutants arising from industrial processes, pesticides and combustion. Their toxicity, persistence and potential to bioaccumulate make rapid and reliable analysis essential for monitoring air, water and soil quality. Efficient separation of these base-neutral analytes supports regulatory compliance and protects human health and ecosystems.
This application note demonstrates a fast gas chromatographic method for the separation of 14 nitro- and halogenated aromatic compounds within 20 minutes. The primary goals were to evaluate the performance of the Agilent VF-1701ms capillary column, optimize chromatographic conditions and verify baseline resolution of structurally similar analytes.
Chromatographic technique used:
The optimized method achieved complete separation of 14 base-neutral nitro- and chlorinated aromatics within 20 minutes. Peak identification order:
Chromatograms showed sharp, well-resolved peaks with minimal tailing, demonstrating the column’s selectivity for substituted aromatics and nitro compounds. The temperature program balanced early elution of lighter dichlorobenzenes with adequate retention for heavier, more chlorinated analytes.
This rapid separation method offers:
It can be applied to monitoring of industrial effluents, soil extracts and air emissions where mixtures of nitro- and halogenated aromatics occur.
Emerging directions include coupling fast GC methods with mass spectrometry for confirmatory analysis and lower detection limits. Development of novel stationary phases tailored for ultra-fast separations may further reduce analysis time. Automation and integration with sample preparation modules will enhance laboratory efficiency.
The Agilent VF-1701ms column delivers efficient, baseline separation of 14 nitro- and chlorinated aromatic compounds in under 20 minutes, making it a valuable tool for environmental monitoring and industrial compliance. The method’s speed, resolution and adaptability support diverse analytical needs in the field of trace organic pollutant analysis.
GC, GC columns, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Significance of the Topic
Halogenated and nitroaromatic compounds are common environmental pollutants arising from industrial processes, pesticides and combustion. Their toxicity, persistence and potential to bioaccumulate make rapid and reliable analysis essential for monitoring air, water and soil quality. Efficient separation of these base-neutral analytes supports regulatory compliance and protects human health and ecosystems.
Objectives and Study Overview
This application note demonstrates a fast gas chromatographic method for the separation of 14 nitro- and halogenated aromatic compounds within 20 minutes. The primary goals were to evaluate the performance of the Agilent VF-1701ms capillary column, optimize chromatographic conditions and verify baseline resolution of structurally similar analytes.
Methodology and Instrumentation
Chromatographic technique used:
- Gas chromatography with capillary column
- Column: Agilent Factor Four VF-1701ms, 0.25 mm × 30 m, 0.25 µm film thickness
- Oven temperature program: 45 °C hold 3 min, ramp at 10 °C/min to 280 °C
- Carrier gas: Helium at 60 kPa (approximately 1 mL/min)
- Injector: Split mode 1:100, injection volume 1 µL
- Detector: Flame ionization detector (FID)
- Sample concentration: 100 µg/mL in methylene chloride
Key Results and Discussion
The optimized method achieved complete separation of 14 base-neutral nitro- and chlorinated aromatics within 20 minutes. Peak identification order:
- 1,3-Dichlorobenzene
- 1,4-Dichlorobenzene
- 1,2-Dichlorobenzene
- Hexachloroethane
- Nitrobenzene
- Isophorone
- 1,2,4-Trichlorobenzene
- Hexachlorobutadiene
- Hexachlorocyclopentadiene
- 2-Chloronaphthalene
- 2,6-Dinitrotoluene
- 2,4-Dinitrotoluene
- Azobenzene
- Hexachlorobenzene
Chromatograms showed sharp, well-resolved peaks with minimal tailing, demonstrating the column’s selectivity for substituted aromatics and nitro compounds. The temperature program balanced early elution of lighter dichlorobenzenes with adequate retention for heavier, more chlorinated analytes.
Benefits and Practical Applications
This rapid separation method offers:
- High throughput analysis for environmental laboratories
- Robust reproducibility for routine QA/QC workflows
- Compatibility with FID and potential for MS detection to enhance specificity
It can be applied to monitoring of industrial effluents, soil extracts and air emissions where mixtures of nitro- and halogenated aromatics occur.
Future Trends and Potential Applications
Emerging directions include coupling fast GC methods with mass spectrometry for confirmatory analysis and lower detection limits. Development of novel stationary phases tailored for ultra-fast separations may further reduce analysis time. Automation and integration with sample preparation modules will enhance laboratory efficiency.
Conclusion
The Agilent VF-1701ms column delivers efficient, baseline separation of 14 nitro- and chlorinated aromatic compounds in under 20 minutes, making it a valuable tool for environmental monitoring and industrial compliance. The method’s speed, resolution and adaptability support diverse analytical needs in the field of trace organic pollutant analysis.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Nitro and chlorinated aromatics
2011|Agilent Technologies|Applications
Nitro and chlorinated aromatics Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC Analysis with Direct injection of nitro- and halogenated aromatic compounds (base neutral) using stabilized 50% phenyl PDMS Conditions Technique : GC Column : Agilent VF-17ms, 0.53 mm…
Key words
nitro, nitroaromatics, aromaticshalogenated, halogenatedchlorinated, chlorinatedcourtesy, courtesyaromatic, aromaticprinted, printedinjector, injectorusa, usadirect, directauthors, authorsenvironmental, environmentalsize, sizecarrier, carriertechnique
Nitro and chlorinated aromatics
2011|Agilent Technologies|Applications
Nitro and chlorinated aromatics Application Note Environmental Authors Introduction Agilent Technologies, Inc. For GC separation of 14 nitrosamines in under 20 minutes use the stabilized 50% phenyl PDMS phase of an Agilent VF-17ms column. Conditions Technique : GC Column :…
Key words
nitro, nitroaromatics, aromaticschlorinated, chlorinatedcourtesy, courtesyprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote
Nitro compounds
2011|Agilent Technologies|Applications
Nitro compounds Application Note Environmental Authors Introduction Agilent Technologies, Inc. Agilent VF-1701ms columns separate eight nitro compounds in 22 minutes. Conditions Technique : GC-capillary Column : Agilent FactorFourVF-1701ms, 0.25 mm x 30 m fused silica (df= 0.25 μm) (Part no.…
Key words
nitro, nitrodibenzofuran, dibenzofurananiline, anilinebenzyl, benzylcourtesy, courtesyalcohol, alcoholprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification
Nitro- and sulfur compounds
2011|Agilent Technologies|Applications
Nitro- and sulfur compounds Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC separation of 13 nitro and sulfur compounds according to EPA 8270 is achieved in 20 minutes with an Agilent VF-17ms column. Conditions Technique : GC Column :…
Key words
nitro, nitrocourtesy, courtesysulfur, sulfurprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizecarrier, carriertechnique, techniqueidentification, identificationdetector, detectorcompounds, compoundsnote