Shimadzu Guide to US EPA Method 8260 for Analysis of Volatile Organic Compounds in Ground Water and Solid Waste
Applications | 2015 | ShimadzuInstrumentation
Volatile organic compounds (VOCs) are critical markers of environmental contamination in groundwater and solid waste. Reliable analytical protocols are essential for regulatory compliance, human health protection and industrial quality control. US EPA Method 8260C remains a cornerstone for VOC determination by GC–MS, and continuous improvements in instrumentation enhance sensitivity, speed and data confidence.
This study evaluates Shimadzu’s GCMS-QP2010 SE system configured for US EPA Method 8260C (Revision 3, August 2006). Key aims include verifying instrument performance, optimizing chromatographic resolution for over 100 target VOCs, and demonstrating method detection limits (MDLs), calibration linearity, precision and accuracy across a wide concentration range.
Samples are introduced by purge-and-trap (P&T) in compliance with EPA protocols for drinking water, wastewater and solid matrices. A Restek SH-RXI-624Sil MS capillary column operates in constant linear velocity mode to deliver sharp, symmetrical peaks. A narrow-ID inlet liner minimizes band broadening during sample transfer. Daily instrument conditioning includes P&T trap bake cycles and oven/source interface temperature stabilization at 220 °C.
BFB tune criteria were met consistently over a six-week sequence without retuning, demonstrating robust mass calibration and stable response. Initial calibration across 0.5–200 µg/L yielded correlation coefficients R² ≥ 0.995 and relative standard deviations (RSDs) of response factors < 20% for all compounds. Periodic continuing calibration verification (CCV) standards showed recoveries of 80–120%. MDL studies at 0.5 and 1.0 µg/L achieved detection limits well below the EPA requirements for all analytes. Precision and accuracy assessments at 10 and 50 µg/L returned average recoveries between 90 and 110% with RSDs generally < 15%. Internal standards and surrogates remained within specified stability ranges (< 10% RSD and 80–120% recovery).
The optimized method offers rapid VOC analysis (16 min run time, 26 min cycle), supports a comprehensive analyte list, and meets stringent EPA performance criteria. It is well suited for environmental monitoring, industrial leachate testing, wastewater compliance and contract laboratory workflows, delivering high throughput with reliable quantitation at trace levels.
Emerging developments include integration of automated data processing for large-scale monitoring, miniaturized P&T modules for field-deployable GC–MS, and coupling with high-resolution MS for non-target screening. Enhanced software algorithms and real-time instrument diagnostics will further streamline compliance testing and broaden VOC speciation capabilities.
The Shimadzu GCMS-QP2010 SE system, combined with advanced P&T automation and optimized chromatographic conditions, delivers exceptional performance for US EPA Method 8260C. It exceeds all regulatory criteria for sensitivity, linearity, precision and accuracy, providing a robust platform for routine VOC analysis in diverse environmental matrices.
GC/MSD, Purge and Trap, GC/SQ
IndustriesEnvironmental
ManufacturerShimadzu, EST Analytical
Summary
Importance of the Topic
Volatile organic compounds (VOCs) are critical markers of environmental contamination in groundwater and solid waste. Reliable analytical protocols are essential for regulatory compliance, human health protection and industrial quality control. US EPA Method 8260C remains a cornerstone for VOC determination by GC–MS, and continuous improvements in instrumentation enhance sensitivity, speed and data confidence.
Objectives and Study Overview
This study evaluates Shimadzu’s GCMS-QP2010 SE system configured for US EPA Method 8260C (Revision 3, August 2006). Key aims include verifying instrument performance, optimizing chromatographic resolution for over 100 target VOCs, and demonstrating method detection limits (MDLs), calibration linearity, precision and accuracy across a wide concentration range.
Methodology
Samples are introduced by purge-and-trap (P&T) in compliance with EPA protocols for drinking water, wastewater and solid matrices. A Restek SH-RXI-624Sil MS capillary column operates in constant linear velocity mode to deliver sharp, symmetrical peaks. A narrow-ID inlet liner minimizes band broadening during sample transfer. Daily instrument conditioning includes P&T trap bake cycles and oven/source interface temperature stabilization at 220 °C.
Instrumentation
- Shimadzu GCMS-QP2010 SE mass spectrometer
- Restek SH-RXI-624Sil MS column (30 m × 0.25 mm × 1.4 µm)
- EST Encon Evolution purge-and-trap concentrator with Centurion Water/Soil autosampler
- VOCARB 3000 trap for volatile analyte capture
- Helium carrier gas; split ratio 40:1; full-scan MS mode m/z 35–270
Results and Discussion
BFB tune criteria were met consistently over a six-week sequence without retuning, demonstrating robust mass calibration and stable response. Initial calibration across 0.5–200 µg/L yielded correlation coefficients R² ≥ 0.995 and relative standard deviations (RSDs) of response factors < 20% for all compounds. Periodic continuing calibration verification (CCV) standards showed recoveries of 80–120%. MDL studies at 0.5 and 1.0 µg/L achieved detection limits well below the EPA requirements for all analytes. Precision and accuracy assessments at 10 and 50 µg/L returned average recoveries between 90 and 110% with RSDs generally < 15%. Internal standards and surrogates remained within specified stability ranges (< 10% RSD and 80–120% recovery).
