Performance Demonstration of a Modern GC-MS Instrument and Novel BFB tune for Analysis of Volatile Compounds by EPA Method 624.1 and 8260C

Applications | 2020 | ShimadzuInstrumentation
GC/MSD, Purge and Trap, GC/SQ
Industries
Environmental
Manufacturer
Shimadzu, EST Analytical

Summary

Significance of the Topic


Growing demand for clean water and stricter environmental regulations drive the need for highly sensitive and reliable analysis of volatile organic compounds (VOCs) in surface water, groundwater and industrial effluents. EPA methods 624.1 and 8260C set stringent detection limits to protect human health and ecosystems, requiring state-of-the-art gas chromatography–mass spectrometry (GC-MS) solutions capable of stable long-term performance and low method detection limits (MDLs).

Objectives and Study Overview


This study evaluates the newly released Shimadzu GCMS-QP2020 NX equipped with a novel Barret-Felton-Bridge (BFB) tune algorithm against EPA Methods 624.1 (purgeable organics) and 8260C (comprehensive VOCs) for:
  • Stability of tune over extended operation
  • Sensitivity assessment via calculation of MDLs at 0.50 µg/L and 1.00 µg/L
  • Compliance with EPA performance criteria for calibration, continuing calibration verification and detection limits

Methodology


A three-day MDL study was conducted using ten replicates of spiked blank water at two concentration levels (0.50 and 1.00 µg/L).
Key steps included:
  1. Instrument tuning: standard autotune followed by novel BFB autotune to achieve and maintain EPA mass spectral criteria (electron emission current 60 µA, ionization voltage 70 eV).
  2. Conditioning: baking the VOCARB 3000 trap and heating the GC column from 35 °C to 280 °C prior to analysis.
  3. Calibration: 10-point curve from 0.50 to 200 µg/L; evaluation by response factor %RSD (<20%) and regression (R2 ≥0.9945 for 8260C, ≥0.9912 for 624.1).
  4. Continuing calibration verification (CCV): daily laboratory control samples assessed for percent recovery and response factor criteria.
  5. MDL calculation: standard deviation of replicate analyses multiplied by the Student’s t value for 99% confidence (n−1 degrees of freedom).

Instrumentation


  • Shimadzu GCMS-QP2020 NX with BFB tuning algorithm
  • EST Analytical Econ Evolution purge and trap concentrator with Centurion WS autosampler
  • VOCARB 3000 analytical trap (baked at 260 °C)
  • Narrow-bore inlet liner for improved peak shape and high split injections
  • Full-scan mass spectrometry (m/z 35–330) with EPA-specified quantitation ions

Main Results and Discussion


  • BFB Tune Stability: single BFB tune file met EPA spectral criteria over three days without retuning.
  • Initial Calibration: >90% of compounds passed the 8260C response factor %RSD requirement (<20%); all compounds exhibited R2 values from 0.9945 to 0.9999.
  • CCV Performance: recoveries ranged from 79.8% to 110.6% (624.1) and all met 8260C QC criteria for response factor, variability, retention time and extracted ion current.
  • MDL Outcomes: for both methods, MDLs at 0.50 µg/L ranged from 0.07 to 0.40 µg/L; at 1.00 µg/L ranged from 0.09 to 0.50 µg/L, satisfying EPA detection limit requirements.

Benefits and Practical Applications


The GCMS-QP2020 NX with BFB tuning offers:
  • High sensitivity enabling lower detection limits for trace VOC monitoring
  • Extended tune stability reducing downtime and maintenance frequency
  • Robust compliance with EPA methods for municipal and industrial wastewater analysis
  • Streamlined workflows for QA/QC laboratories and environmental monitoring agencies

Future Trends and Opportunities


Anticipated developments include:
  • Further automation of tuning and calibration processes using adaptive algorithms
  • Integration with laboratory information management systems (LIMS) for real-time data tracking
  • Expansion of compound libraries and application to emerging contaminants
  • Coupling with miniaturized sample preparation or in-field GC-MS units for rapid onsite analysis

Conclusion


The Shimadzu GCMS-QP2020 NX equipped with the novel BFB tune demonstrates excellent stability and sensitivity, fully meeting EPA Methods 624.1 and 8260C requirements. Its robust performance and low MDLs position it as a leading solution for trace VOC analysis in environmental and wastewater monitoring.

Reference


  • US EPA Method 8260C: Volatile Organic Compounds by GC/MS, Revision 3, 2006
  • US EPA Method 624.1: Purgeables by GC/MS, EPA 821-R-16-008
  • A Guide to Novel GC-MS BFB Tuning for VOC Analysis, GCMS Application News No. GCMS-2001
  • Definition and Procedures for MDL Determination, Fed. Regist. 49 (209), Appendix B to Part 136, 1984
  • US EPA Method 8260D: Rev. 4, 2018

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Performance Demonstration of a New GC-MS Instrument and Novel BFB Tune for Analysis of Volatile Compounds by EPA Method 524.2
No. SSI-GCMS-2002 Gas Chromatography Mass Spectrometry Performance Demonstration of a New GC-MS Instrument and Novel BFB Tune for Analysis of Volatile Compounds by EPA Method 524.2 ■ Abstract The US Environmental Protection Agency has published several versions of its method…
Key words
butylbenzene, butylbenzenebfb, bfbmethyl, methylmdl, mdlacrylonitrile, acrylonitrilechloride, chlorideepa, epamethacrylate, methacrylatetert, terttuning, tuningrsd, rsdxylene, xyleneiodomethane, iodomethaneccv, ccvname
Shimadzu Guide to US EPA Method 8260 for Analysis of Volatile Organic Compounds in Ground Water and Solid Waste
No. SSI-GCMS-1503 Gas Chromatograph Mass Spectrometer No. GCMS-1503 Shimadzu Guide to US EPA Method 8260 for Analysis of Volatile Organic Compounds in Ground Water and Solid Waste ■ Introduction Environmental contamination has been at the forefront of government policy and…
Key words
pass, passistd, istdacetate, acetatebfb, bfbtert, tertbutylbenzene, butylbenzenerestek, restekccv, ccvrsd, rsdmethyl, methylsurr, surrbutyl, butylvinyl, vinylethyl, ethylamyl
Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters
Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters Application Note P a g e |1 Abstract Water quality laboratories across the nation are faced with both a rising level of…
Key words
ether, etherbutylbenzene, butylbenzenechloride, chlorideethyl, ethylpurge, purgeamyl, amyltemp, tempbutyl, butylvinyl, vinylmethyl, methyldichlorodifluoromethane, dichlorodifluoromethanersd, rsdchloromethane, chloromethanemethacrylate, methacrylateavg
2017 EPA Method Update Rule and EPA Method 624.1
2017 EPA Method Update Rule and EPA Method 624.1
|Agilent Technologies|Applications
2017 EPA Method Update Rule and EPA Method 624.1 Introduction Method 624 is for the determination of volatile organic compounds in industrial discharges and other liquid environmental samples by gas chromatography combined with mass spectrometry (GC/MS). The method was developed…
Key words
idoc, idocrse, rseether, ethertert, tertavg, avgtemperature, temperaturepurge, purgemethyl, methylbutylbenzene, butylbenzeneppb, ppbethyl, ethylrpd, rpdmdl, mdlcompounds, compoundsrec
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike