Base neutrals
Applications | 2011 | Agilent TechnologiesInstrumentation
Gas chromatographic separation of base neutrals is crucial for monitoring trace-level organic contaminants in environmental, food and industrial samples. This class of compounds, including nitrosamines, chlorinated ethers and phthalates, poses potential health and regulatory concerns. Effective resolution of these analytes supports quality control, risk assessment and compliance with environmental standards.
This application note demonstrates the performance of the Agilent FactorFour VF-1701ms column for the rapid and selective separation of 14 base neutral compounds. The primary goal was to achieve complete resolution within a practical run time, enhancing throughput in routine environmental analyses.
The analysis was conducted using capillary gas chromatography with flame ionization detection (GC-FID). Key parameters included:
Courtesy of Jan Peene, Agilent Application Laboratory, Middelburg, The Netherlands.
The optimized method achieved baseline separation of 14 base neutrals in under 37 minutes. Analytes resolved include light nitrosamines (e.g., N-nitrosodimethylamine) through heavier phthalates (e.g., di-n-octyl phthalate). Peak identification was confirmed by retention time matching. The VF-1701ms phase provided selectivity for both polar and nonpolar analytes, demonstrating narrow peak widths and consistent retention across the temperature range.
Advances in column technologies and alternative detection methods (e.g., tandem mass spectrometry) may further enhance sensitivity and specificity. Coupling this methodology with automation and inline sample preparation can streamline workflows. Emerging regulatory requirements for new contaminants will drive adaptation of similar approaches across different compound classes.
The Agilent VF-1701ms column delivers reliable, high-resolution separation of diverse base neutrals within a single run, supporting comprehensive environmental analyses. This method offers a practical balance of speed, selectivity and robustness for routine laboratory use.
Agilent Technologies, Inc. Application Note A02386, 2011.
GC, GC columns, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Significance of the Topic
Gas chromatographic separation of base neutrals is crucial for monitoring trace-level organic contaminants in environmental, food and industrial samples. This class of compounds, including nitrosamines, chlorinated ethers and phthalates, poses potential health and regulatory concerns. Effective resolution of these analytes supports quality control, risk assessment and compliance with environmental standards.
Objectives and Study Overview
This application note demonstrates the performance of the Agilent FactorFour VF-1701ms column for the rapid and selective separation of 14 base neutral compounds. The primary goal was to achieve complete resolution within a practical run time, enhancing throughput in routine environmental analyses.
Methodology and Instrumentation
The analysis was conducted using capillary gas chromatography with flame ionization detection (GC-FID). Key parameters included:
- Column: Agilent FactorFour VF-1701ms, 30 m × 0.25 mm ID, 0.25 μm film thickness
- Temperature Program: Initial hold at 45 °C for 3 min, ramp at 10 °C/min to 280 °C
- Carrier Gas: Helium at 60 kPa (approx. 1 mL/min)
- Injector: Split mode, 1:100 split ratio, injection volume 1 μL of a 100 μg/mL sample in methylene chloride
- Detector: Flame ionization detector (FID)
Courtesy of Jan Peene, Agilent Application Laboratory, Middelburg, The Netherlands.
Main Results and Discussion
The optimized method achieved baseline separation of 14 base neutrals in under 37 minutes. Analytes resolved include light nitrosamines (e.g., N-nitrosodimethylamine) through heavier phthalates (e.g., di-n-octyl phthalate). Peak identification was confirmed by retention time matching. The VF-1701ms phase provided selectivity for both polar and nonpolar analytes, demonstrating narrow peak widths and consistent retention across the temperature range.
Benefits and Practical Applications
- High resolution: Distinct separation of structurally similar compounds, reducing false positives.
- Efficiency: Moderate analysis time improves laboratory throughput.
- Robustness: The column maintains performance over repeated temperature cycles.
- Applicability: Suitable for environmental monitoring, industrial QC and forensic testing.
Future Trends and Potential Applications
Advances in column technologies and alternative detection methods (e.g., tandem mass spectrometry) may further enhance sensitivity and specificity. Coupling this methodology with automation and inline sample preparation can streamline workflows. Emerging regulatory requirements for new contaminants will drive adaptation of similar approaches across different compound classes.
Conclusion
The Agilent VF-1701ms column delivers reliable, high-resolution separation of diverse base neutrals within a single run, supporting comprehensive environmental analyses. This method offers a practical balance of speed, selectivity and robustness for routine laboratory use.
Reference
Agilent Technologies, Inc. Application Note A02386, 2011.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Base neutrals
2011|Agilent Technologies|Applications
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC analysis of 14 base neutrals with an Agilent VF-200ms column achieves a separation in 28 minutes. Conditions Technique : GC Column : Agilent FactorFour VF-200ms, 0.25 mm x 30…
Key words
neutrals, neutralscourtesy, courtesybase, baseprinted, printedinjector, injectorusa, usaauthors, authorsprogram, programenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationpressure, pressuresolvent
Base neutrals
2011|Agilent Technologies|Applications
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. The Agilent VF-17ms GC column separates 14 base neutrals in 28 minutes. Conditions Technique : GC Column : Agilent VF-17ms, 0.25 mm x 30 m fused silica (df = 0.25…
Key words
neutrals, neutralsbase, baseprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak, peaktemperature
Base neutrals
2011|Agilent Technologies|Applications
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. Analysis of 14 base neutrals by gas chromatography using an Agilent FactorFour VF-1301ms column in less than 40 minutes. Conditions Technique : GC-capillary Column : Agilent FactorFour VF-1301ms, 0.25 mm…
Key words
neutrals, neutralscourtesy, courtesybase, baseprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentaltechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak, peaktemperature