Practical Applications and Benefits
The optimized method offers rapid VOC analysis (16 min run time, 26 min cycle), supports a comprehensive analyte list, and meets stringent EPA performance criteria. It is well suited for environmental monitoring, industrial leachate testing, wastewater compliance and contract laboratory workflows, delivering high throughput with reliable quantitation at trace levels.
Future Trends and Potential Applications
Emerging developments include integration of automated data processing for large-scale monitoring, miniaturized P&T modules for field-deployable GC–MS, and coupling with high-resolution MS for non-target screening. Enhanced software algorithms and real-time instrument diagnostics will further streamline compliance testing and broaden VOC speciation capabilities.
Conclusion
The Shimadzu GCMS-QP2010 SE system, combined with advanced P&T automation and optimized chromatographic conditions, delivers exceptional performance for US EPA Method 8260C. It exceeds all regulatory criteria for sensitivity, linearity, precision and accuracy, providing a robust platform for routine VOC analysis in diverse environmental matrices.
References
- US EPA Method 8260C, Revision 3, August 2006
- Shimadzu Guide to BFB Tuning for VOC Analysis, GCMS Application News No. GCMS-1405
- Definition and Procedure for Determination of the Method Detection Limit, Federal Register, 49(209), Appendix B, 1984
- Shimadzu Guide to US EPA Method 624 for VOC Analysis in Wastewater, GCMS Application News No. GCMS-1406
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Solutions for Environmental Analysis - Application Notebook
2018|Shimadzu|ApplicationsGuides
�SHIMADZU Excellence in Science Solutions for Environmental Analysis Application Notebook C10G-E063 Solutions for Environmental Analysis Application Notebook Introduction A clean environment is the basis for any life on earth. Whether water, soil or air – keeping the environment clean for…
Key words
pass, passwater, watergas, gasnews, newsanalysis, analysisbfb, bfbrsd, rsdwithout, withouttap, tapconcentration, concentrationistd, istdmethod, methodusing, usingmeasurement, measurementcolumn
Analysis of VOCs according to EPA Method 8260
2020|Thermo Fisher Scientific|Applications
APPLICATION NOTE 73769 Analysis of VOCs according to EPA Method 8260 Authors: Adam Ladak1, Terry Jeffers1, Dwain Cardona1, David Lee2, and Amy Nutter3 Thermo Fisher Scientific, USA Thermo Fisher Scientific, Runcorn, UK 3 Teledyne Tekmar, Mason, OH, USA 1 2…
Key words
acetate, acetatetert, tertpurge, purgeether, etherprecision, precisionbake, bakelinarity, linarityrinse, rinsemethyl, methylbutyl, butylbutylbenzene, butylbenzenesweep, sweepaccuracy, accuracydemonstration, demonstrationchloride
Analysis of VOCs according to EPA Method 8260
2020|Thermo Fisher Scientific|Applications
APPLICATION NOTE 73769 Analysis of VOCs according to EPA Method 8260 Authors: Adam Ladak1, Terry Jeffers1, Dwain Cardona1, David Lee2, and Amy Nutter3 Thermo Fisher Scientific, USA Thermo Fisher Scientific, Runcorn, UK 3 Teledyne Tekmar, Mason, OH, USA 1 2…
Key words
acetate, acetatetert, tertpurge, purgeether, etherbake, bakeprecision, precisionlinarity, linaritymethyl, methylrinse, rinsebutyl, butylbutylbenzene, butylbenzenesweep, sweepaccuracy, accuracydemonstration, demonstrationethyl
Analysis of volatile organic compounds in drinking water according to U.S. EPA Method 524.4 
2024|Thermo Fisher Scientific|Applications
Application note | 003028 Environmental Analysis of volatile organic compounds in drinking water according to U.S. EPA Method 524.4 Goal Authors Adam Ladak , Terry Jeffers , and Demonstration of an analytical method that meets the requirements outlined in U.S.…
Key words
lin, linaquatek, aquatekavgcalfact, avgcalfactether, etherlva, lvabutylbenzene, butylbenzenemethyl, methylbutyl, butylbake, bakeprecision, precisionethyl, ethylamyl, amyllumin, luminchloride, chloridepurge