Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler
Applications | 2004 | Agilent TechnologiesInstrumentation
Blood alcohol analysis is a critical forensic tool for DUI enforcement and legal proceedings. Static headspace gas chromatography reduces sample preparation, lowers maintenance, and enhances throughput when determining volatile compounds in complex biological matrices such as blood.
This study evaluates the performance of the Agilent G1888 Network Headspace Sampler coupled to a 6890N GC with dual-column configurations (0.32 mm and 0.53 mm ID DB-ALC1/DB-ALC2 phases) under isothermal conditions. The goal is to achieve rapid, accurate ethanol quantification with robust confirmation of analytes and minimal carryover.
The G1888 sampler offers a fully inert sample path, uniform heating zones, and programmable vent purge. A 70-vial tray accommodates high throughput. Key settings include headspace oven at 60–70 °C, loop at 70–80 °C, transfer line at 80–90 °C, 10–15 min equilibration with vigorous shake, and GC cycle times of 4–6 min. Split/splitless injection and flame ionization detection at 300 °C record volatile compounds. Internal standards used include n-propanol, methyl-ethyl ketone, or t-butanol (common in EU), depending on regional protocols.
Further advances may include tailored GC temperature programs to extend the scope to inhalant and environmental pollutants, enhanced automation and online LIMS integration, development of novel inert coatings to further reduce carryover, and broader adoption in clinical and industrial QA/QC laboratories.
The G1888 headspace sampler paired with an Agilent 6890N GC delivers reliable, high-throughput blood alcohol analysis with excellent repeatability, linearity, low carryover, and dual-phase confirmation. This setup meets forensic requirements and regulatory standards, offering a robust solution for modern forensic laboratories.
GC, HeadSpace
IndustriesForensics
ManufacturerAgilent Technologies
Summary
Importance of the Topic
Blood alcohol analysis is a critical forensic tool for DUI enforcement and legal proceedings. Static headspace gas chromatography reduces sample preparation, lowers maintenance, and enhances throughput when determining volatile compounds in complex biological matrices such as blood.
Objectives and Study Overview
This study evaluates the performance of the Agilent G1888 Network Headspace Sampler coupled to a 6890N GC with dual-column configurations (0.32 mm and 0.53 mm ID DB-ALC1/DB-ALC2 phases) under isothermal conditions. The goal is to achieve rapid, accurate ethanol quantification with robust confirmation of analytes and minimal carryover.
Methodology and Instrumentation
The G1888 sampler offers a fully inert sample path, uniform heating zones, and programmable vent purge. A 70-vial tray accommodates high throughput. Key settings include headspace oven at 60–70 °C, loop at 70–80 °C, transfer line at 80–90 °C, 10–15 min equilibration with vigorous shake, and GC cycle times of 4–6 min. Split/splitless injection and flame ionization detection at 300 °C record volatile compounds. Internal standards used include n-propanol, methyl-ethyl ketone, or t-butanol (common in EU), depending on regional protocols.
Main Results and Discussion
- Resolution: Both column configurations clearly separate ethanol from common interferences (acetaldehyde, methanol, acetone, isopropanol, acetonitrile, ethyl acetate, MEK) in under 5 min.
- Repeatability: Relative standard deviations for ethanol were typically below 1% over 18–40 consecutive runs; other analytes showed RSDs under 3.5% at 0.1–0.15 g/dL.
- Carryover: Ethanol carryover remained below 1% after high-concentration runs (0.5% mix), demonstrating effective vent purging and inert flow path.
- Linearity: FID response was linear from 0.005 to 1.0 g/dL (R² > 0.9993) for all target compounds.
- System Stability: Calculated system response factors (k-factors) varied by less than ±1% over 40 runs across multiple concentration levels.
- European Applications: Using t-butanol as ISTD produced minor co-elutions with solvents on each phase, highlighting the benefit of dual-phase confirmation.
Practical Benefits and Applications
- High throughput with cycle times under 5 min and a 70-sample tray.
- Minimal maintenance due to inert Siltek-coated flow paths and reduced condensation.
- Dual-column confirmation enhances specificity and reduces false positives.
- Integrated 21 CFR Part 11-compliant software within the ChemStation environment enables secure, automated control and data logging.
Future Trends and Opportunities
Further advances may include tailored GC temperature programs to extend the scope to inhalant and environmental pollutants, enhanced automation and online LIMS integration, development of novel inert coatings to further reduce carryover, and broader adoption in clinical and industrial QA/QC laboratories.
Conclusion
The G1888 headspace sampler paired with an Agilent 6890N GC delivers reliable, high-throughput blood alcohol analysis with excellent repeatability, linearity, low carryover, and dual-phase confirmation. This setup meets forensic requirements and regulatory standards, offering a robust solution for modern forensic laboratories.
References
- ICAP Reports 2002 Issue 11, “Blood Alcohol Concentration Limits Worldwide”.
- E. Kuhn, M. Datta, J. Ellis, “Separation and Identification of Blood Pollutants”, Agilent Technologies, Publication B-0328 Rev.2.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Blood Alcohol Analysis Using an Automated Static Headspace Method
2005|Agilent Technologies|Applications
AppNote 1/2005 Blood Alcohol Analysis Using an Automated Static Headspace Method Vanessa R. Kinton, Edward A. Pfannkoch, and Jacqueline A. Whitecavage Gerstel, Inc., 701 Digital Drive, Suite J, Linthicum, MD 21090, USA ABSTRACT Forensic laboratories face the need to analyze…
Key words
ethanol, ethanolpass, passalcohol, alcoholpropanol, propanolaik, aikaiu, aiuaue, auecke, ckecalifornia, californiaheadspace, headspaceake, akeblood, bloodisopropanol, isopropanolstandard, standardacetone
Determine Blood Alcohol with Dual Column/Dual FID for Precision and Reproducibility
2013|Agilent Technologies|Applications
Determine Blood Alcohol with Dual Column/Dual FID for Precision and Reproducibility Application Note Forensics and Toxicology Authors Abstract Haleigh Boswell and Frank Dorman This application note highlights the use of Agilent J&W DB-ALC1 and DB-ALC2 The Penn State University columns…
Key words
ethanol, ethanolpropanol, propanoldual, dualinternal, internalvalue, valueblood, bloodbutanol, butanolcalculated, calculatedfid, fidcoelutions, coelutionsmek, mekknown, knownalcohol, alcoholprecision, precisionacetaldehyde
Determination of Blood Alcohol with Dual Column/Dual FID and the Agilent Intuvo 9000 GC
2017|Agilent Technologies|Applications
Determination of Blood Alcohol with Dual Column/Dual FID and the Agilent Intuvo 9000 GC Application Note Author Abstract Rebecca Veeneman PhD This Application Note demonstrates dual column/dual detector separation and Agilent Technologies, Inc. detection of blood alcohol components on an…
Key words
ethanol, ethanoldual, dualresponse, responseconcentration, concentrationblood, bloodisopropanol, isopropanolrepeatability, repeatabilityacetone, acetonealcohol, alcoholcalibration, calibrationdetermined, determinedcurves, curvesdetector, detectorflow, flowmethanol
Rapid, automated, and accurate determination of blood alcohol concentration (BAC) by headspace coupled to gas chromatography and flame ionization detection
2019|Thermo Fisher Scientific|Applications
APPLICATION NOTE 10713 Rapid, automated, and accurate determination of blood alcohol concentration (BAC) by headspace coupled to gas chromatography and flame ionization detection Authors Jane Cooper, Giulia Riccardino, and Cristian Cojocariu Thermo Fisher Scientific, Runcorn, UK Keywords Blood alcohol concentration,…
Key words
bac, bacblood, bloodethanol, ethanolcarryover, carryoveralcohol, alcoholacf, acfloop, looptracegold, tracegoldacetone, acetonevial, vialrsd, rsdchromatographic, chromatographicincubation, incubationfid, fidisopropanol