Oxygenates
Applications | | QuadrexInstrumentation
Reliable identification and quantification of oxygenated compounds in petrochemical streams is critical for product quality, process control and regulatory compliance. Alcohols and ethers such as methanol, MTBE and TAME can influence fuel performance, emissions and safety parameters. A robust gas chromatographic method ensures accurate monitoring of these components at trace and bulk levels in refinery and blending operations.
This application note describes a capillary gas chromatography–flame ionization detection (GC-FID) method for the separation of 13 common oxygenates in petroleum matrices. The goal is to demonstrate the performance of a 20 m dimethylpolysiloxane PHAT Phase column (0.18 mm i.d., 6.0 µm film) under a defined temperature program and carrier conditions to achieve baseline resolution of alcohol and ether analytes.
The analysis employs a PHAT Phase dimethylpolysiloxane capillary column (20 m × 0.18 mm i.d. × 6.0 µm film thickness, Cat. No. 007-1-20HS-6.0F). The oven program starts at 40 °C (3.5 min hold), ramps to 85 °C at 10 °C/min, then to 235 °C at 7 °C/min. Injection is performed at 220 °C with splitless mode and detection by FID at 300 °C. Helium at 28 cm/s linear velocity serves as carrier gas.
The optimized temperature gradient achieves clear separation of 13 analytes, including primary, secondary and tertiary alcohols and common fuel ethers such as MTBE, ETBE and TAME. Retention times increase with boiling point and polarity. The method provides sharp, symmetric peaks and consistent retention reproducibility. Analyte order follows: methanol, ethanol, isopropanol, t-butanol, n-propanol, MTBE, sec-butanol, DIPE, isobutanol, ETBE, t-amyl alcohol, butanol and TAME.
Integration with mass spectrometric detection could enhance selectivity and sensitivity for complex hydrocarbon matrices. Advances in column technology, such as shorter or narrower bore formats, promise faster run times and reduced carrier consumption. Green analytical approaches may adopt hydrogen or nitrogen carriers to lower operating costs and environmental impact.
The presented GC-FID method on a PHAT Phase dimethylpolysiloxane column delivers reliable separation and quantitation of key petrochemical oxygenates. It offers a practical solution for laboratories seeking robust, reproducible performance with minimal method development.
No external literature references were provided in the original application note.
GC, GC columns, Consumables
IndustriesEnergy & Chemicals
ManufacturerQuadrex
Summary
Significance of the Topic
Reliable identification and quantification of oxygenated compounds in petrochemical streams is critical for product quality, process control and regulatory compliance. Alcohols and ethers such as methanol, MTBE and TAME can influence fuel performance, emissions and safety parameters. A robust gas chromatographic method ensures accurate monitoring of these components at trace and bulk levels in refinery and blending operations.
Objectives and Overview
This application note describes a capillary gas chromatography–flame ionization detection (GC-FID) method for the separation of 13 common oxygenates in petroleum matrices. The goal is to demonstrate the performance of a 20 m dimethylpolysiloxane PHAT Phase column (0.18 mm i.d., 6.0 µm film) under a defined temperature program and carrier conditions to achieve baseline resolution of alcohol and ether analytes.
Methodology and Instrumentation Used
The analysis employs a PHAT Phase dimethylpolysiloxane capillary column (20 m × 0.18 mm i.d. × 6.0 µm film thickness, Cat. No. 007-1-20HS-6.0F). The oven program starts at 40 °C (3.5 min hold), ramps to 85 °C at 10 °C/min, then to 235 °C at 7 °C/min. Injection is performed at 220 °C with splitless mode and detection by FID at 300 °C. Helium at 28 cm/s linear velocity serves as carrier gas.
Main Results and Discussion
The optimized temperature gradient achieves clear separation of 13 analytes, including primary, secondary and tertiary alcohols and common fuel ethers such as MTBE, ETBE and TAME. Retention times increase with boiling point and polarity. The method provides sharp, symmetric peaks and consistent retention reproducibility. Analyte order follows: methanol, ethanol, isopropanol, t-butanol, n-propanol, MTBE, sec-butanol, DIPE, isobutanol, ETBE, t-amyl alcohol, butanol and TAME.
Benefits and Practical Applications
- High resolution of structurally similar oxygenates in a single run
- Rapid analysis cycle with controlled temperature programming
- Suitable for quality control in refineries and fuel blending facilities
- Quantitative FID response for trace and bulk concentration levels
Future Trends and Opportunities
Integration with mass spectrometric detection could enhance selectivity and sensitivity for complex hydrocarbon matrices. Advances in column technology, such as shorter or narrower bore formats, promise faster run times and reduced carrier consumption. Green analytical approaches may adopt hydrogen or nitrogen carriers to lower operating costs and environmental impact.
Conclusion
The presented GC-FID method on a PHAT Phase dimethylpolysiloxane column delivers reliable separation and quantitation of key petrochemical oxygenates. It offers a practical solution for laboratories seeking robust, reproducible performance with minimal method development.
References
No external literature references were provided in the original application note.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Oxygenates in RF Gasoline
|Quadrex|Applications
GC CAPILLARY COLUMN APPLICATIONS: Petrochemical Oxygenates in RF Gasoline COLUMN: Cat. No.: Temperature: Injector: Detector: Carrier Gas: 007-1, Dimethylpolysiloxane 20M. x 0.18mm I.D. x 6.0μm film PHAT Phase™ 007-1-20HS-6.0F 40° (3.5 min. hold) (10°/min.) 85°C - (7°/min.) - 235°C 220°C…
Key words
phat, phatdimethylpolysiloxane, dimethylpolysiloxaneoxygenates, oxygenatespetrochemical, petrochemicalgasoline, gasolinephase
Residual Solvents in Drugs USP Method 467
|Quadrex|Applications
GC CAPILLARY COLUMN APPLICATIONS: Industrial Solvents Residual Solvents in Drugs USP Method 467 COLUMN: Cat. No.: Temperature: Injector: Detector: Carrier Gas: 007-1, Dimethylpolysiloxane 30M. x 0.32mm I.D. x 10.0μm film 007-1-30W-10.0F 50° (4 min. hold) (5°/min.) - 200°C 220°C 290°C,…
Key words
dimethylpolysiloxane, dimethylpolysiloxanesolvents, solventsindustrial, industrialusp, uspresidual, residualdrugs, drugsmethod
Residual Solvents in Drugs (USP Method 467)
|Quadrex|Applications
GC CAPILLARY COLUMN APPLICATIONS: Clinical/Toxological Residual Solvents in Drugs (USP Method 467) COLUMN: Cat. No.: Temperature: Injector: Detector: Carrier Gas: 007-624, Cyanopropyl Phenyl Methylpolysiloxane 30M. x 0.53mm I.D. x 5.0μm film 007-624-30V-5.0F 50° (5 min. hold) (5°min.) - 70° (8°/min.)…
Key words
toxological, toxologicalcyanopropyl, cyanopropyltetrahydrofuran, tetrahydrofuranphenyl, phenylusp, uspclinical, clinicalresidual, residualdrugs, drugssolvents, solventsmethod
ASTM D5441 - Impurities in MTBE
|Quadrex|Applications
GC CAPILLARY COLUMN APPLICATIONS: Petrochemical ASTM D5441 Impurities in MTBE COLUMN: Cat. No.: Temperature: Flow: Injector: Detector: 007-1, Dimethylpolysiloxane PHAT Phase™ 20M. x 0.18mm I.D. x 6.0μm film 007-1-20HS-6.0F 40°(3 min. hold) (5°/min.) - 100° (12°/min) - 210°C 12 psi,…
Key words
petrochemical, petrochemicalmtbe