USEPA Methods 502.2 and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors

Applications |  | OI AnalyticalInstrumentation
GC, Purge and Trap
Industries
Environmental
Manufacturer
Agilent Technologies, OI Analytical

Summary

Significance of the Topic


Monitoring volatile organic compounds (VOCs) in drinking water is crucial for protecting human health and ensuring compliance with environmental regulations. Aromatic and halogenated VOCs can pose significant toxicity risks even at low concentrations, making sensitive and reliable analytical methods essential for routine water quality assessment.

Objectives and Study Overview


This application note evaluates the performance of the OI Analytical Eclipse 4660 purge-and-trap (P&T) concentrator coupled with tandem photoionization (PID) and electrolytic conductivity (ELCD) detectors. The aim is to demonstrate compliance with USEPA Methods 502.2 and 8021B for the analysis of 60 target VOCs in drinking water, focusing on method development, separation efficiency, calibration, detection limits, accuracy, precision, and real-sample applicability.

Methodology and Instrumentation


  • Purge-and-Trap System: Eclipse 4660 with patented Infra-Sparge heater, Tenax/silica trap, 11-minute purge at 40 °C, 1 minute desorption at 190 °C, and 10 minute bake at 240 °C.
  • Autosampler and pH Control: OI Analytical 4551A autosampler equipped with pHDetect module for automated pH monitoring and electronic record-keeping.
  • Gas Chromatography: Agilent 6890 GC fitted with a Restek Rtx-VGC capillary column (75 m × 0.45 mm × 2.55 μm), helium carrier gas at 11 mL/min, split ratio 9:1, oven program from 35 °C to 205 °C over a 32 minute run.
  • Detection: Tandem PID (10 eV lamp, He makeup gas 25 mL/min, H₂ sweep gas 100 mL/min) and ELCD in halogen mode, both maintained at 220 °C.

Main Results and Discussion


  • Separation Performance: All 60 VOCs achieved baseline resolution within a 32 minute run, reducing analysis time by ~42% compared to previous methods.
  • Detector Selectivity: PID selectively detected unsaturated aromatics, ELCD targeted halogenated species, and combined responses provided confirmation for co-eluting compounds.
  • Calibration and Sensitivity: Six-point calibration (0.5–100 ppb) yielded correlation coefficients (R²) ≥ 0.998 for all analytes. Method detection limits ranged from 0.04 to 1.08 ppb, well below regulatory thresholds.
  • Accuracy and Precision: Average recoveries were 99.9% (PID) and 98.7% (ELCD). Relative standard deviations were ≤ 5.9% (average ~3%), demonstrating robust method precision.
  • Automated pH Monitoring: The pHDetect module eliminated manual titration and recording steps, streamlining compliance with sample preservation requirements (pH < 2).
  • Real Sample Analysis: Tap water testing confirmed method applicability, with quantifiable concentrations of several target compounds detected in both PID and ELCD chromatograms.

Benefits and Practical Applications


  • Significant reduction in run time enhances laboratory throughput and cost-effectiveness.
  • Tandem detection broadens the scope to cover both aromatic and halogenated VOCs in a single analysis.
  • Automated pH measurement ensures regulatory compliance while reducing manual labor and potential for error.
  • Adjustable calibration range allows adaptation to varying regulatory limits and monitoring needs.

Future Trends and Potential Applications


  • Integration of P&T systems with advanced laboratory information management systems (LIMS) for seamless data handling and reporting.
  • Coupling purge-and-trap with mass spectrometry to expand analyte coverage and provide enhanced compound confirmation.
  • Development of portable or field-deployable P&T units for on-site water quality monitoring.
  • Incorporation of real-time sensor technologies for continuous monitoring of emerging contaminants.

Conclusion


The combination of the Eclipse 4660 purge-and-trap concentrator with PID and ELCD tandem detectors meets all USEPA Method 502.2 and 8021B requirements while halving analysis time. The method delivers excellent sensitivity, precision, and accuracy for comprehensive VOC monitoring in drinking water.

References


  1. Ho JS. Volatile Organic Compounds in Water by Purge and Trap Capillary Column GC with PID and ELCD in Series; EPA Method 502.2. U.S. Environmental Protection Agency; 1989.
  2. United States Environmental Protection Agency. Aromatic and Halogenated Volatiles by Gas Chromatography using PID and/or ELCD; EPA Method 8021B. U.S. EPA; 1996.
  3. OI Corporation. US Patent 5,337,619. August 16, 1994.
  4. U.S. Environmental Protection Agency. Appendix B to Part 136: Definition and Procedure for the Determination of the Method Detection Limit (MDL). Federal Register; 1984.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Optimizing the Analysis of Volatile Organic Compounds
Technical Guide 1 Optimizing the Analysis of Volatile Organic Compounds Inside: EPA Method Definitions State GRO Methods Contract Laboratory Program (CLP) The Love Canal Scandal Purge and Trap Theory Sequences and Flow Paths of the Purge and Trap Unit Purge…
Key words
pack, packxylene, xylenertx, rtxtert, tertpid, pidbutylbenzene, butylbenzenepurge, purgeelcd, elcdeach, eachchloride, chloridetrap, trapbenzene, benzenedata, dataacetate, acetatevgc
Analysis of BTEX-MTBE by Purge and Trap (P&T) Concentration and Determination by Gas Chromatography and Photoionization Detection (GC/PID)
Analysis of BTEX-MTBE by Purge and Trap (P&T) Concentration and Determination by Gas Chromatography and Photoionization Detection (GC/PID) Introduction The analysis of Benzene, Toluene, Ethylbenzene, Xylenes, and Methyl-tert-butyl ether is a common analysis performed in many laboratories. BTEX compounds are…
Key words
mtbe, mtbepid, pidpurge, purgebtex, btexphotoionization, photoionizationtrap, trapidp, idpgas, gasrec, recconcentration, concentrationdesorb, desorbtemperature, temperaturedetermination, determinationethylbenzene, ethylbenzenexylene
Volatile organic compounds in water by purge and trap/capillary column GC with photoionization and electrolytic conductivity detectors in series
Volatile organic compounds in water by purge and trap/capillary column GC with photoionization and electrolytic conductivity detectors in series Application Note Environmental Authors Introduction Agilent Technologies, Inc. The 1986 amendments to the Safe Drinking Water Act require the determination of…
Key words
vocs, vocselcd, elcdphotoionization, photoionizationpid, piddetectors, detectorscolumn, columnratioing, ratioingidentification, identificationtrap, trapvolatile, volatilecompounds, compoundsselector, selectorelectrolytic, electrolyticorganic, organicaromatics
Volatile Organics US EPA Method 8021 Rtx®-VRX
Volatile Organics US EPA Method 8021 Rtx®-VRX
|Thermo Fisher Scientific|Applications
Volatile Organics US EPA Method 8021 Rtx®-VRX 39/40 PID • Good choice for waste water analysis. 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24.…
Key words
butylbenzene, butylbenzenetert, tertvrx, vrxpid, pidxylene, xylenefluorobenzene, fluorobenzenebenzene, benzenertx, rtxether, etherbutyl, butylelcd, elcdchloride, chloridesilcosteel, silcosteelmethyl, methylisopropyltoluene
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike