Optimized GC-MS solution for semivolatiles (SVOC) analysis in environmental samples in compliance with the U.S. EPA Method 8270D
Applications | 2018 | Thermo Fisher ScientificInstrumentation
The analysis of semivolatile organic compounds (SVOCs) in environmental matrices is essential for pollution monitoring, risk assessment, and regulatory compliance. U.S. EPA Method 8270D establishes robust criteria for GC-MS detection of SVOCs in soils, wastes, air sampling media, and water extracts. Optimizing instrumentation and methods to meet these criteria supports high-throughput laboratories by ensuring accuracy, reproducibility, and reduced operational costs.
This application note evaluates the Thermo Scientific™ ISQ™ 7000 single quadrupole GC-MS system, equipped with the ExtractaBrite ion source and NeverVent maintenance technology, for compliance with EPA Method 8270D. Key goals included demonstrating a broad working range (0.2–200 ppm) on a single column, assessing method transfer across multiple instruments, and quantifying the impact of an Instant Connect Helium Saver module on operational efficiency.
Continued advancements in ion source technology, automated maintenance workflows, and sustainable carrier gas solutions will further enhance environmental GC-MS performance. Integration of machine-learning–driven method optimization and greener analytical protocols may shape the next generation of semivolatile compound analysis.
The Thermo Scientific™ ISQ™ 7000 GC-MS system, featuring ExtractaBrite source, NeverVent maintenance, and Instant Connect Helium Saver injectors, provides a fully compliant, high-efficiency solution for EPA 8270D SVOC analysis. Laboratories benefit from extended dynamic range, reliable sensitivity, dramatic helium savings, and minimal instrument downtime.
GC/MSD, GC/SQ
IndustriesEnvironmental
ManufacturerThermo Fisher Scientific
Summary
Importance of the Topic
The analysis of semivolatile organic compounds (SVOCs) in environmental matrices is essential for pollution monitoring, risk assessment, and regulatory compliance. U.S. EPA Method 8270D establishes robust criteria for GC-MS detection of SVOCs in soils, wastes, air sampling media, and water extracts. Optimizing instrumentation and methods to meet these criteria supports high-throughput laboratories by ensuring accuracy, reproducibility, and reduced operational costs.
Objectives and Study Overview
This application note evaluates the Thermo Scientific™ ISQ™ 7000 single quadrupole GC-MS system, equipped with the ExtractaBrite ion source and NeverVent maintenance technology, for compliance with EPA Method 8270D. Key goals included demonstrating a broad working range (0.2–200 ppm) on a single column, assessing method transfer across multiple instruments, and quantifying the impact of an Instant Connect Helium Saver module on operational efficiency.
Methodology
- Analytes and Standards: Seventy-six target SVOCs, six surrogate/internal standards prepared in methylene chloride.
- Calibration Ranges: Splitless injections from 0.2 to 50 ppm; split injections (10:1) from 2 to 200 ppm on the same Thermo Scientific™ TraceGOLD™ TG-5MS column.
- Injection Modes: Standard splitless, split (10:1), and both with Instant Connect Helium Saver Injector.
- Detection: Full-scan GC-MS at 70 eV, m/z range 35–500, data managed by Chromeleon™ CDS with dedicated EPA 8270D reporting.
Used Instrumentation
- Thermo Scientific™ ISQ™ 7000 single quadrupole GC-MS with ExtractaBrite ion source, NeverVent VPI and V-lock technologies.
- Thermo Scientific™ TRACE™ 1310 GC system with Instant Connect Split-Splitless and Helium Saver Injectors.
- Thermo Scientific™ TraceGOLD™ TG-5MS 30 m × 0.25 mm × 0.25 µm column.
- Chromeleon™ Chromatography Data System.
Main Results and Discussion
- Regulatory Compliance: All 76 analytes met or exceeded EPA 8270D minimum relative response factors in both injection modes; fewer than 10% required alternative curve fitting.
- Broad Dynamic Range: A single column and liner supported quantitation across 0.2–200 ppm without loss of sensitivity or linearity.
- Retention Time Stability: Over a three-week evaluation, retention times remained reproducible across splitless, split, and Helium Saver configurations despite routine maintenance interventions.
- Helium Savings: Instant Connect Helium Saver module extended helium supply life by up to 14 years and reduced carrier gas consumption by over 90%.
- Minimal Downtime: NeverVent technology enabled ion source cleaning, column changes, and injector maintenance without venting the MS, cutting maintenance time by up to 98%.
Benefits and Practical Applications
- High Throughput: Robust ion source and extended dynamic range enable analysis of challenging environmental extracts with minimal downtime.
- Cost Efficiency: Significant reductions in helium use and maintenance downtime lower cost per analysis.
- Assured Compliance: Automated tune checks and dedicated EPA 8270D reporting templates ensure method conformity and documentation.
- Operational Flexibility: Rapid changeover between injection modes and quick maintenance support demanding laboratory schedules.
Future Trends and Opportunities
Continued advancements in ion source technology, automated maintenance workflows, and sustainable carrier gas solutions will further enhance environmental GC-MS performance. Integration of machine-learning–driven method optimization and greener analytical protocols may shape the next generation of semivolatile compound analysis.
Conclusion
The Thermo Scientific™ ISQ™ 7000 GC-MS system, featuring ExtractaBrite source, NeverVent maintenance, and Instant Connect Helium Saver injectors, provides a fully compliant, high-efficiency solution for EPA 8270D SVOC analysis. Laboratories benefit from extended dynamic range, reliable sensitivity, dramatic helium savings, and minimal instrument downtime.
References
- U.S. EPA Method 8270D: Semivolatile Organic Compounds by GC/MS, Revision 5 (2014).
- J. W. Eichelberger et al., Anal. Chem. 47 (1975) 995–1000.
- J. R. Donnelly, J. Assoc. Off. Anal. Chem. 71 (1988) 434–439.
- Y. Tondeur et al., J. Mass Spectrom. 15 (1988) 429–439.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Optimizing the Analysis of Semi-volatiles by EPA Method 8270
2016|Thermo Fisher Scientific|Applications
assessed in this study to show that significant financial costs savings can be realized throughout the lifetime of the GC-MS instrument without compromising the instrument’s performance. INTRODUCTION The U.S. EPA released the first Semi-Volatile Organic Compounds (SVOC) method by Gas…
Key words
saver, saverisq, isqthermo, thermoscientific, scientifichelium, heliumefs, efsminimum, minimumresponse, responsesurrogates, surrogatesinstant, instantmodule, modulecompounds, compoundstuning, tuningfactors, factorsconnect
Optimized analysis of semi-volatile organic compounds (SVOC) in environmental samples in compliance with U.S. EPA Method 8270 utilizing GC-MS
2022|Thermo Fisher Scientific|Applications
Application note | 000761 Mass spectrometry Optimized analysis of semi-volatile organic compounds (SVOC) in environmental samples in compliance with U.S. EPA Method 8270 utilizing GC-MS Introduction Authors Semi-volatile organic compounds (SVOCs) are ubiquitous in our environment and raise Amit Gujar…
Key words
avgcalfact, avgcalfactphthalate, phthalatebenzo, benzopass, passfluoranthene, fluoranthenefactor, factorbenzyl, benzylresponse, responseanthracene, anthracenepyrene, pyreneminimum, minimumname, namefit, fitdiethylphthalate, diethylphthalatedioctyl
Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and the Agilent HydroInert Source by Gas Chromatography/Mass Spectrometry
2022|Agilent Technologies|Applications
Application Note Environmental Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and the Agilent HydroInert Source by Gas Chromatography/Mass Spectrometry Author Angela Smith Henry, PhD Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry (GC/MS) is integral to the analysis of…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluoranthenepass, passnitrobenzene, nitrobenzeneanthracene, anthracenelinear, lineararamite, aramitemethanesulfonate, methanesulfonatediallate, diallatepyrene, pyreneazobenzene, azobenzenebenzyl, benzylnitroso, nitrosohydroinert
Analysis of Semivolatile Organic Compounds with Hydrogen Carrier Gas and HydroInert Source by Gas Chromatography/Triple Quadrupole Mass Spectrometry (GC/MS/MS)
2022|Agilent Technologies|Applications
Application Note Environmental Analysis of Semivolatile Organic Compounds with Hydrogen Carrier Gas and HydroInert Source by Gas Chromatography/Triple Quadrupole Mass Spectrometry (GC/MS/MS) Author Abstract Angela Smith Henry, PhD Agilent Technologies, Inc. Gas chromatography/mass spectrometry (GC/MS) is integral to the analysis…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluoranthenequadratic, quadraticpass, passaramite, aramitemethanesulfonate, methanesulfonatenitrosodi, nitrosodianthracene, anthracenepyrene, pyrenebenzyl, benzylbutyl, butylparathion, parathionnitroso, nitrosohydroinert