Blood Alcohol Determination with Teledyne Tekmar HT3™ Automated Static/Dynamic Headspace Analyzer
Applications | 2011 | Teledyne LABSInstrumentation
The accurate measurement of blood alcohol concentration is critical in forensic toxicology, clinical diagnostics, and regulatory enforcement of impaired driving laws. Headspace gas chromatography provides a reliable, minimally invasive approach to quantify volatile analytes in biological fluids.
This study aimed to evaluate Teledyne Tekmar’s HT3™ Automated Static/Dynamic Headspace Analyzer for routine determination of ethanol and related volatiles in blood. Key goals included establishing optimal instrument parameters, assessing method linearity, precision, and demonstrating dual-column confirmation.
A static headspace procedure was employed with 70 °C sample equilibration for 10 minutes, followed by automated transfer to an Agilent 6890 GC fitted with dual flame ionization detectors. An internal standard of n-propanol in 0.5 M ammonium sulfate buffer ensured accurate quantitation.
Chromatograms exhibited sharp, well-resolved peaks for ethanol, methanol, acetaldehyde, acetone, and isopropanol. Seven-point calibration over 0.002–0.32 g/dL ethanol (and equivalent levels of other volatiles) produced correlation coefficients ≥0.999, with ethanol achieving r² = 1.0000. Precision trials at 0.008 g/dL ethanol yielded internal-standard RSD values below 1% on the quantitation column and below 2.5% on the confirmation column.
The described method delivers rapid (<3 min) analysis, high linearity, and robust intra-day precision, supporting high-throughput workflows in forensic and clinical laboratories. Dual-column confirmation enhances result credibility and defensibility in legal contexts.
Advances may include coupling headspace analyzers with mass spectrometric detection for increased specificity, miniaturized and portable systems for field use, integration with laboratory information management systems for automated data handling, and expansion to additional volatile biomarkers in clinical toxicology.
Teledyne Tekmar’s HT3™ Automated Static/Dynamic Headspace Analyzer, combined with dual Zebron ZB-BAC columns and dual FIDs, provides a fast, precise, and linear method for blood alcohol determination. The approach meets forensic and regulatory requirements for accuracy, repeatability, and confirmatory analysis.
GC, HeadSpace
IndustriesForensics
ManufacturerAgilent Technologies, Teledyne LABS
Summary
Importance of the Topic
The accurate measurement of blood alcohol concentration is critical in forensic toxicology, clinical diagnostics, and regulatory enforcement of impaired driving laws. Headspace gas chromatography provides a reliable, minimally invasive approach to quantify volatile analytes in biological fluids.
Objectives and Study Overview
This study aimed to evaluate Teledyne Tekmar’s HT3™ Automated Static/Dynamic Headspace Analyzer for routine determination of ethanol and related volatiles in blood. Key goals included establishing optimal instrument parameters, assessing method linearity, precision, and demonstrating dual-column confirmation.
Methodology and Instrumentation
A static headspace procedure was employed with 70 °C sample equilibration for 10 minutes, followed by automated transfer to an Agilent 6890 GC fitted with dual flame ionization detectors. An internal standard of n-propanol in 0.5 M ammonium sulfate buffer ensured accurate quantitation.
- Headspace system: Teledyne Tekmar HT3™ Automated Static/Dynamic Headspace Analyzer
- Gas chromatograph: Agilent 6890 GC with dual FIDs, constant oven at 40 °C
- Quantitation column: Phenomenex Zebron ZB-BAC-1 (30 m × 0.53 mm, 3 µm film)
- Confirmation column: Phenomenex Zebron ZB-BAC-2 (30 m × 0.53 mm, 2 µm film)
- Carrier gas flows: 12 mL/min column flow, FID gases hydrogen and air at 35 mL/min and 300 mL/min respectively
Main Results and Discussion
Chromatograms exhibited sharp, well-resolved peaks for ethanol, methanol, acetaldehyde, acetone, and isopropanol. Seven-point calibration over 0.002–0.32 g/dL ethanol (and equivalent levels of other volatiles) produced correlation coefficients ≥0.999, with ethanol achieving r² = 1.0000. Precision trials at 0.008 g/dL ethanol yielded internal-standard RSD values below 1% on the quantitation column and below 2.5% on the confirmation column.
Benefits and Practical Applications
The described method delivers rapid (<3 min) analysis, high linearity, and robust intra-day precision, supporting high-throughput workflows in forensic and clinical laboratories. Dual-column confirmation enhances result credibility and defensibility in legal contexts.
Future Trends and Possibilities
Advances may include coupling headspace analyzers with mass spectrometric detection for increased specificity, miniaturized and portable systems for field use, integration with laboratory information management systems for automated data handling, and expansion to additional volatile biomarkers in clinical toxicology.
Conclusion
Teledyne Tekmar’s HT3™ Automated Static/Dynamic Headspace Analyzer, combined with dual Zebron ZB-BAC columns and dual FIDs, provides a fast, precise, and linear method for blood alcohol determination. The approach meets forensic and regulatory requirements for accuracy, repeatability, and confirmatory analysis.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Blood Alcohol Determination with the New Teledyne Tekmar Versa Headspace Instrument
2011|Shimadzu|Applications
Blood Alcohol Determination with the New Teledyne Tekmar Versa Headspace Instrument Application Note Introduction Recently, Teledyne Tekmar released the Versa, their latest innovative headspace instrument. Initially developed for the pharmaceutical industry to provide an economical headspace system for residual solvent…
Key words
versa, versaarea, areaequil, equilblood, bloodalcohol, alcoholpressurize, pressurizeheadspace, headspaceacetaldehyde, acetaldehydetemp, tempisopropanol, isopropanolplaten, platenacetone, acetoneethanol, ethanolpropanol, propanoleconomical
Blood Alcohol Determination with the Teledyne Tekmar HT3 and Young Lin YL6500 Gas Chromatograph
2012|Teledyne LABS|Applications
Blood Alcohol Determination with the Teledyne Tekmar HT3 and Young Lin YL6500 Gas Chromatograph Application Note Introduction Police and forensic laboratories around the world routinely use headspace-gas chromatography for volatile components in biological fluids. One of the more common headspace…
Key words
tbut, tbutnpro, nproext, extblood, bloodpressurize, pressurizealcohol, alcoholtemp, temppropanol, propanolacetaldehyde, acetaldehydebutanol, butanoltime, timemixer, mixerisopropanol, isopropanolacetone, acetonefill
Alcoholic Beverage Fusel Alcohol Testing with Static Headspace
2014|Teledyne LABS|Applications
Alcoholic Beverage Fusel Alcohol Testing with Static Headspace Application Note Abstract A static headspace method was developed using Teledyne Tekmar automated headspace vial samplers (Figure 1 and Figure 2) to meet the method requirements of the Alcohol and Tobacco Tax…
Key words
fusel, fuselheadspace, headspaceequil, equilalcohol, alcoholpressurize, pressurizestatic, statictemp, temptime, timeplaten, platentable, tableisobutanol, isobutanolisoamyl, isoamylteledyne, teledynepropanol, propanolamyl
Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler
2004|Agilent Technologies|Applications
Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application Forensics Author Introduction Roger L. Firor and Chin-Kai Meng Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Blood alcohol analysis is a widely used, highthroughput application…
Key words
acetone, acetoneethanol, ethanolpropanol, propanolheadspace, headspaceisopropanol, isopropanolloop, loopblood, bloodmek, meksplit, splitetoh, etohcarrier, carrieralcohol, alcoholmethanol, methanoltemp, tempsplitless