GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Lost and Found: Troubleshooting Missing Peaks in GC Analysis

Presentations | 2022 | Agilent TechnologiesInstrumentation
GC, GC/MSD
Industries
Manufacturer
Agilent Technologies

Summary

Significance of the topic


Gas chromatography remains a workhorse technique in analytical chemistry, but missing or distorted peaks compromise data integrity and laboratory productivity. Systematic troubleshooting of peak failures is essential to maintain reliable results in QA/QC, environmental monitoring, food safety and research.

Objectives and overview of the article


The article "Lost and Found: Troubleshooting Missing Peaks in GC Analysis" aims to provide a logical framework to isolate, diagnose and correct common gas chromatographic issues. It reviews sources of missing or abnormal peaks, offers practical tips and illustrates stepwise repair strategies.

Methodology and instrumentation used


The author outlines a methodology based on:
  • Problem isolation through blank runs and nonretained probes
  • Half-splitting to separate GC and detector contributions
  • Sequential variable changes to pinpoint root causes

Instrumentation and accessories include:
  • Agilent 7693A autosampler with 10 µL syringes
  • Agilent 7890B GC with multimode inlet options
  • J&W HP-5ms Ultra Inert capillary columns
  • 5977A GC/MSD for detector tuning
  • New Universal Fit GC detector jets and Agilent CrossLab CS electronic leak detector
  • UltiMetal and Ultra Inert flow components to minimize adsorption

Main results and discussion


The paper identifies five primary GC problem areas: injector, carrier gas flow, column, detector and electronics. Key observations include:
  • Missing peaks often stem from malfunctioning syringes or blocked inlets.
  • Peak shape distortions (tailing, fronting, split peaks) point to liner activity, column damage or injector leaks.
  • Response drift arises from matrix buildup, active sites in liners or tubes and requires preventive maintenance or matrix removal techniques.
  • MS tune reports detect contamination via degraded mass peak shapes or abnormal ion ratios.
  • Leak detection with cartridge-based, flow-meter-equipped devices prevents gas loss and baseline noise.

Benefits and practical application of the method


A logical, stepwise approach reduces nonproductive downtime and repetitive repairs. Tracking maintenance actions and logging system changes fosters reproducibility. End-users can anticipate component lifetimes, optimize inert flow solutions and deploy preventive QC injections to monitor system health.

Future trends and potential applications


Advances in inert surface technologies (UltiMetal Plus, Ultra Inert ferrules), mid-column backflush, automated matrix removal cartridges and AI-enabled diagnostics are on the horizon. Integration of leak detection with IoT connectivity, real-time flow monitoring and predictive maintenance algorithms will further streamline GC reliability.

Conclusion


Effective GC troubleshooting relies on isolating individual subsystems, altering one variable at a time and comparing against known good data. Regular preventive maintenance, use of inert consumables and thorough logging of events dramatically improve peak recovery and analytical confidence.

Reference


No external literature references were provided in the original text.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
A Tale of Two Peaks: Troubleshooting Poor Peak Shape
A Tale of Two Peaks: Troubleshooting Poor Peak Shape
2019|Agilent Technologies|Presentations
A Tale of Two Peaks: Troubleshooting Poor Peak Shape Alexander Ucci January 17, 2019 1 Agilent Restricted “Everything was Just Fine and then this Happened!” “How do I go about Troubleshooting?” Track events- log book • Changed column, liner, septum,…
Key words
restricted, restrictedagilent, agilentproblem, probleminjector, injectorwhat, whatelectronics, electronicsdetector, detectorcolumn, columnsyringe, syringedoes, doeslook, lookoccurrence, occurrencejumper, jumperput, puttroubleshooting
It’s Peak Season  for Great Peak  Shape - Tips and tricks on troubleshooting in GC chromatography
It’s Peak Season for Great Peak Shape Tips and tricks on troubleshooting in GC chromatography Mark Sinnott Alexander Ucci 6 August, 2020 1 August 6, 2020 It’s Peak Season for Great Peak Shape DE.2851967593 “Everything Was Just Fine.. and Then…
Key words
season, seasonpeak, peakgreat, greatshape, shapeproblem, problemwhat, whatinjector, injectorelectronics, electronicscolumn, columndetector, detectorsyringe, syringetridecane, tridecanedoes, doeslook, lookoccurrence
Mass Spectrometer Troubleshooting - What to Check When Your Results Go Wrong
Mass Spectrometer Troubleshooting What to Check When Your Results Go Wrong Angela Smith Henry, Ph.D. Applications Chemist CSD Supplies Division 1 November 29, 2018 How do we know something is wrong? Recent Chromatogram I have a “gut feeling.” Compared to…
Key words
check, checkbackground, backgroundwhat, whatgasclean, gascleanlook, lookyour, yourleaks, leaksleak, leaktune, tuneduster, dusteruse, usenoisy, noisyyou, youcolumn, columnpeaks
Mass Spectrometer Optimization
Mass Spectrometer Optimization
2020|Agilent Technologies|Presentations
Mass Spectrometer Optimization Angela Smith Henry, Ph.D. Applications Chemist CSD Supplies Division 1 January 20, 2020 Mass Spectrometer Optimization Agilent Restricted Agilent Inert Flow Solution Agilent UltiMetal Plus – TCD, FPD, NPD/FID jets Agilent UltiMetal Plus inlet weldment, shell and…
Key words
restricted, restrictedoptimization, optimizationspectrometer, spectrometeragilent, agilentmass, massleak, leakduster, dustertune, tunecheck, checknut, nuttightening, tighteningtransfer, transferwhat, whatyou, youuse
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike