Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention
Presentations | 2019 | Agilent TechnologiesInstrumentation
Ghost peaks can obscure true analyte signals, compromise quantitation and lead to misinterpretation. Ensuring baseline stability and peak integrity is critical in high-resolution gas chromatography analyses for environmental monitoring, food testing, pharmaceutical quality control and research.
This work presents a systematic troubleshooting approach to identify and prevent ghost or unexpected peaks in GC and GC–MS systems. The authors illustrate common contamination sources, evaluate sample preparation techniques and propose maintenance strategies to maintain consistent, reliable chromatographic performance.
The study examines multiple potential contamination origins through targeted tests:
Key findings indicate that ghost peaks originate from multiple sources:
A holistic protocol combining gas quality control, autosampler hygiene, inlet and column maintenance and targeted sample cleanup effectively prevents ghost peaks. Adoption of these best practices enhances GC performance, data quality and lab efficiency.
GC, GC/MSD, Consumables
IndustriesManufacturerAgilent Technologies
Summary
Significance of the Topic
Ghost peaks can obscure true analyte signals, compromise quantitation and lead to misinterpretation. Ensuring baseline stability and peak integrity is critical in high-resolution gas chromatography analyses for environmental monitoring, food testing, pharmaceutical quality control and research.
Objectives and Study Overview
This work presents a systematic troubleshooting approach to identify and prevent ghost or unexpected peaks in GC and GC–MS systems. The authors illustrate common contamination sources, evaluate sample preparation techniques and propose maintenance strategies to maintain consistent, reliable chromatographic performance.
Methodology
The study examines multiple potential contamination origins through targeted tests:
- Blank injections and non-injection runs to isolate inlet and detector contributions.
- Jumper tube tests to systematically disconnect the injector, column or detector and pinpoint the faulty component.
- Condensation tests after extended equilibration at low temperature to detect system-wide contamination.
- Comparative sample preparation workflows including liquid–liquid extraction, solid supported liquid extraction (SLE) and Enhanced Matrix Removal lipid cleanup (EMR-Lipid).
- Evaluation of autosampler wash routines, syringe maintenance and vial handling to minimize carry-over.
Instrumentation
- Agilent 7890 GC with split/splitless inlet and inert flow path options.
- Agilent 5977 Series MSD or single-quadrupole MS for total ion chromatograms and deconvolution.
- Flame ionization (FID), nitrogen-phosphorus (NPD) and flame photometric (FPD) detectors.
- Autosampler with programmable needle wash and diffusion caps.
- Agilent J&W GC columns: DB-5ms, DB-624, DB-HeavyWAX and Ultra Inert coatings, plus specialty liners and ferrules.
- Sample cleanup cartridges: Chem Elut S, Captiva EMR-Lipid, Bond Elut SPE and filtration plates.
Results and Discussion
Key findings indicate that ghost peaks originate from multiple sources:
- Carrier gas impurities produce erratic baselines and siloxane peaks; ultrahigh purity gas and traps mitigate these effects.
- Autosampler carry-over arises from insufficient needle washes, contaminated wash vials and worn syringes; optimizing wash cycles, replacing consumables and using diffusion caps reduces residual peaks.
- Inlet contamination such as septa bleed, liner residues or septum coring can be diagnosed via siloxane polymer signals in TICs; regular septa, liner and trap maintenance is essential.
- Column contamination by nonvolatile matrix residues leads to ghost peaks and retention shifts; periodic trimming, bake-out and use of glass wool or sintered frit liners improve peak shape.
- Robust sample cleanup with SLE and EMR-Lipid effectively removes lipids and salts, enhancing signal-to-noise and quantitative accuracy in pesticide and PAH analyses.
Benefits and Practical Applications
- Structured troubleshooting saves time by rapidly isolating malfunctioning components.
- Enhanced cleanup workflows yield reproducible results across diverse matrices.
- Proactive consumable replacement schedules minimize downtime and repair costs.
Future Trends and Opportunities
- New inert flow path materials and coatings to reduce adsorption and thermal degradation.
- Automated, high-throughput sample preparation integrating SPE, SLE and EMR technologies.
- Predictive maintenance using real-time chromatographic data to anticipate wear.
- Novel stationary phases with extended temperature stability and broader analyte range.
Conclusion
A holistic protocol combining gas quality control, autosampler hygiene, inlet and column maintenance and targeted sample cleanup effectively prevents ghost peaks. Adoption of these best practices enhances GC performance, data quality and lab efficiency.
References
- Agilent application note 5994-0951EN: Chem Elut S performance for aromatic amines.
- Agilent application note 5994-0593EN: EU priority PAHs in seed oil using EMR-Lipid.
- Agilent application note 5991-9078EN: Carry-over ghost peaks characterization.
- Agilent brochure 5990-8532EN: Ultra Inert flow path components.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention
2019|Agilent Technologies|Presentations
Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention Mark Sinnott Online Application Engineer Alexander Ucci Online Application Engineer October 10, 2019 1 October 10, 2019 Don’t Fear the Ghost or Unexpected Peaks: Troubleshooting and Prevention Agilent Restricted Exclusive…
Key words
restricted, restrictedagilent, agilentcontamination, contaminationsyringe, syringeinlet, inletghost, ghostsplit, splitjumper, jumpercleanup, cleanupemr, emrliner, linercolumn, columndirty, dirtyvial, vialvent
Making a Grand Entrance: Techniques for Efficient Sample Introduction, Inlet Types, and Maintenance
2020|Agilent Technologies|Presentations
Making a Grand Entrance: Techniques for Efficient Sample Introduction, Inlet Types, and Maintenance Mark Sinnott – Application Engineer Alexander Ucci - Application Engineer March 19, 2019 1 March 29, 2019 Techniques for Efficient Sample Introduction Agilent Restricted Sample Injection Goals…
Key words
restricted, restrictedefficient, efficienttechniques, techniquesagilent, agilentsplitless, splitlesssplit, splitsample, sampleintroduction, introductioninlet, inletliner, linervent, ventcolumn, columndiscrimination, discriminationsolvent, solventseptum
A Tale of Two Peaks: Troubleshooting Poor Peak Shape
2019|Agilent Technologies|Presentations
A Tale of Two Peaks: Troubleshooting Poor Peak Shape Alexander Ucci January 17, 2019 1 Agilent Restricted “Everything was Just Fine and then this Happened!” “How do I go about Troubleshooting?” Track events- log book • Changed column, liner, septum,…
Key words
restricted, restrictedagilent, agilentproblem, probleminjector, injectorwhat, whatelectronics, electronicsdetector, detectorcolumn, columnsyringe, syringedoes, doeslook, lookoccurrence, occurrencejumper, jumperput, puttroubleshooting
Don’t Get Carried Away by Carryover: Troubleshooting GC Chromatography
2024|Agilent Technologies|Presentations
Don’t Get Carried Away by Carryover: Troubleshooting GC Chromatography Alexander Ucci Online Application Engineer March 20, 2024 1 March 20, 2024 Don’t Get Carried Away by Carryover DE75430999 “Everything Was Just Fine...and Then This Happened” “How do I troubleshoot?” Track…
Key words
don’t, don’tcarryover, carryoveraway, awayget, getcarried, carriedpeaks, peaksjumper, jumperwash, washtest, testcolumn, columnblobby, blobbybleed, bleedblank, blanktroubleshooting, troubleshootingbackflush