GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Optimizing the Analysis of Volatile Organic Compounds

Guides | 2003 | RestekInstrumentation
GC, GC/MSD, Purge and Trap, Consumables
Industries
Manufacturer
Restek

Summary

Importance of the Topic


Volatile organic compounds (VOCs) are environmental contaminants of significant concern due to their toxicity, potential carcinogenicity, and widespread occurrence in water, wastewater, and soil. Regulatory agencies such as the US EPA have established methods that employ purge and trap concentration followed by gas chromatography (GC) with selective detectors or mass spectrometry (MS) to detect VOCs at trace levels. This guide summarizes best practices and innovations that enhance the reliability, sensitivity, and throughput of VOC analysis.

Objectives and Study Overview


This guide compiles expert experience and industry best practices to:
  • Describe key EPA and state methods for drinking water, wastewater, hazardous waste, and LUST analyses.
  • Explain the theory and operation of purge and trap concentration.
  • Identify critical instrument components and troubleshooting techniques.
  • Evaluate GC system configurations, detectors, and MS interfaces.
  • Showcase practical applications and optimized conditions for a variety of VOC assays.

Methodology and Instrumentation


Purge and trap systems dynamically sweep an inert gas through the sample to transfer VOCs onto an adsorbent trap, followed by thermal desorption to the GC. Key steps include standby, wet purge, dry purge, desorb preheat, desorb, and trap bake. Optimal trap selection (e.g. Vocarb 3000 “type K” for broad volatility range) and moisture control (dry purge, MCS bypass) minimize water and methanol interference. Heated, inert transfer lines (Silcosteel® or deactivated fused silica) and low-dead-volume unions ensure sample integrity. MS interfaces such as split injection or jet separators reduce carrier flow to <1 mL/min for quadrupoles and ion traps. Detector configurations in series or tandem (PID/ELCD, FID/ELCD, or GC/MS) provide selectivity without sensitivity loss.

Main Results and Discussion


Optimized GC columns and conditions significantly improve separation and throughput:
  • Wide-bore (0.45–0.53 mm ID) Rtx®-VGC/Rtx®-VRX for PID/ELCD tandem detection of EPA Methods 502.2/8021 in 28 min.
  • Narrow-bore (0.18–0.25 mm ID) Rtx®-VMS for GC/MS Method 8260B complete in <12 min. without cryofocusing.
  • Enhanced MS tuning (PFTBA, 4-bromofluorobenzene) and leak checking yield stable baselines, low noise, and improved VOC recovery.
  • Dual purge and trap with synchronized GC/MS (Duet®) doubles sample throughput (80 samples/day).

Benefits and Practical Applications


Applying these optimized methods yields:
  • Lower method detection limits (MDLs) for regulated VOCs.
  • Reduced analysis time and consumable costs.
  • Robust performance across diverse matrices (drinking water, wastewater, soils).
  • High data quality for regulatory compliance (EPA, state programs, CLP).

Future Trends and Potential Uses


Emerging advances include:
  • Automated sample preparation with robotics and autosamplers.
  • New adsorbent materials for broader volatility ranges and reduced moisture adsorption.
  • High-throughput GC/MS and real-time screening via portable MS systems.
  • Machine-learning algorithms for automated peak deconvolution and compound identification.

Conclusion


Comprehensive understanding of purge and trap theory, careful selection of traps and columns, precise moisture control, and optimized GC and MS conditions are crucial for reliable VOC analysis. Adoption of advanced phases (Rtx®-VMS, VGC), inert components (Silcosteel®), and dual-system workflows significantly improves sensitivity, resolution, and throughput, enabling laboratories to meet stringent regulatory and contractual requirements.

References


  • EPA Method 502.2: Volatile Organic Compounds in Drinking Water.
  • EPA Method 8260B: Volatile Organic Compounds by GC/MS.
  • EPA SW-846 Methods 8010, 8020, 624, 8240.
  • Dorman FL et al., Anal. Chem. 74 (2002) 2133–2138.
  • McLafferty FW, Tureček F., Interpretation of Mass Spectra, 4th Ed., 1993.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
USEPA Methods 502.2 and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors
USEPA Methods 502.2 and 8021B using the Eclipse Purge-and-Trap (P&T) Sample Concentrator and PID/ELCD Tandem Detectors Application Note 31630508 Keywords Volatile Organic Compounds (VOCs) Purge-and-Trap (P&T) USEPA Method 502.2 USEPA Method 8021B Eclipse Purge-and-trap Sample Concentrator Photoionization Detector (PID) Electrolyic…
Key words
yes, yespid, pidelcd, elcdbutylbenzene, butylbenzenephotoionization, photoionizationpurge, purgeeclipse, eclipsetrap, trapphdetect, phdetectdetector, detectorxylene, xylenemdl, mdlelectrolytic, electrolyticdibromochloromethane, dibromochloromethanebromodichloromethane
Volatile Organics US EPA Method 8021 Rtx®-VRX
Volatile Organics US EPA Method 8021 Rtx®-VRX
|Thermo Fisher Scientific|Applications
Volatile Organics US EPA Method 8021 Rtx®-VRX 39/40 PID • Good choice for waste water analysis. 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24.…
Key words
butylbenzene, butylbenzenetert, tertvrx, vrxpid, pidxylene, xylenefluorobenzene, fluorobenzenebenzene, benzenertx, rtxether, etherbutyl, butylelcd, elcdchloride, chloridesilcosteel, silcosteelmethyl, methylisopropyltoluene
Volatile Organics US EPA Method 502.2 Rtx®-502.2
Volatile Organics US EPA Method 502.2 Rtx®-502.2
|Thermo Fisher Scientific|Applications
Volatile Organics US EPA Method 502.2 Rtx®-502.2 for more info A confirmational analysis can be found on www.restek.com/chromatograms Search for chromatogram number GC_EV00006 54/55 69 16 14 17 25 Rtx®-502.2 75m, 0.45mm ID, 2.55µm (cat.# 10986) 14 67 21 42/43…
Key words
butylbenzene, butylbenzenechloride, chloridexylene, xylenebenzene, benzenetert, tertelcd, elcdconfirmational, confirmationalpid, pidfluorobenzene, fluorobenzeneisopropyltoluene, isopropyltolueneinfo, infoallyl, allyldichlorodifluoromethane, dichlorodifluoromethanedibromomethane, dibromomethanetrichlorofluoromethane
Fast, Optimized GC Purge-and-Trap Analysis of Volatiles for Soil and Water Methods
Featured Application: Optimized volatiles analysis on an Rtx-VMS column with Restek CRMs means better VOC data in less time Fast, Optimized GC Purge-and-Trap Analysis of Volatiles for Soil and Water Methods • Save set-up time—we’ve optimized volatiles analysis for you!…
Key words
tert, tertether, etherbutylbenzene, butylbenzenemethyl, methylxylene, xylenebutyl, butylchloride, chlorideethyl, ethylamyl, amylcarbon, carbonalcohol, alcoholmethacrylate, methacrylatestyrene, styrenetba, tbavms
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike