Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention

Presentations | 2019 | Agilent TechnologiesInstrumentation
GC, Consumables
Industries
Manufacturer
Agilent Technologies

Summary

Význam tématu


Ghost peaks and unexpected chromatographic features can undermine method robustness in GC workflows. Understanding their origins and prevention is essential for reliable analysis in applications ranging from environmental monitoring to quality control.

Cíle a přehled studie


This article examines common sources of spurious peaks in gas chromatography and outlines systematic troubleshooting strategies. It also showcases sample cleanup approaches to minimize matrix interferences.

Použitá metodika a instrumentace


The authors describe use of Agilent 7890 GC systems with split/splitless inlets, various J&W GC columns (DB-5ms, DB-624, HeavyWAX), FID and MS detectors, Ultrimetal Plus inert flow path components, and sample cleanup cartridges (Bond Elut SPE, Captiva EMR-Lipid, SLE sorbents). Ultrahigh-purity carrier gases (H2, He) with GasClean filters were recommended.

Hlavní výsledky a diskuse


Key findings include identification of ghost peaks originating from:
  • Autosampler carryover due to inadequate needle washes, dirty syringes or vial contamination.
  • Inlet issues such as septum bleed, septum coring, liner contamination and backflash of solvent expansions.
  • Column degradation manifesting as elevated baseline, tailing peaks and semivolatile ghost peaks.
  • Detector contamination or electronic artifacts.
Systematic isolation using blank runs, condensation tests and jumper-tube diagnostics enables pinpointing faulty components. Sample cleanup by solid supported liquid extraction (Chem Elut S), EMR-Lipid cartridges and liquid–liquid extraction mitigates matrix effects in oils, biological tissues and cosmetics.

Přínosy a praktické využití metody


Implementing these practices improves sensitivity, reproducibility and instrument uptime. Proactive maintenance schedules for septa, liners, gold seals, split-vent traps and columns maximize method reliability. Optimized sample cleanup enhances analyte signal-to-noise and quantitation accuracy in complex matrices.

Budoucí trendy a možnosti využití


Future developments include expanded use of inert liners with sintered frits, touchless packaging to reduce manual contamination, next-generation WAX phases (DB-HeavyWAX) with higher temperature limits, and integrated automated diagnostics within GC software. Emerging sorbent technologies for matrix removal and on-line cleanup modules will further streamline workflows.

Závěr


A structured approach to identifying and eliminating ghost peaks ensures robust GC analyses. Combining preventive maintenance, advanced inert materials, rigorous sample preparation, and systematic troubleshooting safeguards data quality and instrument performance.

Reference


Sinnott M., Ucci A. Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention. Agilent Technologies Online Application Note, October 2019.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Don’t Fear the Ghost or Unexpected Peaks:  Troubleshooting and Prevention
Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention Mark Sinnott Online Application Engineer Alexander Ucci Online Application Engineer October 10, 2019 1 October 10, 2019 Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention Agilent Restricted Exclusive…
Key words
restricted, restrictedagilent, agilentcontamination, contaminationsyringe, syringeinlet, inletghost, ghostsplit, splitjumper, jumperliner, linercleanup, cleanupemr, emrcolumn, columndirty, dirtyvent, ventvial
Making a Grand Entrance: Techniques for Efficient Sample Introduction, Inlet Types, and Maintenance
Making a Grand Entrance: Techniques for Efficient Sample Introduction, Inlet Types, and Maintenance Mark Sinnott – Application Engineer Alexander Ucci - Application Engineer March 19, 2019 1 March 29, 2019 Techniques for Efficient Sample Introduction Agilent Restricted Sample Injection Goals…
Key words
restricted, restrictedefficient, efficienttechniques, techniquessplitless, splitlessagilent, agilentsplit, splitsample, sampleintroduction, introductioninlet, inletliner, linervent, ventcolumn, columndiscrimination, discriminationsolvent, solventseptum
A Tale of Two Peaks: Troubleshooting Poor Peak Shape
A Tale of Two Peaks: Troubleshooting Poor Peak Shape
2019|Agilent Technologies|Presentations
A Tale of Two Peaks: Troubleshooting Poor Peak Shape Alexander Ucci January 17, 2019 1 Agilent Restricted “Everything was Just Fine and then this Happened!” “How do I go about Troubleshooting?” Track events- log book • Changed column, liner, septum,…
Key words
restricted, restrictedagilent, agilentproblem, probleminjector, injectorwhat, whatelectronics, electronicsdetector, detectorcolumn, columnsyringe, syringedoes, doeslook, lookoccurrence, occurrencejumper, jumperput, puttroubleshooting
Don’t Get Carried Away by Carryover: Troubleshooting GC Chromatography
Don’t Get Carried Away by Carryover: Troubleshooting GC Chromatography Alexander Ucci Online Application Engineer March 20, 2024 1 March 20, 2024 Don’t Get Carried Away by Carryover DE75430999 “Everything Was Just Fine...and Then This Happened” “How do I troubleshoot?” Track…
Key words
don’t, don’tcarryover, carryoveraway, awayget, getcarried, carriedpeaks, peaksjumper, jumperwash, washtest, testcolumn, columnblobby, blobbybleed, bleedblank, blanktroubleshooting, troubleshootingbackflush
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike