Optimized analysis of semi-volatile organic compounds (SVOC) in environmental samples in compliance with U.S. EPA Method 8270 utilizing GC-MS
Applications | 2022 | Thermo Fisher ScientificInstrumentation
Semi-volatile organic compounds occur widely in soil water air and waste streams and pose risks to human health and ecosystems. Regulatory frameworks such as US EPA Method 8270D require sensitive reliable analysis of these contaminants to ensure environmental safety and compliance.
This study demonstrates how the Thermo Scientific ISQ 7610 GC-MS system meets all EPA Method 8270D requirements for the analysis of 76 target SVOCs in environmental sample extracts. Key goals include evaluation of dynamic tuning performance relative retention time stability calibration linearity response factor reproducibility and instrument uptime.
Sample extracts spiked with internal standards and surrogates were injected in split mode using an SSL inlet. Eleven calibration levels from 1 to 400 parts per million were prepared in methylene chloride. Dynamic MS tuning employed decafluorotriphenyl phosphine with criteria for ion abundance ratios as specified in Method 8270D. Peak tailing factors for acidic and basic check compounds were monitored. Relative retention times were calculated against deuterated or isotopically labeled internal standards. Calibration employed average response factor fits with alternative linear fits where required.
Tuning results passed all EPA ion abundance checks with stable ion ratios. Tailing factors for pentachlorophenol and benzidine were below the maximum criterion of two. No degradation of DDT was observed. Relative retention time deviations across calibration levels remained well under the allowed 0.06 units for all analytes. Calibration response factors exhibited less than 20 percent relative standard deviation for all but a few compounds, with correlation coefficients above 0.99 where required. Minimum response factor thresholds specified in Method 8270D were met or exceeded for all compounds.
The extended dynamic range of the XLXR detector enables analysis from low ppm to high ppm in a single run reducing reruns. NeverVent technology minimizes downtime by allowing ion source cleaning column changes and inlet maintenance without venting the MS. Automated reporting in Chromeleon CDS ensures streamlined compliance and data integrity for environmental laboratories.
Advances in detector design and ion source automation will further enhance sensitivity and robustness for complex environmental matrices. Integration of high-throughput sample preparation robotics and AI-driven data analysis may improve laboratory productivity and real-time contaminant monitoring. Expansion of compound libraries and use of chemical ionization modes could broaden SVOC screening capabilities.
The ISQ 7610 GC-MS with ExtractaBrite source and NeverVent technology fully satisfies EPA Method 8270D requirements for semivolatile organic compound analysis. It delivers reliable tuning stability calibration performance and extended dynamic range while significantly reducing maintenance downtime, making it well suited for high-volume environmental testing.
GC/MSD, GC/SQ
IndustriesEnvironmental
ManufacturerThermo Fisher Scientific
Summary
Importance of the topic
Semi-volatile organic compounds occur widely in soil water air and waste streams and pose risks to human health and ecosystems. Regulatory frameworks such as US EPA Method 8270D require sensitive reliable analysis of these contaminants to ensure environmental safety and compliance.
Objectives and overview
This study demonstrates how the Thermo Scientific ISQ 7610 GC-MS system meets all EPA Method 8270D requirements for the analysis of 76 target SVOCs in environmental sample extracts. Key goals include evaluation of dynamic tuning performance relative retention time stability calibration linearity response factor reproducibility and instrument uptime.
Methodology and instrumentation
Sample extracts spiked with internal standards and surrogates were injected in split mode using an SSL inlet. Eleven calibration levels from 1 to 400 parts per million were prepared in methylene chloride. Dynamic MS tuning employed decafluorotriphenyl phosphine with criteria for ion abundance ratios as specified in Method 8270D. Peak tailing factors for acidic and basic check compounds were monitored. Relative retention times were calculated against deuterated or isotopically labeled internal standards. Calibration employed average response factor fits with alternative linear fits where required.
Used instrumentation
- ISQ 7610 single quadrupole GC-MS system with XLXR detector
- TRACE 1610 GC equipped with iConnect split-splitless inlet module
- ExtractaBrite ion source with Vacuum Probe Interlock and V-lock technology
- Thermo Scientific Chromeleon Chromatography Data System
Main results and discussion
Tuning results passed all EPA ion abundance checks with stable ion ratios. Tailing factors for pentachlorophenol and benzidine were below the maximum criterion of two. No degradation of DDT was observed. Relative retention time deviations across calibration levels remained well under the allowed 0.06 units for all analytes. Calibration response factors exhibited less than 20 percent relative standard deviation for all but a few compounds, with correlation coefficients above 0.99 where required. Minimum response factor thresholds specified in Method 8270D were met or exceeded for all compounds.
Benefits and practical applications
The extended dynamic range of the XLXR detector enables analysis from low ppm to high ppm in a single run reducing reruns. NeverVent technology minimizes downtime by allowing ion source cleaning column changes and inlet maintenance without venting the MS. Automated reporting in Chromeleon CDS ensures streamlined compliance and data integrity for environmental laboratories.
Future trends and opportunities
Advances in detector design and ion source automation will further enhance sensitivity and robustness for complex environmental matrices. Integration of high-throughput sample preparation robotics and AI-driven data analysis may improve laboratory productivity and real-time contaminant monitoring. Expansion of compound libraries and use of chemical ionization modes could broaden SVOC screening capabilities.
Conclusion
The ISQ 7610 GC-MS with ExtractaBrite source and NeverVent technology fully satisfies EPA Method 8270D requirements for semivolatile organic compound analysis. It delivers reliable tuning stability calibration performance and extended dynamic range while significantly reducing maintenance downtime, making it well suited for high-volume environmental testing.
References
- US Environmental Protection Agency Method 8270D Semivolatile Organic Compounds by GC/MS Revision 5 July 2014
- Thermo Fisher Scientific Application Note AN000761 Optimized Analysis of SVOCs per EPA 8270D
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Optimized GC-MS solution for semivolatiles (SVOC) analysis in environmental samples in compliance with the U.S. EPA Method 8270D
2018|Thermo Fisher Scientific|Applications
APPLICATION NOTE 10522 Optimized GC-MS solution for semivolatiles (SVOC) analysis in environmental samples in compliance with the U.S. EPA Method 8270D Authors Richard Law,1 Cristian Cojocariu,1 Daniela Cavagnino2 Thermo Fisher Scientific, Runcorn, UK 1 2 Thermo Fisher Scientific, Milan, Italy…
Key words
phthalate, phthalatefluoranthene, fluoranthenebenzo, benzosaver, saveranthracene, anthracenepyrene, pyrenebenzyl, benzylpass, passmin, minbutyl, butylsplitless, splitlesschloroaniline, chloroanilineoctylphthalate, octylphthalatehelium, heliumboldface
Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms
2005|Merck|Applications
Application Report 351 Author: Katherine Stenerson Acquisition System: 6890 GC / 5973 MSD Notebook Reference: 1509-098 Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms US EPA Method 8270D specifies a procedure for the determination of semivolatiles in solid wastes…
Key words
benzo, benzophthalate, phthalatefluoranthene, fluoranthenebenzidine, benzidineexcellent, excellentpentachlorophenol, pentachlorophenolanthracene, anthracenepyrene, pyrenebutylbenzyl, butylbenzylremainder, remaindermillions, millionsnitrosodiphenylamine, nitrosodiphenylaminecarbazole, carbazoledibenzofuran, dibenzofuranazobenzene
Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms
2006|Merck|Applications
Application Report 391 Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms US EPA Method 8270D specifies a procedure for the determination of semivolatiles in solid wastes and ground waters. The suitability of the SLB-5ms for the separation of 72…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenebutylbenzyl, butylbenzylremainder, remaindernitrosodiphenylamine, nitrosodiphenylaminecarbazole, carbazoledibenzofuran, dibenzofuranbenzidine, benzidineazobenzene, azobenzenepropylamine, propylaminehexachlorocyclopentadiene, hexachlorocyclopentadieneisophorone
Analysis of US EPA Method 8270D Semivolatiles Using a 30 m x 0.25 mm I.D., 0.50 μm SLB-5ms
2006|Merck|Applications
Application Report 399 Analysis of US EPA Method 8270D Semivolatiles Using a 30 m x 0.25 mm I.D., 0.50 µm SLB-5ms US EPA Method 8270D specifies a procedure for the determination of semivolatiles in solid waste and ground water samples.…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenebutylbenzyl, butylbenzylnitrosodiphenylamine, nitrosodiphenylaminecarbazole, carbazoledibenzofuran, dibenzofuranazobenzene, azobenzenebenzidine, benzidinepropylamine, propylaminehexachlorocyclopentadiene, hexachlorocyclopentadieneisophorone, isophoronenitroso