Presentations | 2019 | Agilent TechnologiesInstrumentation
GC
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Gas chromatography (GC) remains a cornerstone technique in analytical chemistry for separating and quantifying volatile and semivolatile compounds. Maintaining sharp, reproducible chromatographic peaks is critical for accurate identification and quantitation. When peak shape deteriorates, laboratories face prolonged downtime, repeated analyses and compromised data quality. A structured troubleshooting protocol helps analysts quickly isolate and correct root causes, ensuring robust performance in research, quality control and industrial applications.

Objectives and Article Overview


This article provides a systematic approach to diagnosing and resolving common peak shape and baseline issues in GC and GC–MS systems. It catalogs typical symptoms—tailing, fronting, ghost peaks, split peaks, retention shifts, resolution loss and noisy baselines—and offers step-by-step guidance for isolating problems in the injector, flow path, column, detector or electronics. Practical examples illustrate the application of diagnostic test mixes and simple bench tests to restore optimal performance.

Methodology and Instrumentation


  • Component isolation: sequentially test injector, carrier flow path, column, detector and electronics to pinpoint the source of the anomaly.
  • Diagnostic tests: non-retained compound injections, bleed profiles, condensation tests and jumper-tube bypass experiments.
  • Test mixtures: hydrocarbon standards for inertness and retention checks; mixed analyte blends to assess efficiency, activity and polarity effects.
  • Instrument platform: typical configuration includes an automated sampler (e.g. 7693A), a capillary GC (e.g. 7890B) fitted with deactivated liners, inert ferrules and transfer lines, choose capillary columns (e.g. HP-5ms UI, DB-624) and detectors (FID, TCD, ECD or MS).

Main Results and Discussion


The article identifies a spectrum of chromatographic issues. Tailing often arises from active sites in liners or columns, solved by upgrading to inert flow path consumables. Ghost peaks can signal septum bleed or carry-over, diagnosed by blank runs and gas purity checks. Split peaks and missing signals point to injector misconfiguration or syringe wear. Retention shifts and resolution loss frequently stem from leaks, column aging or contamination by matrix residues. Baseline drift and noise implicate gas flow fluctuations, electronics instability or detector contamination. Examples demonstrate how trimming contaminated column sections and targeted cleaning restores performance.

Benefits and Practical Applications


Applying a logical troubleshooting workflow minimizes guesswork and instrument downtime. Laboratories improve data reliability, reduce repeat analyses and extend column and syringe life by addressing specific fault sources. Standardized diagnostic tests accelerate problem resolution and support consistent method transfer across instruments and sites.

Future Trends and Applications


Emerging developments include automated self-diagnostic routines integrated into GC control software, AI-assisted fault prediction based on chromatogram pattern recognition and advanced inert flow path technologies. Enhanced consumables and real-time monitoring sensors will further streamline maintenance, enabling proactive interventions before performance degrades.

Conclusion


A methodical troubleshooting strategy—grounded in component isolation, standardized tests and incremental repairs—ensures dependable GC and GC–MS operation. By understanding common failure modes and employing targeted diagnostics, analysts can swiftly restore optimal peak shape and baseline stability, safeguarding analytical productivity and data integrity.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
It’s Peak Season  for Great Peak  Shape - Tips and tricks on troubleshooting in GC chromatography
It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography Mark Sinnott Alexander Ucci 6 August, 2020 1 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593 “Everything Was Just Fine.. and Then…
Key words
season, seasonpeak, peakgreat, greatshape, shapeproblem, problemwhat, whatinjector, injectorelectronics, electronicscolumn, columndetector, detectorsyringe, syringetridecane, tridecanedoes, doeslook, lookoccurrence
Lost and Found: Troubleshooting Missing Peaks in GC Analysis
Lost and Found: Troubleshooting Missing Peaks in GC Analysis Alexander Ucci February 22, 2022 1 February 22, 2022 Lost and Found DE86650637 “Everything Was Just Fine... and Then This Happened!” “How do I troubleshoot?” Track your actions/keep a logbook of…
Key words
lost, lostfound, foundwhat, whatproblem, problemtroubleshooting, troubleshootingdoes, doessyringe, syringeelectronics, electronicslook, lookdetector, detectoroccurrence, occurrenceput, puthalf, halfcolumn, columnagilent
Don’t Fear the Ghost or Unexpected Peaks:  Troubleshooting and Prevention
Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention Mark Sinnott Online Application Engineer Alexander Ucci Online Application Engineer October 10, 2019 1 October 10, 2019 Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention Agilent Restricted Exclusive…
Key words
restricted, restrictedagilent, agilentcontamination, contaminationsyringe, syringeinlet, inletghost, ghostsplit, splitjumper, jumperliner, linercleanup, cleanupemr, emrcolumn, columndirty, dirtyvent, ventvial
Don’t Fear the Ghost or Unexpected Peaks:  Troubleshooting and Prevention
Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention Mark Sinnott Online Application Engineer Alexander Ucci Online Application Engineer October 10, 2019 1 October 10, 2019 Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention Agilent Restricted Exclusive…
Key words
restricted, restrictedagilent, agilentcontamination, contaminationsyringe, syringeinlet, inletghost, ghostsplit, splitjumper, jumpercleanup, cleanupemr, emrliner, linercolumn, columndirty, dirtyvial, vialvent
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike