Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms
Applications | 2005 | MerckInstrumentation
The analysis of semivolatile organic compounds as specified by US EPA Method 8270D plays a crucial role in environmental monitoring and waste characterization. It enables accurate detection of hazardous chemicals in soil, water, and industrial samples, supporting regulatory compliance and risk assessment.
This study evaluates the performance of the SLB-5ms capillary column in separating 72 target semivolatile analytes, eight surrogate standards, and six internal standards per EPA Method 8270D. It aims to demonstrate optimal chromatographic conditions for reliable quantitation.
The optimized chromatogram achieved baseline resolution for all 72 analytes within a 24-minute run. Key observations:
Instrumental stability was demonstrated by consistent retention times and reproducible peak shapes, validating the SLB-5ms for complex environmental matrices.
The SLB-5ms column provides a reliable, high-performance solution for US EPA Method 8270D semivolatile analysis. Its combination of thermal endurance, inertness, and chromatographic resolution makes it suitable for routine environmental monitoring and regulatory compliance.
Stenerson K. Application Report 351: Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms. Sigma-Aldrich, 2005.
GC/MSD, GC columns, Consumables
IndustriesEnvironmental
ManufacturerMerck
Summary
Significance of the Topic
The analysis of semivolatile organic compounds as specified by US EPA Method 8270D plays a crucial role in environmental monitoring and waste characterization. It enables accurate detection of hazardous chemicals in soil, water, and industrial samples, supporting regulatory compliance and risk assessment.
Objectives and Study Overview
This study evaluates the performance of the SLB-5ms capillary column in separating 72 target semivolatile analytes, eight surrogate standards, and six internal standards per EPA Method 8270D. It aims to demonstrate optimal chromatographic conditions for reliable quantitation.
Methodology and Instrumentation
- Instrumentation:
- Gas chromatograph: Agilent 6890 GC
- Mass spectrometer: Agilent 5973 MSD, electron impact ionization, scan range 40–450 amu
- Column: SLB-5ms (30 m × 0.25 mm I.D., 0.25 μm film)
- Oven program:
- 40 °C hold for 3 min
- Ramp 20 °C/min to 100 °C
- Ramp 10 °C/min to 200 °C
- Ramp 30 °C/min to 325 °C, hold 5 min
- Injection: 1.0 μL splitless (liner 4 mm I.D.), injector at 250 °C for 0.75 min
- Carrier gas: helium, pressure programming (20 psi initial, ramp to 80 psi, then to 16.5 psi, hold at 25 psi)
- Sample: 50 ng on-column of semivolatile standard mix plus eight surrogates and six internal standards at 40 ng
Main Results and Discussion
The optimized chromatogram achieved baseline resolution for all 72 analytes within a 24-minute run. Key observations:
- Sharp, well-defined peaks with minimal column bleed under high temperature
- Excellent response factors for chlorinated phenols, nitroaromatics, and polycyclic aromatic hydrocarbons
- Surrogate compounds co-eluted predictably, confirming method accuracy
Instrumental stability was demonstrated by consistent retention times and reproducible peak shapes, validating the SLB-5ms for complex environmental matrices.
Benefits and Practical Applications
- High chemical inertness reduces irreversible adsorption of active analytes
- Robust thermal stability up to 325 °C supports a broad analyte range
- Compatibility with regulatory methods (EPA 8270D, RCRA BNA) streamlines compliance workflows
- Sensitive detection at low nanogram levels facilitates trace analysis in environmental and industrial samples
Future Trends and Opportunities
- Integration with high-resolution MS and tandem MS for improved selectivity
- Adoption of two-dimensional GC for enhanced separation of complex mixtures
- Development of faster temperature programming for reduced analysis time
- Implementation of automated data processing and machine learning for peak identification
Conclusion
The SLB-5ms column provides a reliable, high-performance solution for US EPA Method 8270D semivolatile analysis. Its combination of thermal endurance, inertness, and chromatographic resolution makes it suitable for routine environmental monitoring and regulatory compliance.
Reference
Stenerson K. Application Report 351: Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms. Sigma-Aldrich, 2005.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms
2006|Merck|Applications
Application Report 391 Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms US EPA Method 8270D specifies a procedure for the determination of semivolatiles in solid wastes and ground waters. The suitability of the SLB-5ms for the separation of 72…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenebutylbenzyl, butylbenzylremainder, remaindernitrosodiphenylamine, nitrosodiphenylaminecarbazole, carbazolebenzidine, benzidineazobenzene, azobenzenedibenzofuran, dibenzofuranpropylamine, propylaminehexachlorocyclopentadiene, hexachlorocyclopentadieneisophorone
Shorter Analysis Time for Method 8270D on the SLB-5ms
2006|Merck|Applications
Application Report 398 Shorter Analysis Time for Method 8270D on the SLB-5ms Typical run times for US EPA Method 8270 range from 23 to 45 minutes. In this application, the high maximum operating temperature of the SLB-5ms was utilized to…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenebutylbenzyl, butylbenzylnitrosodiphenylamine, nitrosodiphenylaminecarbazole, carbazolebenzidine, benzidineazobenzene, azobenzenedibenzofuran, dibenzofuranpropylamine, propylaminehexachlorocyclopentadiene, hexachlorocyclopentadieneisophorone, isophoronenitroso
Analysis of US EPA Method 8270D Semivolatiles Using a 30 m x 0.25 mm I.D., 0.50 μm SLB-5ms
2006|Merck|Applications
Application Report 399 Analysis of US EPA Method 8270D Semivolatiles Using a 30 m x 0.25 mm I.D., 0.50 µm SLB-5ms US EPA Method 8270D specifies a procedure for the determination of semivolatiles in solid waste and ground water samples.…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenebutylbenzyl, butylbenzylnitrosodiphenylamine, nitrosodiphenylaminecarbazole, carbazolebenzidine, benzidineazobenzene, azobenzenedibenzofuran, dibenzofuranpropylamine, propylaminehexachlorocyclopentadiene, hexachlorocyclopentadieneisophorone, isophoronenitroso
Analysis of US EPA Method 8270D Semivolatiles Using a 20 m x 0.18 mm I.D., 0.36 μm SLB-5ms
2006|Merck|Applications
Application Report 406 Author: Katherine Stenerson Acquisition System: 6890GC/ 5973MSD Notebook Reference: 1569-44 Analysis of US EPA Method 8270D Semivolatiles Using a 20 m x 0.18 mm I.D., 0.36 µm SLB-5ms US EPA Method 8270D specifies a procedure for the…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenebutylbenzyl, butylbenzylnitrosodiphenylamine, nitrosodiphenylaminecarbazole, carbazoledibenzofuran, dibenzofuranazobenzene, azobenzenebenzidine, benzidinepropylamine, propylamineisophorone, isophoronehexachlorocyclopentadiene, hexachlorocyclopentadienenitroso