Oxygenates in RF Gasoline

Applications |  | QuadrexInstrumentation
GC, GC columns, Consumables
Industries
Energy & Chemicals
Manufacturer
Quadrex

Summary

Significance of the Topic


Fuel oxygenates are commonly added to reformulated gasoline to enhance combustion efficiency and reduce emissions. Accurate quantification of these compounds is essential for regulatory compliance, performance optimization, and environmental monitoring.

Objectives and Overview


This application demonstrates a gas chromatography method for the simultaneous separation and detection of 13 oxygenated compounds in reformulated gasoline. The focus is on achieving baseline resolution within a practical analysis time using a high-polarity capillary column.

Methodology


The method employs a dimethylpolysiloxane PHAT Phase™ capillary column (20 m × 0.18 mm I.D., 6 µm film thickness). The temperature program starts at 40 °C (3.5 min hold), ramps at 10 °C/min to 85 °C, then at 7 °C/min to 235 °C. The injector is set to 220 °C and the flame ionization detector to 300 °C. Helium at 28 cm/s linear velocity serves as the carrier gas.

Applied Instrumentation


  • Capillary column: 007-1-20HS-6.0F PHAT Phase™
  • Gas chromatograph with split/splitless injector and FID
  • Carrier gas: Helium

Main Results and Discussion


The optimized temperature program provides baseline separation of 13 analytes, including short-chain alcohols and ethers such as methanol, ethanol, MTBE, ETBE, and TAME. The demonstration confirms reproducible retention times and peak shapes suitable for quantitative analysis in complex matrices.

Benefits and Practical Applications


  • Compliance with fuel quality and emission standards
  • Rapid screening of oxygenate content in gasoline production
  • Robust method adaptable to routine QC and R&D laboratories

Future Trends and Opportunities


Advances in stationary phases may further reduce analysis time and improve resolution. Integration with automated sampling and data processing will enhance throughput. Emerging detectors and coupling with mass spectrometry could expand capabilities for trace-level analysis.

Conclusion


The described GC method offers a reliable, efficient approach for the determination of multiple oxygenates in reformulated gasoline, supporting quality control and regulatory compliance.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Oxygenates
Oxygenates
|Quadrex|Applications
GC CAPILLARY COLUMN APPLICATIONS: Petrochemical Oxygenates COLUMN: Cat. No.: Temperature: Injector: Detector: Carrier Gas: 007-1, Dimethylpolysiloxane PHAT Phase™ 20M. x 0.18mm I.D. x 6.0μm film 007-1-20HS-6.0F 40° (3.5 min. hold) (10°/min.) - 85°C (7°/min.) - 235°C 220°C 300°C, FID 28…
Key words
phat, phatdimethylpolysiloxane, dimethylpolysiloxaneoxygenates, oxygenatespetrochemical, petrochemicalphase
Residual Solvents in Drugs USP Method 467
GC CAPILLARY COLUMN APPLICATIONS: Industrial Solvents Residual Solvents in Drugs USP Method 467 COLUMN: Cat. No.: Temperature: Injector: Detector: Carrier Gas: 007-1, Dimethylpolysiloxane 30M. x 0.32mm I.D. x 10.0μm film 007-1-30W-10.0F 50° (4 min. hold) (5°/min.) - 200°C 220°C 290°C,…
Key words
dimethylpolysiloxane, dimethylpolysiloxanesolvents, solventsindustrial, industrialusp, uspresidual, residualdrugs, drugsmethod
Residual Solvents in Drugs (USP Method 467)
GC CAPILLARY COLUMN APPLICATIONS: Clinical/Toxological Residual Solvents in Drugs (USP Method 467) COLUMN: Cat. No.: Temperature: Injector: Detector: Carrier Gas: 007-624, Cyanopropyl Phenyl Methylpolysiloxane 30M. x 0.53mm I.D. x 5.0μm film 007-624-30V-5.0F 50° (5 min. hold) (5°min.) - 70° (8°/min.)…
Key words
toxological, toxologicalcyanopropyl, cyanopropyltetrahydrofuran, tetrahydrofuranphenyl, phenylusp, uspclinical, clinicalresidual, residualdrugs, drugssolvents, solventsmethod
ASTM D5441 - Impurities in MTBE
ASTM D5441 - Impurities in MTBE
|Quadrex|Applications
GC CAPILLARY COLUMN APPLICATIONS: Petrochemical ASTM D5441 Impurities in MTBE COLUMN: Cat. No.: Temperature: Flow: Injector: Detector: 007-1, Dimethylpolysiloxane PHAT Phase™ 20M. x 0.18mm I.D. x 6.0μm film 007-1-20HS-6.0F 40°(3 min. hold) (5°/min.) - 100° (12°/min) - 210°C 12 psi,…
Key words
petrochemical, petrochemicalmtbe
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike