Author
James Little/Mass Spec Interpretation Services
James Little/Mass Spec Interpretation Services
My main interest is the identification of organic compounds by mass spectrometry in organic mixtures.
Tags
Article
Academy
Software
Video
LinkedIn Logo

RI Calibration in NIST26 Chromatogram and Applying to Calculating RI in Samples

Tu, 30.6.2026
| Original article from: Mass Spec Interpretation Services/James Little
Learn how to create retention index calibrations in NIST 26, calculate RI values automatically, and improve GC-MS compound identification by combining spectral library searching with retention indices.
Video placeholder
  • Photo: James Little: RI Calibration in NIST26 Chromatogram and Applying to Calculating RI in Samples
  • Video: James Little: RI Calibration in NIST26 Chromatogram and Applying to Calculating RI in Samples

Gas chromatography–mass spectrometry (GC-MS) remains one of the most widely used analytical techniques for the identification of volatile and semi-volatile organic compounds. While electron ionization (EI) mass spectra provide highly reproducible fragmentation patterns that enable library searching against databases such as the NIST EI Mass Spectral Library, spectral similarity alone is not always sufficient for reliable compound identification. Structural isomers frequently produce nearly identical EI spectra, making additional orthogonal information essential for confident identification.

Retention indices (RI) provide exactly such an orthogonal parameter. By normalizing chromatographic retention relative to a homologous series of n-alkanes, RI values allow chromatographic behavior to be compared across instruments, laboratories, and analytical methods using similar stationary phases. When combined with EI spectral matching, retention indices substantially improve identification confidence by excluding unlikely candidates that possess similar spectra but inconsistent chromatographic behavior.

The latest NIST 26 software introduces a significantly improved workflow for RI calibration directly within the integrated Chromatogram Window. Compared with earlier stand-alone AMDIS implementations, the new workflow simplifies calibration generation while integrating RI calculations directly into deconvolution and library searching. The presentation by James Little demonstrates practical approaches for creating calibration files, refining them manually when necessary, and applying RI information during routine GC-MS data analysis.

This guide expands upon that presentation by explaining the scientific principles behind retention indices, describing recommended calibration procedures, and discussing practical considerations for implementing RI-based identification workflows in analytical laboratories.

You can download PDF Handout for this guide HERE

Why Retention Indices Matter

Electron ionization (EI) mass spectra are highly reproducible, making spectral library searching the standard approach for compound identification. However, many structurally related compounds generate nearly identical EI spectra. Examples include:

  • Structural isomers
  • Alkyl-substituted aromatic compounds
  • Branched hydrocarbons
  • Terpenes
  • Many environmental contaminants

In these situations, spectral similarity alone may not be sufficient for confident identification.

Retention indices provide an additional analytical dimension by describing chromatographic retention relative to a homologous series of n-alkanes. Because RI values are largely independent of instrument configuration when identical stationary phases are used, they provide an effective way to distinguish compounds that produce similar mass spectra.

Combining spectral matching with retention index agreement offers several advantages:

  • Increased confidence in library identifications
  • Reduced number of false-positive matches
  • Improved discrimination between structural isomers
  • Better consistency across laboratories
  • Additional verification for automated data processing

Rather than replacing spectral matching, retention indices complement library searching by providing an orthogonal confirmation parameter.

The Principle of Retention Index Calibration

Retention indices are calculated from the retention times of a series of reference compounds, most commonly n-alkanes.

During calibration:

  • A standard alkane mixture is analyzed using the same GC method as analytical samples.
  • The retention times of each alkane are recorded.
  • NIST 26 creates a calibration curve relating retention time to retention index.
  • Unknown compounds are assigned RI values through interpolation between adjacent calibration points.

The resulting calibration file can then be reused for subsequent chromatograms acquired under the same chromatographic conditions.

Because RI values are based on relative retention rather than absolute retention times, they remain considerably more robust than retention times alone when comparing results across multiple analyses.

Calibration Standards

The accuracy of calculated retention indices depends primarily on the quality of the calibration standard.

A typical calibration mixture contains a homologous series of straight-chain hydrocarbons covering the retention range of interest.

For example:

  • C7–C30 for volatile compounds
  • C8–C40 for general-purpose GC-MS
  • Extended alkane mixtures for high-boiling analytes

Each alkane defines one calibration point whose retention index is assigned according to the Kovats or linear retention index system.

For linear temperature-programmed GC methods, the retention indices typically increase by 100 units for each successive carbon number:

  • n-Heptane = RI 700
  • n-Octane = RI 800
  • n-Nonane = RI 900
  • n-Decane = RI 1000

Intermediate compounds receive interpolated RI values between neighboring calibration compounds.

The broader the calibration range, the more reliable the calculated RI values become across the complete chromatogram.

Preparing for RI Calibration

Before generating a calibration file, several conditions should be satisfied:

  • Analyze the alkane standard using exactly the same GC method that will be used for analytical samples.
  • Use identical column dimensions and stationary phase.
  • Maintain identical carrier gas flow conditions.
  • Apply the same temperature program.
  • Use identical inlet and detector settings whenever possible.
  • Ensure that all calibration peaks are well resolved and correctly identified.

Consistent chromatographic conditions are essential because any significant change in the GC method will alter retention behavior and require a new calibration.

In practice, laboratories often generate one calibration file for each validated GC method and reuse it until chromatographic conditions change significantly.

Creating a Retention Index Calibration in NIST 26

NIST 26 integrates retention index calibration directly into the Chromatogram Window, allowing users to build calibration files without relying on external software or manual calculations. The workflow is designed to be straightforward while providing sufficient flexibility for laboratories using different GC methods and alkane standards.

The general calibration workflow consists of four steps:

  • Open the chromatogram containing the alkane standard.
  • Launch the Retention Index Calibration tool.
  • Assign each alkane peak to its corresponding carbon number.
  • Generate and save the calibration file.

Once the calibration has been created, it can be applied repeatedly to chromatograms acquired under the same analytical conditions.

Loading the Calibration Chromatogram

The first step is to process a chromatogram acquired from an n-alkane calibration mixture.

Ideally, the chromatogram should exhibit:

  • Stable baseline
  • Good peak symmetry
  • Complete separation of alkane peaks
  • No overloaded peaks
  • Minimal contamination

Because retention index calculations depend entirely on the retention times of the calibration compounds, any poorly integrated or incorrectly assigned peak may introduce errors throughout the entire calibration range.

Before beginning calibration, it is therefore good practice to inspect the chromatogram carefully and verify that every calibration compound has been detected correctly.

Assigning Calibration Peaks

The calibration procedure requires each alkane peak to be associated with its correct carbon number.

For example:

  • C7
  • C8
  • C9
  • C10
  • C11
  • C12

and so forth throughout the calibration range.

NIST 26 provides tools for assigning these peaks efficiently while allowing manual correction whenever necessary.

Depending on chromatogram quality, peak assignments may be largely automatic, but the user should always verify:

  • Peak identity
  • Peak order
  • Retention time
  • Missing calibration compounds
  • Unexpected extra peaks

Errors made during this stage propagate into every calculated retention index, making this the single most important quality-control step of the calibration workflow.

Automatic Peak Recognition

When the chromatogram quality is good, NIST 26 can automatically recognize many calibration peaks.

Automatic assignment offers several advantages:

  • Faster calibration
  • Reduced manual work
  • Improved consistency
  • Lower risk of typographical errors

Nevertheless, automatic assignment should always be reviewed manually.

Unexpected chromatographic artifacts such as solvent peaks, contaminants, column bleed, or coeluting compounds may occasionally interfere with automatic recognition, particularly at the beginning or end of the chromatogram.

A brief visual inspection generally ensures that the calibration remains accurate.

Manual Peak Assignment

Manual editing remains available whenever automatic assignment requires correction.

Users can:

  • Change peak assignments.
  • Remove incorrect calibration points.
  • Insert missing peaks.
  • Correct carbon numbers.
  • Update retention times if necessary.

This flexibility is especially valuable when working with:

  • Customized alkane mixtures
  • Partial calibration standards
  • Extended high-carbon-number standards
  • Specialized chromatographic methods

Manual verification is recommended even when the automatic calibration appears successful.

Building the Calibration Curve

After all calibration compounds have been assigned, NIST 26 generates the retention index calibration.

The software establishes the relationship between:

  • Retention time
  • Carbon number
  • Retention index

Unknown compounds located between adjacent calibration compounds subsequently receive interpolated RI values.

For compounds eluting before the first calibration point or after the final alkane, RI calculations become less reliable because extrapolation is required. For this reason, the calibration standard should ideally cover the complete retention range of expected analytes.

Once generated, the calibration curve can be inspected to confirm that all calibration points follow the expected chromatographic trend.

Smooth progression without unexpected discontinuities generally indicates a successful calibration.

Saving the Calibration File

After successful calibration, the resulting RI calibration file can be saved for future analyses.

Rather than repeating calibration before every analytical sequence, laboratories typically maintain one calibration file for each validated GC method.

A calibration file can therefore be reused whenever chromatographic conditions remain unchanged, including:

  • GC column chemistry
  • Column dimensions
  • Carrier gas conditions
  • Temperature program
  • Instrument configuration

If any of these parameters are modified, a new calibration should be generated to maintain accurate retention index calculations.

Careful naming of calibration files can simplify long-term method management. Many laboratories include information such as:

  • Instrument
  • GC method
  • Column type
  • Column dimensions
  • Temperature program
  • Calibration date

This approach makes it easy to select the appropriate calibration during future data processing.

Applying RI Calibration to Sample Analysis

Once a retention index calibration has been created, it can be applied to analytical chromatograms processed in the Chromatogram Window.

The workflow is straightforward:

  • Open the sample chromatogram.
  • Load the appropriate RI calibration file.
  • Process the chromatogram.
  • Calculate retention indices automatically.
  • Include RI values in library searching.

The calculated retention index becomes an additional property associated with every detected chromatographic component.

Unlike retention time alone, the calculated RI can be compared directly with reference values stored in the NIST library or in user-created libraries.

Automatic RI Calculation

After calibration has been loaded, NIST 26 automatically determines the retention index for each detected chromatographic component.

The calculation is based on interpolation between the two surrounding alkane calibration points.

This process is performed automatically during chromatogram processing and requires no additional calculations by the user.

Each detected component therefore receives:

  • Retention time
  • Calculated retention index
  • Mass spectrum
  • Peak area
  • Additional processing metadata

These values become available throughout the remainder of the identification workflow.

RI as an Additional Identification Criterion

Mass spectral similarity remains the primary identification parameter in NIST 26. Retention indices provide an additional independent criterion that strengthens confidence in the proposed identification.

An ideal library match therefore demonstrates:

  • High spectral similarity
  • Small retention index deviation
  • Good chromatographic peak quality
  • Appropriate isotopic pattern when applicable

When these criteria agree, identification confidence increases substantially.

Conversely, disagreement between spectral similarity and retention index may indicate:

  • Incorrect library match
  • Coeluting compounds
  • Peak integration problems
  • Chromatographic method differences
  • Library compounds measured on a different stationary phase

Rather than rejecting a match automatically, the RI value provides additional evidence that can guide expert interpretation.

Improving Identification Confidence

Retention indices are particularly valuable for compounds whose EI spectra are highly similar.

Examples include:

  • Positional isomers
  • Alkyl-substituted benzenes
  • Polycyclic aromatic hydrocarbons
  • Essential oil components
  • Flavor and fragrance compounds
  • Environmental contaminants

In these cases, multiple library candidates may produce nearly identical spectral match scores.

The calculated RI often provides the additional information needed to distinguish between otherwise similar candidates.

Instead of relying solely on spectral similarity, analysts can evaluate whether both the spectrum and chromatographic behavior support the proposed identification.

This combined approach significantly reduces the probability of false-positive identifications while improving the overall robustness of automated library searching.

Searching the NIST Library Using Retention Indices

One of the major advantages of retention index calibration is the ability to compare experimentally determined RI values with those stored in spectral libraries.

When retention index information is available for library compounds, NIST 26 can display both:

  • Spectral similarity
  • Retention index agreement

These complementary parameters allow analysts to evaluate whether a proposed identification is supported by both mass spectral and chromatographic evidence.

A close agreement between the experimental and library RI values generally increases confidence in the identification, whereas large deviations may indicate that the proposed library match should be examined more carefully.

It is important to remember that retention indices depend on the chromatographic stationary phase. Meaningful RI comparisons should therefore be performed only between data acquired on equivalent or very similar column chemistries.

Building User Libraries with Retention Indices

In addition to using the extensive retention index information available within the NIST database, laboratories can create their own spectral libraries that include experimentally determined RI values.

User libraries containing retention indices provide several advantages:

  • Consistent identification criteria across laboratories
  • Improved confidence for frequently analyzed compounds
  • Better support for regulated analytical workflows
  • Standardized identification of laboratory-specific compounds
  • Long-term accumulation of high-quality reference data

Many laboratories routinely acquire reference standards under validated chromatographic conditions and build internal libraries that combine:

  • Electron ionization mass spectra
  • Retention indices
  • Compound metadata
  • Analytical notes
  • Quality-control information

Over time, these customized libraries become valuable institutional resources that improve consistency and reduce uncertainty in routine analyses.

Best Practices for Reliable RI Calibration

Accurate retention indices require careful calibration and consistent chromatographic conditions.

The following recommendations help ensure reliable RI calculations:

  • Use fresh and well-characterized alkane standards.
  • Cover the complete retention range of expected analytes.
  • Verify every calibration peak before saving the calibration.
  • Recalibrate whenever chromatographic conditions change.
  • Avoid using calibration standards with missing alkane peaks.
  • Monitor column performance over time.
  • Replace calibration files after significant column trimming or column replacement.
  • Use the same stationary phase for calibration and sample analysis.

Routine verification of calibration quality helps maintain consistent RI values over long analytical sequences and minimizes the risk of systematic identification errors.

Common Sources of Error

Although retention index calculations are generally robust, several factors can reduce their accuracy.

Common causes include:

  • Incorrect alkane peak assignment
  • Missing calibration compounds
  • Poor chromatographic peak integration
  • Changes in carrier gas flow
  • Altered temperature programs
  • Column aging
  • Column replacement
  • Different stationary phases
  • Contaminated calibration mixtures

Even relatively small chromatographic changes can produce measurable shifts in retention indices.

For this reason, retention index calibration should be considered part of routine method maintenance rather than a one-time procedure.

Practical Applications

Retention index calibration benefits a wide range of GC-MS applications.

Typical examples include:

  • Environmental analysis
  • Food and flavor analysis
  • Fragrance characterization
  • Petrochemical analysis
  • Clinical toxicology
  • Forensic investigations
  • Pharmaceutical impurity profiling
  • Metabolomics
  • Natural product research

In many of these applications, large numbers of structurally similar compounds are encountered. Combining spectral matching with retention index confirmation substantially improves identification reliability and reduces manual review.

Summary

Retention index calibration represents an important enhancement to GC-MS library searching by combining chromatographic and mass spectral information within a single identification workflow.

NIST 26 integrates RI calibration directly into the Chromatogram Window, allowing users to create calibration files, calculate retention indices automatically, and compare experimental values with those stored in spectral libraries.

By incorporating retention indices into routine data processing, laboratories can:

  • Increase confidence in compound identification.
  • Reduce false-positive library matches.
  • Differentiate structurally similar compounds.
  • Improve consistency between analytical runs.
  • Standardize identification workflows.
  • Build high-quality in-house reference libraries.

When combined with the powerful deconvolution, chromatogram processing, and library searching capabilities introduced in NIST 26, retention index calibration provides an additional level of analytical confidence for both routine and research GC-MS applications.

James Little/Mass Spec Interpretation Services
LinkedIn Logo
 

Related content

Is Your Next Sip Safe? A Simple, Rapid Method for Measuring volatile PFAS in Juices.

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, HeadSpace
Manufacturer
Shimadzu
Industries
Food & Agriculture

Thermo Scientific™ Orbitrap Exploris™ GC S Mass Spectrometer

Brochures and specifications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD, GC/MS/MS, GC/Orbitrap, GC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Other

Operational efficiency for your GC and GC/MS labs with the Agilent 8860B gas chromatography system

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8890B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8860B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other
 

Related articles

Effective Use of EI Mass Spectral Databases in Unknown Identifications - Part VI: Creating and Using Retention Indices in NIST Software
Article | Video

Effective Use of EI Mass Spectral Databases in Unknown Identifications - Part VI: Creating and Using Retention Indices in NIST Software

This series of webinars will demonstrate the use of the NIST software suite including the NIST Search, AMDIS, and MS Interpreter for searching EI libraries.
James Little/Mass Spec Interpretation Services
tag
share
more
Leveraging Local Libraries for Positively Parsimonious Peak Tables (James Harynuk, MDCW 2026)
Presentation | Video

Leveraging Local Libraries for Positively Parsimonious Peak Tables (James Harynuk, MDCW 2026)

Discover how custom MS libraries and retention index filtering improve GC×GC–MS peak identification, alignment, and data quality in complex samples.
The Multidimensional Chromatography Workshop
tag
share
more
Processing EI GC-MS Data in Chromatogram Window NIST26
Article | Academy

Processing EI GC-MS Data in Chromatogram Window NIST26

Learn how to process EI GC-MS data in NIST 26 using integrated AMDIS deconvolution, library searching, filtering, spectral review, and Excel reporting for confident compound identification.
James Little/Mass Spec Interpretation Services
tag
share
more
Processing of Accurate Mass GC-MS with NIST26 Chromatogram
Article | Academy

Processing of Accurate Mass GC-MS with NIST26 Chromatogram

Learn how to process accurate-mass GC-MS data in NIST 26 using integrated deconvolution, library searching, MS Interpreter validation, filtering tools, and Excel reporting workflows.
James Little/Mass Spec Interpretation Services
tag
share
more
 

Related content

Is Your Next Sip Safe? A Simple, Rapid Method for Measuring volatile PFAS in Juices.

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, HeadSpace
Manufacturer
Shimadzu
Industries
Food & Agriculture

Thermo Scientific™ Orbitrap Exploris™ GC S Mass Spectrometer

Brochures and specifications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD, GC/MS/MS, GC/Orbitrap, GC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Other

Operational efficiency for your GC and GC/MS labs with the Agilent 8860B gas chromatography system

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8890B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8860B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other
 

Related articles

Effective Use of EI Mass Spectral Databases in Unknown Identifications - Part VI: Creating and Using Retention Indices in NIST Software
Article | Video

Effective Use of EI Mass Spectral Databases in Unknown Identifications - Part VI: Creating and Using Retention Indices in NIST Software

This series of webinars will demonstrate the use of the NIST software suite including the NIST Search, AMDIS, and MS Interpreter for searching EI libraries.
James Little/Mass Spec Interpretation Services
tag
share
more
Leveraging Local Libraries for Positively Parsimonious Peak Tables (James Harynuk, MDCW 2026)
Presentation | Video

Leveraging Local Libraries for Positively Parsimonious Peak Tables (James Harynuk, MDCW 2026)

Discover how custom MS libraries and retention index filtering improve GC×GC–MS peak identification, alignment, and data quality in complex samples.
The Multidimensional Chromatography Workshop
tag
share
more
Processing EI GC-MS Data in Chromatogram Window NIST26
Article | Academy

Processing EI GC-MS Data in Chromatogram Window NIST26

Learn how to process EI GC-MS data in NIST 26 using integrated AMDIS deconvolution, library searching, filtering, spectral review, and Excel reporting for confident compound identification.
James Little/Mass Spec Interpretation Services
tag
share
more
Processing of Accurate Mass GC-MS with NIST26 Chromatogram
Article | Academy

Processing of Accurate Mass GC-MS with NIST26 Chromatogram

Learn how to process accurate-mass GC-MS data in NIST 26 using integrated deconvolution, library searching, MS Interpreter validation, filtering tools, and Excel reporting workflows.
James Little/Mass Spec Interpretation Services
tag
share
more
 

Related content

Is Your Next Sip Safe? A Simple, Rapid Method for Measuring volatile PFAS in Juices.

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, HeadSpace
Manufacturer
Shimadzu
Industries
Food & Agriculture

Thermo Scientific™ Orbitrap Exploris™ GC S Mass Spectrometer

Brochures and specifications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD, GC/MS/MS, GC/Orbitrap, GC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Other

Operational efficiency for your GC and GC/MS labs with the Agilent 8860B gas chromatography system

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8890B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8860B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other
 

Related articles

Effective Use of EI Mass Spectral Databases in Unknown Identifications - Part VI: Creating and Using Retention Indices in NIST Software
Article | Video

Effective Use of EI Mass Spectral Databases in Unknown Identifications - Part VI: Creating and Using Retention Indices in NIST Software

This series of webinars will demonstrate the use of the NIST software suite including the NIST Search, AMDIS, and MS Interpreter for searching EI libraries.
James Little/Mass Spec Interpretation Services
tag
share
more
Leveraging Local Libraries for Positively Parsimonious Peak Tables (James Harynuk, MDCW 2026)
Presentation | Video

Leveraging Local Libraries for Positively Parsimonious Peak Tables (James Harynuk, MDCW 2026)

Discover how custom MS libraries and retention index filtering improve GC×GC–MS peak identification, alignment, and data quality in complex samples.
The Multidimensional Chromatography Workshop
tag
share
more
Processing EI GC-MS Data in Chromatogram Window NIST26
Article | Academy

Processing EI GC-MS Data in Chromatogram Window NIST26

Learn how to process EI GC-MS data in NIST 26 using integrated AMDIS deconvolution, library searching, filtering, spectral review, and Excel reporting for confident compound identification.
James Little/Mass Spec Interpretation Services
tag
share
more
Processing of Accurate Mass GC-MS with NIST26 Chromatogram
Article | Academy

Processing of Accurate Mass GC-MS with NIST26 Chromatogram

Learn how to process accurate-mass GC-MS data in NIST 26 using integrated deconvolution, library searching, MS Interpreter validation, filtering tools, and Excel reporting workflows.
James Little/Mass Spec Interpretation Services
tag
share
more
 

Related content

Is Your Next Sip Safe? A Simple, Rapid Method for Measuring volatile PFAS in Juices.

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, HeadSpace
Manufacturer
Shimadzu
Industries
Food & Agriculture

Thermo Scientific™ Orbitrap Exploris™ GC S Mass Spectrometer

Brochures and specifications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD, GC/MS/MS, GC/Orbitrap, GC/HRMS
Manufacturer
Thermo Fisher Scientific
Industries
Other

Operational efficiency for your GC and GC/MS labs with the Agilent 8860B gas chromatography system

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8890B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other

Agilent 8860B Gas Chromatograph

Brochures and specifications
| 2026 | Agilent Technologies
Instrumentation
GC
Manufacturer
Agilent Technologies
Industries
Other
 

Related articles

Effective Use of EI Mass Spectral Databases in Unknown Identifications - Part VI: Creating and Using Retention Indices in NIST Software
Article | Video

Effective Use of EI Mass Spectral Databases in Unknown Identifications - Part VI: Creating and Using Retention Indices in NIST Software

This series of webinars will demonstrate the use of the NIST software suite including the NIST Search, AMDIS, and MS Interpreter for searching EI libraries.
James Little/Mass Spec Interpretation Services
tag
share
more
Leveraging Local Libraries for Positively Parsimonious Peak Tables (James Harynuk, MDCW 2026)
Presentation | Video

Leveraging Local Libraries for Positively Parsimonious Peak Tables (James Harynuk, MDCW 2026)

Discover how custom MS libraries and retention index filtering improve GC×GC–MS peak identification, alignment, and data quality in complex samples.
The Multidimensional Chromatography Workshop
tag
share
more
Processing EI GC-MS Data in Chromatogram Window NIST26
Article | Academy

Processing EI GC-MS Data in Chromatogram Window NIST26

Learn how to process EI GC-MS data in NIST 26 using integrated AMDIS deconvolution, library searching, filtering, spectral review, and Excel reporting for confident compound identification.
James Little/Mass Spec Interpretation Services
tag
share
more
Processing of Accurate Mass GC-MS with NIST26 Chromatogram
Article | Academy

Processing of Accurate Mass GC-MS with NIST26 Chromatogram

Learn how to process accurate-mass GC-MS data in NIST 26 using integrated deconvolution, library searching, MS Interpreter validation, filtering tools, and Excel reporting workflows.
James Little/Mass Spec Interpretation Services
tag
share
more
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike