Base neutrals

Applications | 2011 | Agilent TechnologiesInstrumentation
GC, GC columns, Consumables
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Separation and quantification of neutral organic contaminants are essential for assessing environmental pollution and ensuring compliance with regulations.
Compounds such as phthalates, nitrosamines and ethers can pose health risks, making reliable analytical methods crucial in environmental laboratories.

Objectives and Overview


This study demonstrates a gas chromatography (GC) method using an Agilent VF-17ms column for the rapid separation of 14 key base neutral analytes within 28 minutes.
The goal is to provide a robust protocol for routine environmental monitoring and quality control applications.

Methodology


  • Column: Agilent VF-17ms, 0.25 mm × 30 m fused silica, df = 0.25 μm
  • Temperature program: 50 °C initial, ramp at 10 °C/min to 300 °C
  • Carrier gas: Helium at 70 kPa
  • Injection: Split mode, 1:100 ratio, sample volume 1 μL
  • Detector: Flame ionization detector (FID)
  • Concentration range: 200 μg/mL

Instrumentation


  • Agilent gas chromatograph configured with FID
  • Agilent VF-17ms capillary column (Part No. CP8982)

Main Results and Discussion


The described method achieved baseline resolution of 14 base neutral compounds in under 28 minutes.
Analytes include various ethers, phthalates and nitrosamines, with consistent retention times and sharp peak shapes.
Peak identification ranged from an unknown early-eluting compound to di-n-octyl phthalate.

Benefits and Practical Applications


  • Rapid analysis suitable for high-throughput environmental screening
  • High resolution enables accurate quantitation of structurally similar compounds
  • Robustness and reproducibility support routine QA/QC operations

Future Trends and Applications


Advances may include coupling GC with mass spectrometry for enhanced selectivity, faster temperature ramps for reduced run time, and portable GC systems for field analysis.
Method adaptations could expand to broader classes of polar and semi-volatile environmental contaminants.

Conclusion


The Agilent VF-17ms GC method provides a fast, reliable approach for separating and quantifying a suite of base neutral pollutants in environmental samples.
Its robustness and simplicity make it well-suited for routine monitoring and regulatory compliance programs.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Base neutrals
Base neutrals
2011|Agilent Technologies|Applications
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC analysis of 14 base neutrals with an Agilent VF-200ms column achieves a separation in 28 minutes. Conditions Technique : GC Column : Agilent FactorFour VF-200ms, 0.25 mm x 30…
Key words
neutrals, neutralscourtesy, courtesybase, baseprinted, printedinjector, injectorusa, usaauthors, authorsprogram, programenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationpressure, pressuresolvent
Base neutrals
Base neutrals
2011|Agilent Technologies|Applications
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. Analysis of 14 base neutrals by gas chromatography using an Agilent FactorFour VF-1301ms column in less than 40 minutes. Conditions Technique : GC-capillary Column : Agilent FactorFour VF-1301ms, 0.25 mm…
Key words
neutrals, neutralscourtesy, courtesybase, baseprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentaltechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak, peaktemperature
Base neutrals
Base neutrals
2011|Agilent Technologies|Applications
Base neutrals Application Note Environmental Authors Introduction Agilent Technologies, Inc. Agilent VF-1701ms columns separate 14 base neutrals in 37 minutes. Conditions Technique : GC-capillary Column : Agilent FactorFourVF-1701ms, 0.25 mm x 30 m fused silica (df= 0.25 μm) (Part no.…
Key words
neutrals, neutralscourtesy, courtesybase, baseprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak
Analysis of Nonylphenols and Phthalates from Food Contact Packaging using GC/MS/MS
Poster Reprint ASMS 2023 Poster number WP 199 Analysis of Nonylphenols and Phthalates from Food Contact Packaging using GC/MS/MS Matthew Curtis; David A. Weil Agilent Technologies, Santa Clara, CA Introduction The production of antioxidants and surfactants used for the manufacturing…
Key words
phth, phthterephth, terephthnonylphenols, nonylphenolsdioctyl, dioctylnonylphenol, nonylphenoldiethyl, diethylphthalates, phthalatesdicyclohhexyl, dicyclohhexylisohexyl, isohexyldimethyl, dimethyldibutyl, dibutylmonoethoxylate, monoethoxylateisobutyl, isobutylxms, xmsmonoethoxylates
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike