Residual Solvents in Drugs (USP Method 467)

Applications |  | QuadrexInstrumentation
GC, GC columns, Consumables
Industries
Pharma & Biopharma
Manufacturer
Quadrex

Summary

Significance of the Topic


Residual solvents in pharmaceutical products can present health risks if above threshold levels. The United States Pharmacopeia (USP) Method 467 outlines guidelines for quantifying these impurities to ensure drug safety and regulatory compliance.

Objectives and Overview of the Study


This application demonstrates a gas chromatographic method with flame ionization detection (GC-FID) to separate and quantify 21 common residual solvents in drug substances. The protocol follows USP Method 467 to achieve reliable, reproducible results consistent with pharmacopeial requirements.

Methodology and Instrumentation


The procedure employs a cyanopropyl phenyl methylpolysiloxane stationary phase column (30 m × 0.53 mm i.d., 5.0 µm film) with a temperature program starting at 50 °C (5 min hold), ramping to 70 °C at 8 °C/min, then increasing to 220 °C. Injector and detector temperatures are set at 220 °C and 300 °C respectively, using helium as the carrier gas at 10 mL/min.

Used Instrumentation


  • Column: 007-624-30V-5.0F, Cyanopropyl Phenyl Methylpolysiloxane (30 m × 0.53 mm i.d. × 5.0 µm)
  • Injector Temperature: 220 °C (split/splitless)
  • Detector: Flame Ionization Detector at 300 °C
  • Carrier Gas: Helium, 10 mL/min

Main Results and Discussion


The chromatographic method successfully resolved all 21 residual solvents listed in USP 467, including alcohols, ketones, ethers, and halogenated hydrocarbons. Baseline separation was achieved with clear peak shapes and retention times reproducible within acceptable limits. The FID provided sufficient sensitivity to detect solvents at trace levels below pharmacopeial thresholds.

Benefits and Practical Applications


  • Compliance: Meets USP 467 criteria for residual solvent testing in pharmaceuticals
  • Robustness: High resolution and repeatability across diverse solvent classes
  • Efficiency: Moderate analysis time with a single temperature program
  • Versatility: Applicable to routine quality control and method validation

Future Trends and Potential Applications


Advancements may include coupling GC to mass spectrometry for enhanced specificity, adoption of faster temperature ramps or shorter columns for reduced run times, and integration with automated sampling systems. Emerging regulatory updates and green chemistry initiatives may drive further method optimization to lower solvent usage and environmental impact.

Conclusion


The described GC-FID method provides a reliable, pharmacopeial-compliant approach for residual solvent analysis in pharmaceuticals. Its robust separation, sensitivity, and adaptability make it well-suited for quality control and regulatory testing.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Residual Solvents in Drugs USP Method 467
GC CAPILLARY COLUMN APPLICATIONS: Industrial Solvents Residual Solvents in Drugs USP Method 467 COLUMN: Cat. No.: Temperature: Injector: Detector: Carrier Gas: 007-1, Dimethylpolysiloxane 30M. x 0.32mm I.D. x 10.0μm film 007-1-30W-10.0F 50° (4 min. hold) (5°/min.) - 200°C 220°C 290°C,…
Key words
dimethylpolysiloxane, dimethylpolysiloxanesolvents, solventsindustrial, industrialusp, uspresidual, residualdrugs, drugsmethod
Oxygenates in RF Gasoline
Oxygenates in RF Gasoline
|Quadrex|Applications
GC CAPILLARY COLUMN APPLICATIONS: Petrochemical Oxygenates in RF Gasoline COLUMN: Cat. No.: Temperature: Injector: Detector: Carrier Gas: 007-1, Dimethylpolysiloxane 20M. x 0.18mm I.D. x 6.0μm film PHAT Phase™ 007-1-20HS-6.0F 40° (3.5 min. hold) (10°/min.) 85°C - (7°/min.) - 235°C 220°C…
Key words
phat, phatdimethylpolysiloxane, dimethylpolysiloxaneoxygenates, oxygenatespetrochemical, petrochemicalgasoline, gasolinephase
Oxygenates
Oxygenates
|Quadrex|Applications
GC CAPILLARY COLUMN APPLICATIONS: Petrochemical Oxygenates COLUMN: Cat. No.: Temperature: Injector: Detector: Carrier Gas: 007-1, Dimethylpolysiloxane PHAT Phase™ 20M. x 0.18mm I.D. x 6.0μm film 007-1-20HS-6.0F 40° (3.5 min. hold) (10°/min.) - 85°C (7°/min.) - 235°C 220°C 300°C, FID 28…
Key words
phat, phatdimethylpolysiloxane, dimethylpolysiloxaneoxygenates, oxygenatespetrochemical, petrochemicalphase
The Retention Index System in Gas Chromatography: McReynolds Constants
Bulletin 880 The Retention Index System in Gas Chromatography: McReynolds Constants Key Words: McReynolds constants l Rohrschneider constants retention index The selectivity of a stationary phase can be accurately estimated by determining the degree to which polar compounds are retained…
Key words
mcreynolds, mcreynoldsglycol, glycolsebacate, sebacateadipate, adipateapiezon, apiezonphthalate, phthalatesupelco, supelcophases, phasesigepal, igepalconstants, constantsneopentyl, neopentylsee, seewax, waxsuccinate, succinatesqualane
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike