Phthalic acid esters
Applications | | GL SciencesInstrumentation
The analysis of phthalic acid esters is critical for monitoring plasticizer contamination in environmental, consumer, and industrial samples. Phthalates are widely used as plasticizers and can pose health and ecological risks when released into air or other matrices. Robust and inert gas chromatography methods enable accurate quantification of trace phthalates in complex samples.
This study evaluates the chromatographic performance of an InertCap 5MS/Sil capillary column for the separation and detection of a comprehensive mixture of 15 phthalate esters. The goal is to demonstrate baseline resolution, reproducibility, and inertness of the stationary phase when analyzing phthalates in an atmospheric sample matrix.
Key methodological parameters and instrument configuration are as follows:
The optimized method achieved baseline separation of all 15 target phthalate esters within a 32.5 min run time. Retention times increased systematically with alkyl chain length, confirming consistent phase selectivity. Peak shapes were sharp and symmetrical, indicating minimal interaction between analytes and the inert stationary phase. Mass spectral data provided clear identification without interference or carryover.
The InertCap 5MS/Sil column offers exceptional inertness towards polar phthalate esters, reducing adsorption and peak tailing. This translates into improved sensitivity and quantitation limits suitable for trace-level environmental monitoring, quality control in plastic materials, and regulatory compliance testing. The single-column method simplifies workflow and minimizes method development time.
Continued demand for phthalate analysis in emerging matrices such as electronic waste, food packaging, and indoor air will drive further refinement of GC–MS workflows. Potential advancements include coupling inert columns with high-resolution mass spectrometry for enhanced selectivity, automated sample preparation for high throughput, and development of low-bleed stationary phases for ultra-trace detection.
This application note demonstrates that the InertCap 5MS/Sil column provides reliable, high-resolution separation of a broad range of phthalic acid esters. The robust method supports accurate quantification in environmental and industrial samples, highlighting its suitability for routine analysis in laboratories focused on plasticizer monitoring.
GC/MSD, GC columns, Consumables
IndustriesManufacturerGL Sciences
Summary
Significance of the Topic
The analysis of phthalic acid esters is critical for monitoring plasticizer contamination in environmental, consumer, and industrial samples. Phthalates are widely used as plasticizers and can pose health and ecological risks when released into air or other matrices. Robust and inert gas chromatography methods enable accurate quantification of trace phthalates in complex samples.
Objectives and Study Overview
This study evaluates the chromatographic performance of an InertCap 5MS/Sil capillary column for the separation and detection of a comprehensive mixture of 15 phthalate esters. The goal is to demonstrate baseline resolution, reproducibility, and inertness of the stationary phase when analyzing phthalates in an atmospheric sample matrix.
Methodology and Instrumentation
Key methodological parameters and instrument configuration are as follows:
- Instrument: Gas chromatography–mass spectrometry (GC/MS) system configured for phthalate analysis
- Column: InertCap 5MS/Sil, 0.25 mm I.D. × 30 m, 0.25 μm film thickness
- Carrier gas: Helium at a linear velocity of 36.5 cm/s
- Injection: Splitless mode, injector temperature 250 °C, 1 min hold
- Temperature program: 60 °C initial (3 min hold), ramp at 7 °C/min to 260 °C, final hold 19 min
- Detection: MS scan mode, m/z 35–500, transfer line temperature 280 °C
- Sample preparation: Phthalate ester standard mixture at 1 μg/mL in acetone, injection volume 1 μL
Main Results and Discussion
The optimized method achieved baseline separation of all 15 target phthalate esters within a 32.5 min run time. Retention times increased systematically with alkyl chain length, confirming consistent phase selectivity. Peak shapes were sharp and symmetrical, indicating minimal interaction between analytes and the inert stationary phase. Mass spectral data provided clear identification without interference or carryover.
Benefits and Practical Applications
The InertCap 5MS/Sil column offers exceptional inertness towards polar phthalate esters, reducing adsorption and peak tailing. This translates into improved sensitivity and quantitation limits suitable for trace-level environmental monitoring, quality control in plastic materials, and regulatory compliance testing. The single-column method simplifies workflow and minimizes method development time.
Future Trends and Opportunities
Continued demand for phthalate analysis in emerging matrices such as electronic waste, food packaging, and indoor air will drive further refinement of GC–MS workflows. Potential advancements include coupling inert columns with high-resolution mass spectrometry for enhanced selectivity, automated sample preparation for high throughput, and development of low-bleed stationary phases for ultra-trace detection.
Conclusion
This application note demonstrates that the InertCap 5MS/Sil column provides reliable, high-resolution separation of a broad range of phthalic acid esters. The robust method supports accurate quantification in environmental and industrial samples, highlighting its suitability for routine analysis in laboratories focused on plasticizer monitoring.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Phthalic acid esters by the GB/T 21911-2008
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Phthalic acid esters by the GB/T 21911-2008 Data No. GA178-0644 6.0 (x100,000) 5.5 5.0 2 4.5 3 4 0 4.0 3.5 1 4 3.0 7 8 5 2.5 2.0 9 10 6 13 11…
Key words
phthalate, phthalatehth, hthdep, depinertcap, inertcapジシクロヘキシル, ジシクロヘキシルジシクロヘキシルフタレ, ジシクロヘキシルフタレdith, dithフタルサンジエチル, フタルサンジエチルphth, phthフタル, フタルhold, holdmin, minnonyl, nonylphthalic, phthalicdiisobutyl
Phthalic acid esters by the GB/T 21911-2008
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Phthalic acid esters by the GB/T 21911-2008 Data No. GA179-0644 6.0 (x100,000) 5.5 5.0 2 3 4.5 4.0 1 4 3.5 8 3.0 2.5 5 2 0 2.0 6 13 9 10 7 15…
Key words
hth, hthphthalate, phthalatephth, phthhold, holdinertcap, inertcaphexyl, hexyliddi, iddiジブチ, ジブチdib, diblli, llinonyl, nonyldioctyl, dioctyldiisobutyl, diisobutylpentyl, pentyldibutyl
Identification and Quantification of 19 Phthalic Acid Esters in Chinese Liquors Using GC-MS/MS
2013|Shimadzu|Posters
PO-CON1339E Identification and Quantification of 19 Phthalic Acid Esters in Chinese Liquors Using GC-MS/MS ASMS 2013 TP-748 Jiangtao Xing, Peng Gao, Furong Wang, Jun Fan, Shin-ichi Kawano Shimadzu Global COE, Shimadzu (China) Co., Ltd., China Identification and Quantification of 19…
Key words
phthalic, phthalicester, esteracid, acidliquors, liquorschinese, chinesebranched, branchedisomers, isomerspaes, paeschain, chaindiallyl, diallyldipentyl, dipentyldihexyl, dihexyldiisodecyl, diisodecyldipropyl, dipropyldiisononyl
Phthalic acid esters
|GL Sciences|Applications
InertSearch for GC TM InertCap® Applications Phthalic acid esters Data No. GA146-0875 (x1,000,000) 4.5 TIC 3 4.0 4 3.5 2 1 3.0 2.5 2.0 5 1.5 6 1.0 0.5 0.0 5.0 7.5 10.0 12.5 15.0 17.5 20.0 (x1,000,000) 6.0 149.00…
Key words
フタル, フタルphthalate, phthalatephthalic, phthalicoctyl, octylester, esterdidp, didpacid, acidbutyl, butylinertsearch, inertsearchinertcap, inertcapジイソデシル, ジイソデシルジイソノニル, ジイソノニルジイソノニルフタラート, ジイソノニルフタラートビニル, ビニルホルモン