Aromatic volatiles
Applications | 2011 | Agilent TechnologiesInstrumentation
Aromatic volatile compounds such as benzene, toluene, and xylenes are common environmental contaminants and industrial pollutants. Their rapid and reliable analysis is essential for monitoring air quality, ensuring compliance with regulatory standards, and protecting human health. Gas chromatography (GC) coupled with flame ionization detection (FID) remains a standard approach for separating and quantifying these compounds in environmental and industrial samples.
This application note demonstrates the performance of the Agilent VF-1701ms capillary column for the separation of 11 aromatic volatiles within a ten-minute analysis time. The study aims to optimize chromatographic conditions and showcase the column's resolution and efficiency for trace-level analysis.
The optimized GC method achieved baseline separation of 11 target compounds within ten minutes. Key observations include:
Advancements in GC column technology and faster detection systems may further reduce analysis time while enhancing resolution. Emerging applications include onsite portable GC-FID systems for field monitoring and integration with mass spectrometry for compound confirmation at trace levels. Method adaptations for complex matrices (e.g., water, soil, biological samples) and automation of sample preparation are expected to expand the utility of this approach.
The Agilent VF-1701ms column provides a rapid, efficient, and robust solution for the separation of key aromatic volatile compounds. Its performance supports routine environmental and industrial analyses, balancing speed and resolution to meet stringent analytical requirements.
No external references were cited in the source document.
GC, GC columns, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Importance of the Topic
Aromatic volatile compounds such as benzene, toluene, and xylenes are common environmental contaminants and industrial pollutants. Their rapid and reliable analysis is essential for monitoring air quality, ensuring compliance with regulatory standards, and protecting human health. Gas chromatography (GC) coupled with flame ionization detection (FID) remains a standard approach for separating and quantifying these compounds in environmental and industrial samples.
Objectives and Study Overview
This application note demonstrates the performance of the Agilent VF-1701ms capillary column for the separation of 11 aromatic volatiles within a ten-minute analysis time. The study aims to optimize chromatographic conditions and showcase the column's resolution and efficiency for trace-level analysis.
Methodology
- Technique: Gas chromatography with capillary column separation
- Sample: 100 µg/mL standard solution of aromatic volatiles in methylene chloride
- Injection: Split mode (1:100), injection volume 1.0 µL
- Carrier gas: Helium at 60 kPa resulting in a flow rate of 1 mL/min
- Temperature program: Initial hold at 45 °C for 3 minutes, then ramp to 280 °C at 10 °C/min
- Detection: Flame ionization detector (FID)
Instrumentation
- GC System: Agilent gas chromatograph equipped with FID
- Column: Agilent FactorFour VF-1701ms, 0.25 mm internal diameter × 30 m length, 0.25 µm film thickness (Part no. CP9151)
- Laboratory: Agilent Application Laboratory, Middelburg, The Netherlands
Results and Discussion
The optimized GC method achieved baseline separation of 11 target compounds within ten minutes. Key observations include:
- Sharp, symmetric peaks for mono-substituted benzenes (benzene, toluene, ethylbenzene, chlorobenzene) enabling accurate quantification.
- Effective resolution of xylene isomers (m-, p-, o-) demonstrating good selectivity of the VF-1701ms stationary phase.
- Clear separation of dichlorobenzene isomers (1,2-, 1,3-, 1,4-) and styrene, which are critical for distinguishing environmental pollutants.
- The method's linear temperature program balanced analysis speed and peak capacity.
Benefits and Practical Applications
- Rapid turnaround: Complete separation in ten minutes facilitates high sample throughput in environmental monitoring and industrial quality control.
- Sensitivity and reproducibility: Consistent retention times and peak shapes support reliable quantitation at trace levels.
- Regulatory compliance: Suitable for routine analysis of VOCs to meet environmental regulations.
- Versatility: Applicable to a range of volatile organic solvents and pollutants in various matrices.
Future Trends and Applications
Advancements in GC column technology and faster detection systems may further reduce analysis time while enhancing resolution. Emerging applications include onsite portable GC-FID systems for field monitoring and integration with mass spectrometry for compound confirmation at trace levels. Method adaptations for complex matrices (e.g., water, soil, biological samples) and automation of sample preparation are expected to expand the utility of this approach.
Conclusion
The Agilent VF-1701ms column provides a rapid, efficient, and robust solution for the separation of key aromatic volatile compounds. Its performance supports routine environmental and industrial analyses, balancing speed and resolution to meet stringent analytical requirements.
References
No external references were cited in the source document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Nitro- and chlorinated aromatics - Separation of nitro-and halogenated aromatic compounds (base neutrals)
2011|Agilent Technologies|Applications
Nitro- and chlorinated aromatics Separation of nitro-and halogenated aromatic compounds (base neutrals) Application Note Environmental Authors Introduction Agilent Technologies, Inc. Separation of 14 nitro-and halogenated aromatic compounds (base neutrals) by Agilent VF-1701ms in 20 minutes. Conditions Technique : GC-capillary Column…
Key words
nitro, nitroneutrals, neutralshalogenated, halogenatedaromatics, aromaticschlorinated, chlorinatedcourtesy, courtesyaromatic, aromaticbase, baseprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique
Aromatic volatiles
2011|Agilent Technologies|Applications
Aromatic volatiles Application Note Environmental Authors Introduction Agilent Technologies, Inc. Separation of 11 aromatic volatiles on an Agilent VF-200ms column in nine minutes. Conditions Technique : GC Column : Agilent FactorFour VF-200ms, 0.25 mm x 30 m (df = 1.0…
Key words
courtesy, courtesyvolatiles, volatilesaromatic, aromaticprinted, printedinjector, injectorusa, usaauthors, authorsprogram, programenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationpressure, pressuresolvent
Aromatic volatiles
2011|Agilent Technologies|Applications
Aromatic volatiles Application Note Environmental Authors Introduction Agilent Technologies, Inc. Analysis of 11 aromatic volatile compounds by gas chromatography using an Agilent FactorFour VF-1301ms column takes less than 15 minutes. Conditions Technique : GC-capillary Column : Agilent FactorFour VF-1301ms, 0.25…
Key words
courtesy, courtesyvolatiles, volatilesaromatic, aromaticprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentaltechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak, peaktemperature
Aromatic volatile organic compounds
2011|Agilent Technologies|Applications
Aromatic volatile organic compounds Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC analysis of nine aromatic volatiles uses an Agilent FactorFour VF-200ms column to achieve a separation in less than 12 minutes. Conditions Technique : GC Column : Agilent…
Key words
erthylbenzene, erthylbenzenexylene, xylenemeta, metaortho, orthopara, parachlorobenzene, chlorobenzenecourtesy, courtesyaromatic, aromaticbenzene, benzenetoluene, tolueneprinted, printedvolatile, volatileinjector, injectorusa, usaauthors