Aromatic volatiles
Applications | 2011 | Agilent TechnologiesInstrumentation
The analysis of aromatic volatile organic compounds (VOCs) is critical for environmental monitoring, industrial hygiene, and quality control across chemical and pharmaceutical industries. Rapid and accurate determination of these compounds helps assess air and water quality, detect contamination events, and ensure compliance with regulatory standards. The ability to separate a complex mixture of aromatics in a short time frame enhances laboratory throughput and reduces operational costs.
This application note demonstrates a nine-minute gas chromatographic method for simultaneous separation and identification of 11 common aromatic volatiles. The primary goal is to showcase the performance of the Agilent VF-200ms capillary column coupled with mass spectrometric detection, evaluating resolution, run time, and sensitivity for key environmental analytes.
The method employs gas chromatography with mass spectrometry (GC-MS). Key operating parameters include:
The optimized temperature program and high-efficiency VF-200ms column achieved baseline resolution of all 11 aromatics—including benzene, toluene, chlorobenzene, ethylbenzene, the three xylene isomers, styrene, and dichlorobenzene isomers—within a total runtime of nine minutes. Peak elution order was confirmed by characteristic MS spectra. Rapid heating rate and low dead volume ensured sharp peaks and minimized coelution risks, demonstrating excellent reproducibility and retention time stability.
Ultrafast separation of environmental VOCs reduces sample turnaround time, increases sample throughput, and lowers operational costs in routine monitoring laboratories. The method’s robustness and sensitivity make it suitable for trace-level detection in air and water matrices, supporting compliance testing and contamination investigations.
Advancements in stationary phase chemistry and micro-fabricated column technologies promise even faster separations with enhanced selectivity. Integration of high-resolution MS and automated sample preparation will further improve sensitivity and data quality. Expanding applications into real-time field analysis using portable GC-MS platforms may revolutionize on-site environmental monitoring.
This application note validates a rapid GC-MS protocol for efficient separation of 11 aromatic volatiles using the Agilent VF-200ms column. The method delivers exceptional resolution and speed, meeting the demands of high-throughput environmental and industrial testing.
GC/MSD, GC columns, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Importance of the Topic
The analysis of aromatic volatile organic compounds (VOCs) is critical for environmental monitoring, industrial hygiene, and quality control across chemical and pharmaceutical industries. Rapid and accurate determination of these compounds helps assess air and water quality, detect contamination events, and ensure compliance with regulatory standards. The ability to separate a complex mixture of aromatics in a short time frame enhances laboratory throughput and reduces operational costs.
Objectives and Scope of the Study
This application note demonstrates a nine-minute gas chromatographic method for simultaneous separation and identification of 11 common aromatic volatiles. The primary goal is to showcase the performance of the Agilent VF-200ms capillary column coupled with mass spectrometric detection, evaluating resolution, run time, and sensitivity for key environmental analytes.
Methodology and Used Instrumentation
The method employs gas chromatography with mass spectrometry (GC-MS). Key operating parameters include:
- Column: Agilent FactorFour VF-200ms, 0.25 mm × 30 m, 1.0 μm film thickness
- Carrier Gas: Helium at ~1.0 mL/min, constant pressure of 60 kPa
- Oven Program: 45 °C (3 min hold) ramped at 10 °C/min to 325 °C
- Injection: Split mode (1:100), 1 μL sample volume, solvent methylene chloride (2000 μg/mL)
- Detector: Mass spectrometer (MS)
Results and Discussion
The optimized temperature program and high-efficiency VF-200ms column achieved baseline resolution of all 11 aromatics—including benzene, toluene, chlorobenzene, ethylbenzene, the three xylene isomers, styrene, and dichlorobenzene isomers—within a total runtime of nine minutes. Peak elution order was confirmed by characteristic MS spectra. Rapid heating rate and low dead volume ensured sharp peaks and minimized coelution risks, demonstrating excellent reproducibility and retention time stability.
Benefits and Practical Applications
Ultrafast separation of environmental VOCs reduces sample turnaround time, increases sample throughput, and lowers operational costs in routine monitoring laboratories. The method’s robustness and sensitivity make it suitable for trace-level detection in air and water matrices, supporting compliance testing and contamination investigations.
Future Trends and Opportunities
Advancements in stationary phase chemistry and micro-fabricated column technologies promise even faster separations with enhanced selectivity. Integration of high-resolution MS and automated sample preparation will further improve sensitivity and data quality. Expanding applications into real-time field analysis using portable GC-MS platforms may revolutionize on-site environmental monitoring.
Conclusion
This application note validates a rapid GC-MS protocol for efficient separation of 11 aromatic volatiles using the Agilent VF-200ms column. The method delivers exceptional resolution and speed, meeting the demands of high-throughput environmental and industrial testing.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Aromatic volatile organic compounds
2011|Agilent Technologies|Applications
Aromatic volatile organic compounds Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC analysis of nine aromatic volatiles uses an Agilent FactorFour VF-200ms column to achieve a separation in less than 12 minutes. Conditions Technique : GC Column : Agilent…
Key words
erthylbenzene, erthylbenzenexylene, xylenemeta, metaortho, orthopara, parachlorobenzene, chlorobenzenecourtesy, courtesyaromatic, aromaticbenzene, benzenetoluene, tolueneprinted, printedvolatile, volatileinjector, injectorusa, usaprogram
Aromatic volatiles
2011|Agilent Technologies|Applications
Aromatic volatiles Application Note Environmental Authors Introduction Agilent Technologies, Inc. An Agilent VF-1701ms column separates 11 aromatic volatiles in ten minutes. Conditions Technique : GC-capillary Column : Agilent FactorFourVF-1701ms, 0.25 mm x 30 m fused silica (df= 0.25 μm) (Part…
Key words
courtesy, courtesyvolatiles, volatilesaromatic, aromaticprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak
Aromatic volatiles
2011|Agilent Technologies|Applications
Aromatic volatiles Application Note Environmental Authors Introduction Agilent Technologies, Inc. Analysis of 11 aromatic volatile compounds by gas chromatography using an Agilent FactorFour VF-1301ms column takes less than 15 minutes. Conditions Technique : GC-capillary Column : Agilent FactorFour VF-1301ms, 0.25…
Key words
courtesy, courtesyvolatiles, volatilesaromatic, aromaticprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentaltechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notetemperature, temperaturepeak
Base neutrals
2011|Agilent Technologies|Applications
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. Using GC/MS, an Agilent FactorFour VF-200ms column separates 14 base neutrals in 20 minutes. Conditions Technique : GC Column : Agilent FactorFour VF-200ms, 0.25mm x 30 m (df = 1.0…
Key words
neutrals, neutralscourtesy, courtesybase, baseprinted, printedinjector, injectorusa, usaauthors, authorsprogram, programenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationpressure, pressuresolvent