Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Applications | 2015 | Teledyne LABSInstrumentation
GC/MSD, Purge and Trap, GC/SQ
Industries
Environmental
Manufacturer
Agilent Technologies, Teledyne LABS

Summary

Significance of the topic


Rising public awareness of water quality and increasing monitoring costs drive the need for streamlined analytical workflows in environmental laboratories. Validating multiple US EPA volatile organic compound (VOC) methods on a single gas chromatograph/mass spectrometer system can significantly reduce downtime and operational complexity. Efficient water analysis methods support regulatory compliance, protect public health and optimize laboratory throughput.

Objectives and study overview


This work aimed to validate four EPA methods—524.2, 524.3, 8260B and 8260C—using one automated purge and trap (P&T) system coupled to a single GC/MS platform. Key goals included:
  • Comparing sample volumes and purge parameters for methods 524.2 and 524.3
  • Assessing quality control criteria for methods 8260B and 8260C
  • Demonstrating seamless method switching without hardware changes
  • Achieving method detection limits (MDLs) and calibration performance within EPA requirements

Methodology and instrumentation


Sample preparation and analysis relied exclusively on parameter adjustments rather than physical modifications to the system. The workflow comprised two P&T methods and two GC/MS acquisition methods:
  • Purge and trap parameters optimized for 524.2 (larger volume, extended dry purge) and default water method for 524.3, 8260B and 8260C
  • GC oven and inlet settings tuned to accommodate water vapor load in 524.2 and standard VOC separations for the others

Used instrumentation


  • Teledyne Tekmar Atomx automated P&T system
  • Agilent 7890B gas chromatograph
  • Agilent 5977A mass spectrometer detector
  • Agilent DB-624 capillary column (20 m × 0.18 mm × 1.0 µm)

Main results and discussion


Method 524.2 vs. 524.3
  • 524.2 recommended 25 mL sample and 4 min desorb resulted in higher water load. Extended dry purge and increased split ratio managed moisture but required balance to maintain sensitivity for volatile analytes.
  • Optimized 524.2 parameters yielded single-digit RSDs for most compounds and MDLs around 0.1 ppb, matching performance of 524.3.
  • 524.3, with 5 mL sample and shorter desorb, afforded broader purge parameter flexibility, lower MDLs and robust calibrations (average response factors with < 10 % RSD for most targets).

Methods 8260B and 8260C QC considerations
  • Both methods share a seven-point calibration in water (0.5–200 ppb). Method 8260B limits RF variability to 15 % while 8260C allows 20 % but limits failures to < 10 % of compounds.
  • In this study, 86 of 90 compounds met the 15 % RF criterion for 8260B; all compounds met the 20 % criterion for 8260C.
  • System Performance Check Compounds and Calibration Check Compounds under 8260B all satisfied minimum RFs and RSD requirements (< 30 %), while only two of 49 priority compounds recommended by 8260C fell below suggested RF thresholds.

Benefits and practical applications


Consolidating multiple VOC methods on a single GC/MS platform enables:
  • Reduced downtime through method transitions driven solely by software parameter changes
  • Lower capital and maintenance costs by avoiding extra hardware and column changes
  • Consistent sensitivity and precision across a wide range of analytes and concentrations
  • Compliance with EPA QC requirements for regulatory and research laboratories

Future trends and opportunities


Ongoing developments may include:
  • Advanced software-driven method scheduling that adapts purge and GC parameters in real time
  • Integration of real-time moisture sensors to further optimize water management in P&T systems
  • Machine learning algorithms to predict calibration stability and flag potential performance issues before analysis
  • Expansion of single-system workflows to include other environmental matrices such as soil and air

Conclusion


This study demonstrated that four key EPA VOC methods can be effectively validated on one GC/MS system using only parameter adjustments. The Agilent 7890B/5977A in combination with the Teledyne Tekmar Atomx achieved low detection limits, robust calibrations and met all QC criteria for methods 524.2, 524.3, 8260B and 8260C. Laboratories can adopt this unified workflow to improve efficiency, reduce costs and maintain regulatory compliance.

References


  1. US EPA Method 524.2 Revision 4.1, 1995.
  2. US EPA Method 524.3 Revision 1.0, 2009.
  3. US EPA Method 8260B Revision 2, 1996.
  4. US EPA Method 8260C Revision 3, 2006.
Downloadable PDF for viewing
 

Similar PDF

Toggle
US EPA Method 8260 with the Atomx XYZ and the Thermo Scientific™ TRACE™ 1310 GC and ISQ™ MS
APPLICATION NOTE: TECHNOLOGY: INDUSTRY: AN1803 P+T VOC ENV US EPA Method 8260 with the Atomx XYZ and the Thermo Scientific™ TRACE™ 1310 GC and ISQ™ MS Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA…
Key words
temp, temppurge, purgebake, bakeacetate, acetatedesorb, desorbrinse, rinseppb, ppbtert, tertsweep, sweepether, ethertime, timecondensate, condensateaverage, averagemethyl, methylbutyl
Using Alternative Carrier Gases for US EPA VOC Drinking Water Methods
Using Alternative Carrier Gases for US EPA VOC Drinking Water Methods Application Note Abstract Limited supplies and the rising cost of helium have spurred the need for analytical methods allowing more economical alternatives. With this in mind, new purge and…
Key words
temp, tempbake, bakerinse, rinsepurge, purgevoc, vocepa, epasweep, sweepgases, gasesdesorb, desorbalternative, alternativecondensate, condensatetime, timecarrier, carrierbutylbenzene, butylbenzenemethanol
US EPA Method 8260 with the Tekmar Lumin and the Agilent 7890B GC / 5977A MS
US EPA Method 8260 with the Tekmar Lumin and the Agilent 7890B GC / 5977A MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method 8260 is used primarily by environmental labs…
Key words
temp, tempppb, ppbpurge, purgebake, bakeprecision, precisionrsd, rsdaccuracy, accuracyacetate, acetatetert, tertmdl, mdlether, etherlumin, lumindesorb, desorbsurrogate, surrogatevariable
US EPA Method 8260 with the Tekmar Lumin P&T Concentrator and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ MS
US EPA Method 8260 with the Tekmar Lumin P&T Concentrator and Thermo Scientific™ TRACE™ 1310 GC and ISQ™ MS Application Note Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e |1 Abstract US EPA Method 8260 is used primarily…
Key words
temp, tempppb, ppbpurge, purgebake, bakeprecision, precisionlumin, luminrsd, rsdaccuracy, accuracyacetate, acetatemcs, mcstert, tertmdl, mdlether, etherdesorb, desorbvariable
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike