Blood Alcohol Determination with the Teledyne Tekmar HT3 and Young Lin YL6500 Gas Chromatograph

Applications | 2012 | Teledyne LABSInstrumentation
GC, HeadSpace
Industries
Forensics
Manufacturer
Teledyne LABS

Summary

Significance of the Topic


Headspace gas chromatography for blood alcohol testing is a cornerstone analytical technique in forensic and clinical laboratories. Reliable determination of ethanol concentration and related volatile compounds is critical for legal compliance, patient care, and forensic investigations.

Study Objectives and Overview


This study aimed to assess the performance of the Teledyne Tekmar HT3 automated headspace sampler coupled with the Young Lin YL6500 gas chromatograph and flame ionization detector. Target analytes included ethanol, its metabolic byproducts (acetaldehyde, acetone), and other potential contaminants or endogenous compounds (methanol, isopropanol).

Methodology


Samples and calibration standards were prepared by spiking aqueous matrices with analytes at concentrations ranging from 0.002% to 0.10% (w/v). An internal standard approach using n-propanol or tert-butanol was employed. Each vial contained sample, salt for salting-out, and internal standard solution, followed by equilibration and automated headspace sampling.

Instrumentation


The HT3 sampler provided precise vial heating, agitation, and headspace transfer. Key parameters included an 80°C sample temperature, 15 min equilibration, and controlled vial pressurization. The YL6500 GC used a polar wax column (30 m×0.32 mm, 0.25 µm film) with helium carrier at 2 mL/min. The oven ramp started at 35°C held for 6 min, then rose at 25°C/min to 150°C. The FID operated at 230°C with optimized hydrogen, air, and makeup gas flows for enhanced sensitivity.

Main Results and Discussion


Calibration curves demonstrated excellent linearity (r² ≥ 0.998) for all analytes using both external and internal standard methods. Precision studies at 0.002%, 0.05%, and 0.10% levels yielded percent relative standard deviations below 5% for most compounds, confirming robust reproducibility. Chromatograms showed clear baseline separation of acetaldehyde, acetone, methanol, isopropanol, ethanol, and internal standards.

Benefits and Practical Applications


Automated headspace sampling combined with GC/FID delivers high throughput, consistent sample handling, and reliable quantification tailored for forensic casework, clinical toxicology, and regulatory compliance. The method supports low detection limits and can be adjusted to meet various legal blood alcohol thresholds.

Future Trends and Applications


Emerging trends include integration with laboratory information management systems, AI-assisted real-time data interpretation, portable GC platforms for on-site testing, and expanded multi-analyte panels covering additional biomarkers and volatile substances.

Conclusion


The combination of the Teledyne Tekmar HT3 headspace sampler and Young Lin YL6500 GC/FID offers a validated, reliable approach for blood alcohol determination. High linearity, precision, and automation efficiency position this workflow as an effective solution for modern forensic and clinical laboratories.

References


Teledyne Tekmar Application Note: Blood Alcohol Determination with the HT3 and YL6500 GC, December 2012.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Blood Alcohol Determination with Teledyne Tekmar HT3™ Automated Static/Dynamic Headspace Analyzer
Blood Alcohol Determination with Teledyne Tekmar HT3™ Automated Static/Dynamic Headspace Analyzer Application Note By: Roger Bardsley Introduction Analysis of biological fluids for volatile components by headspace analysis is routinely used by forensic laboratories around the world. One of the more…
Key words
area, arearatio, ratioacetaldehyde, acetaldehydeheadspace, headspaceblood, bloodethanol, ethanolalcohol, alcoholisopropanol, isopropanolacetone, acetonestatic, staticpressurize, pressurizemethanol, methanoltemp, temppropanol, propanolloop
Blood Alcohol Determination with the New Teledyne Tekmar Versa Headspace Instrument
Blood Alcohol Determination with the New Teledyne Tekmar Versa Headspace Instrument Application Note Introduction Recently, Teledyne Tekmar released the Versa, their latest innovative headspace instrument. Initially developed for the pharmaceutical industry to provide an economical headspace system for residual solvent…
Key words
versa, versaarea, areaequil, equilblood, bloodalcohol, alcoholpressurize, pressurizeheadspace, headspaceacetaldehyde, acetaldehydetemp, tempisopropanol, isopropanolplaten, platenacetone, acetoneethanol, ethanolpropanol, propanoleconomical
Alcoholic Beverage Fusel Alcohol Testing with Static Headspace
Alcoholic Beverage Fusel Alcohol Testing with Static Headspace Application Note Abstract A static headspace method was developed using Teledyne Tekmar automated headspace vial samplers (Figure 1 and Figure 2) to meet the method requirements of the Alcohol and Tobacco Tax…
Key words
fusel, fuselheadspace, headspaceequil, equilalcohol, alcoholpressurize, pressurizestatic, statictemp, temptime, timeplaten, platentable, tableisobutanol, isobutanolisoamyl, isoamylteledyne, teledynepropanol, propanolamyl
Rtx-BAC Plus 1 and Rtx-BAC Plus 2 Columns
Rtx-BAC Plus 1 and Rtx-BAC Plus 2 Columns
2018|Agilent Technologies|Brochures and specifications
Restek GC LC Rtx-BAC Plus 1 and Rtx-BAC Plus 2 Columns Advanced Technology for Fast, Reliable Measurement of Alcohol in Blood • Optimized column selectivities guarantee resolution of ethanol, internal standards, and frequently encountered interferences. • Robust and reproducible column…
Key words
restek, restekleak, leakpeaks, peaksadaptor, adaptoruniversal, universalalcohol, alcoholglycol, glycolessentials, essentialscharger, chargertailing, tailingacetone, acetoneblood, blooddescription, descriptionthread, threadbutanol
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike