Performance Demonstration of a New GC-MS Instrument and Novel BFB Tune for Analysis of Volatile Compounds by EPA Method 524.2
Applications | 2020 | ShimadzuInstrumentation
Volatile organic compounds in drinking water represent a significant health and regulatory concern. Reliable and sensitive analysis according to US EPA method 524.2 requires gas chromatography mass spectrometry instrumentation with robust tuning and stability under purge and trap conditions.
The study evaluated the performance of the Shimadzu GCMS-QP2020 NX coupled with a novel backflush filtration and background tuning algorithm for analysis of 30 target VOCs. Key goals included verification of tuning stability over a 15 hour period, initial calibration precision, continuing calibration verification, and determination of method detection limits at spiked levels of 0.25 and 0.50 micrograms per liter.
An initial standard autotune preceded application of the new BFB tuning algorithm. Daily spectra checks were performed over three days to confirm consistent tuning. Purge and trap used an EST Analytical Econ Evolution concentrator with a VOCARB 3000 trap, and a Centurion WS autosampler. A narrow bore inlet liner enabled improved peak shape and split injections. The GCMS-QP2020 NX acquired data in full scan mode from mass to charge ratios 35 to 330. Compliance with EPA tuning emission current and ionization voltage parameters was maintained throughout.
The enhanced sensitivity and robustness of the GCMS-QP2020 NX with novel BFB tuning reduces instability due to water carryover and purge and trap variability. This enables environmental laboratories to perform compliance testing more reliably, with confidence in meeting detection limits and quality control criteria for drinking water analysis.
The Shimadzu GCMS-QP2020 NX combined with novel BFB tuning demonstrated full compliance with US EPA method 524.2 requirements. Stable tuning, precise initial and continuing calibrations, and low method detection limits establish this system as a robust solution for routine analysis of volatile organic compounds in drinking water.
GC/MSD, GC/SQ
IndustriesEnvironmental
ManufacturerShimadzu
Summary
Importance of the Topic
Volatile organic compounds in drinking water represent a significant health and regulatory concern. Reliable and sensitive analysis according to US EPA method 524.2 requires gas chromatography mass spectrometry instrumentation with robust tuning and stability under purge and trap conditions.
Objectives and Study Overview
The study evaluated the performance of the Shimadzu GCMS-QP2020 NX coupled with a novel backflush filtration and background tuning algorithm for analysis of 30 target VOCs. Key goals included verification of tuning stability over a 15 hour period, initial calibration precision, continuing calibration verification, and determination of method detection limits at spiked levels of 0.25 and 0.50 micrograms per liter.
Methodology and Instrumentation
An initial standard autotune preceded application of the new BFB tuning algorithm. Daily spectra checks were performed over three days to confirm consistent tuning. Purge and trap used an EST Analytical Econ Evolution concentrator with a VOCARB 3000 trap, and a Centurion WS autosampler. A narrow bore inlet liner enabled improved peak shape and split injections. The GCMS-QP2020 NX acquired data in full scan mode from mass to charge ratios 35 to 330. Compliance with EPA tuning emission current and ionization voltage parameters was maintained throughout.
Instrumentation Used
- Shimadzu GCMS-QP2020 NX gas chromatograph mass spectrometer
- EST Analytical Econ Evolution purge and trap concentrator
- Centurion WS autosampler
- VOCARB 3000 P&T trap and narrow bore inlet liner
Main Results and Discussion
- BFB tuning remained stable over 30 samples in 15 hours, meeting EPA spectra acceptance criteria each day.
- Initial calibration showed relative standard deviations of response factors below 20 percent for most analytes; remaining compounds met linearity criteria with r squared values above 0.9915.
- Continuing calibration verification recoveries of internal standards and surrogate standards ranged from 95 to 103 percent, passing EPA method 524.2 criteria.
- Method detection limits at 0.25 micrograms per liter ranged from 0.020 to 0.190 micrograms per liter; at 0.50 micrograms per liter limits ranged from 0.050 to 0.790 micrograms per liter, all meeting regulatory requirements.
Benefits and Practical Applications
The enhanced sensitivity and robustness of the GCMS-QP2020 NX with novel BFB tuning reduces instability due to water carryover and purge and trap variability. This enables environmental laboratories to perform compliance testing more reliably, with confidence in meeting detection limits and quality control criteria for drinking water analysis.
Future Trends and Opportunities
- Further refinement of tuning algorithms may improve stability for challenging analytes.
- Advances in purge and trap hardware and automation can enhance throughput and reproducibility.
- Integration with high resolution mass spectrometry and real time data processing will expand applications to broader environmental monitoring.
- Benefits of machine learning for predictive maintenance and automated method optimization are emerging.
Conclusion
The Shimadzu GCMS-QP2020 NX combined with novel BFB tuning demonstrated full compliance with US EPA method 524.2 requirements. Stable tuning, precise initial and continuing calibrations, and low method detection limits establish this system as a robust solution for routine analysis of volatile organic compounds in drinking water.
References
- US Environmental Protection Agency Method 524.2 Measurement of Purgeable Organic Compounds in Water by Capillary Column Gas Chromatography Mass Spectrometry Revision 4.1 June 2009
- Guide to the Novel GC MS BFB Tuning for Analysis of Volatile Organic Compounds in Environmental Samples GC MS Application News No GCMS 2001
- Definition and Procedures for the Determination of the Method Detection Limit Federal Register 1984 49 209 Appendix B to Part 136
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Shimadzu Guide to US EPA Method 8260 for Analysis of Volatile Organic Compounds in Ground Water and Solid Waste
2015|Shimadzu|Applications
No. SSI-GCMS-1503 Gas Chromatograph Mass Spectrometer No. GCMS-1503 Shimadzu Guide to US EPA Method 8260 for Analysis of Volatile Organic Compounds in Ground Water and Solid Waste ■ Introduction Environmental contamination has been at the forefront of government policy and…
Key words
pass, passistd, istdacetate, acetatebfb, bfbtert, tertbutylbenzene, butylbenzenerestek, restekccv, ccvrsd, rsdmethyl, methylsurr, surrbutyl, butylvinyl, vinylethyl, ethylamyl
Performance Demonstration of a Modern GC-MS Instrument and Novel BFB tune for Analysis of Volatile Compounds by EPA Method 624.1 and 8260C
2020|Shimadzu|Applications
No. SSI-GCMS-2003 Gas Chromatography Mass Spectrometry No. GCMS-2003 Performance Demonstration of a Modern GC-MS Instrument and Novel BFB tune for Analysis of Volatile Compounds by EPA Method 624.1 and 8260C ■ Abstract With global population growth and demand for safe…
Key words
butylbenzene, butylbenzenexylene, xylenemdl, mdlvinyl, vinylchloride, chlorideiodomethane, iodomethanebfb, bfbepa, epaacrolein, acroleinname, nameccv, ccvchloroethane, chloroethanetetrachloroethene, tetrachloroethenebromomethane, bromomethanetrichloroethene
USEPA 524.2 Method Validation Using the Evolution Purge and Trap Concentrator and the Centurion WS Autosampler
|Agilent Technologies|Applications
Solid Phase Micro Extraction of Flavor Compounds in Beer ANNE JUREK USEPA 524.2 Method Validation Using the Evolution Purge and Trap Concentrator and the Centurion WS Autosampler Application Note Environmental Author Anne Jurek Applications Chemist EST Analytical Cincinnati, OH Abstract…
Key words
butylbenzene, butylbenzenepurge, purgeest, estcenturion, centurionjurek, jurekanne, annechloride, chlorideevolution, evolutionmethacrylate, methacrylatedesorb, desorbtrap, trapxylene, xylenebake, bakeconcentrator, concentratorbeer
US EPA Method 524.2 - Atomx XYZ and Agilent 7890B GC/5977A MS
2018|Agilent Technologies|Applications
Application Note Environmental US EPA Method 524.2 - Atomx XYZ and Agilent 7890B GC/5977A MS Author Amy Nutter Applications Chemist, Teledyne Tekmar Abstract US EPA Method 524.2 was used to determine the concentration of volatile organic compounds (VOCs) in drinking…
Key words
temp, tempbake, bakepurge, purgerinse, rinsebutylbenzene, butylbenzeneppb, ppbdesorb, desorbsweep, sweepxyz, xyzatomx, atomxmethanol, methanolchloride, chloridemdl, mdlmethacrylate, methacrylatetime