Eurachem is a European network promoting international traceability of chemical measurements and good analytical practice. It provides expert guidance, supports method validation, uncertainty evaluation, proficiency testing, and facilitates discussion on technical and policy issues in analytical quality.
Tags
Article
Academy
LinkedIn Logo

Performance Assessment of Binary Output Examinations in Medical Laboratories

Tu, 31.3.2026
| Original article from: Eurachem
Learn how to assess clinical test performance using sensitivity, specificity, and confidence intervals, and how to define fitness-for-purpose criteria for reliable diagnostic decisions.
<p>Eurachem: Performance Assessment of Binary Output Examinations in Medical Laboratories</p>

Eurachem: Performance Assessment of Binary Output Examinations in Medical Laboratories

1 - What is fitness for purpose? 

The test meets the necessary performance criteria for a clinical binary examination to ensure it is suitable for making accurate and reliable clinical decisions in the intended patient population. Among others, performance criteria include clinical sensitivity and clinical specificity. 

2 - How to assess clinical performance? 

Assessing the clinical performance of a binary examination, stating results as 'positive' or 'negative,' involves calculating the examination's Clinical Sensitivity and Clinical Specificity. From the physician's point of view, clinical performance is evaluated by using predictive values, i.e., how likely is it that a positive or negative result corresponds to the presence or absence of the condition, which is also influenced by the prevalence of the condition in the population being tested. 

3 - What is Clinical Sensitivity? 

Clinical sensitivity (SS), also known as the True Positive Rate, measures the proportion of true positives correctly identified by the test. Data from a group of subjects who have been confirmed to have the disease or condition through diagnosis or a gold standard test (D1 in Table 1) are needed to calculate clinical sensitivity.

Eurachem: Table 1. 2x2 Contingency table for clinical diagnosis.Eurachem: Table 1. 2x2 Contingency table for clinical diagnosis.

The formula for clinical sensitivity is: 

SS = tp / (tp + fn), with: 

  • tp, True Positives are the cases where the test correctly identifies the presence of the disease. 
  • fn, False Negatives are the cases where the test incorrectly identifies the absence of the disease when it is actually present. 

Table 1 represents the components of clinical sensitivity and specificity. Figure 1 illustrates an example of the overlapping of the number of components. The symbol "∩" represents the intersection of groups. For example, D1∩n denotes the set of negative results obtained for known positive cases, i.e., false negatives.

4 - What is Clinical Specificity? 

Clinical Specificity (SP), also known as the True Negative Rate, measures the proportion of true negatives correctly identified by the test. Data from a group of subjects who have been confirmed not to have the disease or condition through a gold standard test (D0 in Table 1) are needed to calculate clinical specificity.

The formula for clinical specificity is: 

SP = tn / (tn + fp), where: 

  • tn, True Negatives are the cases where the test correctly identifies the absence of the disease. 
  • fp, False Positives are the cases where the test incorrectly identifies the presence of the disease when it is actually absent. 

Eurachem: Figure 1. Graphical representation of an example of tp (D1∩p), tn (D0∩n), fn (D1∩n), and fp (D0∩p).Eurachem: Figure 1. Graphical representation of an example of tp (D1∩p), tn (D0∩n), fn (D1∩n), and fp (D0∩p).

5 - How to calculate the uncertainty of clinical sensitivity and specificity?

Calculating clinical sensitivity and specificity uncertainty involves statistical methods that estimate each parameter's confidence interval (CI). These intervals provide a range of values within which the true sensitivity or specificity is expected to lie with a certain level of confidence, typically 95 %. Please refer to the AQA:2021 for information about statistical models to compute the 95 % CI [1].

6 - How to define targets? 

Target minimum values for clinical sensitivity and specificity should be identified. These values may originate from regulations or criteria based on the intended use of the binary results. A target or minimum value for the lower limit of the 95 % CI for result rate, i.e., SS or SP, (LLSS.95 or LLSP.95) or the upper limit of the 95 % CI for result rate (HLSS.95 or HLSS.95) should also be defined according to the purpose of the analysis. The target or minimum value (LLSStg or LLSPtg) is particularly significant when the impact of false results is critical. For example, for blood components used in transfusions, the screening for infectious diseases should be performed with tests associated with an LLSS.95 close to 100 %. 

Figure 2 illustrates five different tests, each assessed for clinical sensitivity and its associated uncertainty. These cases reflect different levels of compliance with a target sensitivity of 0.5. Case 1 is non-compliant, Cases 2 to 4 are partially compliant, and Case 5 is compliant. From a simplified statistical power view, the laboratory should always take into account the minimum number of samples when defining its target. This number can be estimated by simulation. For example, for a clinical sensitivity of 90 % the evaluation will require a minimum of 10 samples. It should be emphasized that samples must be representative of the target population for a clinical power evaluation - for example, epidemiologically representative samples of the types and subtypes of a given virological agent are required.

Eurachem: Figure 2. Compliance assessment with clinical sensitivity uncertainty using the lower 95% confidence interval limit for five performance assessment scenarios.Eurachem: Figure 2. Compliance assessment with clinical sensitivity uncertainty using the lower 95% confidence interval limit for five performance assessment scenarios.

Resources

[1] R. Bettencourt da Silva and S. L. R. Ellison (eds.) Eurachem/CITAC Guide: Assessment of performance and uncertainty in qualitative chemical analysis. (1st ed. 2021). 

Eurachem
LinkedIn Logo
 

Related content

Run Smarter, Go Greener: Smart Energy Saving In Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Gas Chromatograph Mass Spectrometer GCMS-TQ RX Series

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Shimadzu
Industries
Other

Hydrogen Carrier Gas Method Translation in Comprehensive Two- Dimensional Gas Chromatography for Sustainable Nontargeted Analysis

Scientific articles
| 2026 | William & Mary
Instrumentation
GCxGC, GC/MSD, GC/MS/MS, GC/TOF, GC/HRMS
Manufacturer
LECO
Industries
Other

Novel data processing workflow including in-silico derivatization for integrated CI & EI MS data acquired with the GC-ecTOFinstrument

Posters
| 2026 | Bruker (ASMS)
Instrumentation
GC/MS/MS, GC/MSD, GC/TOF, GC/HRMS
Manufacturer
Bruker
Industries
Metabolomics

Is Your Next Sip Safe? A Simple, Rapid Method for Measuring volatile PFAS in Juices.

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, HeadSpace
Manufacturer
Shimadzu
Industries
Food & Agriculture
 

Related articles

Exploring PFAS in consumer goods using GCxGC-high-resolution mass spectrometry (David Alonso, MDCW 2026)
Presentation | Video

Exploring PFAS in consumer goods using GCxGC-high-resolution mass spectrometry (David Alonso, MDCW 2026)

GC×GC-HRTOFMS enables fast and reliable screening of PFAS and emerging contaminants in complex consumer goods, combining EI/PCI/NCI and mass defect workflows for confident identification.
The Multidimensional Chromatography Workshop
tag
share
more
On-Site Microfluidic Assay for the Detection of Multiple Drugs in Urine
Scientific article | Science and research

On-Site Microfluidic Assay for the Detection of Multiple Drugs in Urine

This study presents a fluorescence-based microfluidic assay for rapid, simultaneous detection of multiple drugs in urine with high sensitivity and specificity.
LabRulez
tag
share
more
A Rapid, Antigen-Independent Diagnostic Strategy for Chronic Chagas Disease Based on Serum ATR-FTIR Spectroscopy and Machine Learning
Scientific article | Science and research

A Rapid, Antigen-Independent Diagnostic Strategy for Chronic Chagas Disease Based on Serum ATR-FTIR Spectroscopy and Machine Learning

This study combines ATR-FTIR serum spectroscopy and machine learning to accurately diagnose chronic Chagas disease without antigen-based assays.
LabRulez
tag
share
more
Use of uncertainty in compliance
Article | Academy

Use of uncertainty in compliance

In this leaflet we present the Eurachem/CITAC guide on how to assess compliance with a specification or a regulation.
Eurachem
tag
share
more
 

Related content

Run Smarter, Go Greener: Smart Energy Saving In Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Gas Chromatograph Mass Spectrometer GCMS-TQ RX Series

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Shimadzu
Industries
Other

Hydrogen Carrier Gas Method Translation in Comprehensive Two- Dimensional Gas Chromatography for Sustainable Nontargeted Analysis

Scientific articles
| 2026 | William & Mary
Instrumentation
GCxGC, GC/MSD, GC/MS/MS, GC/TOF, GC/HRMS
Manufacturer
LECO
Industries
Other

Novel data processing workflow including in-silico derivatization for integrated CI & EI MS data acquired with the GC-ecTOFinstrument

Posters
| 2026 | Bruker (ASMS)
Instrumentation
GC/MS/MS, GC/MSD, GC/TOF, GC/HRMS
Manufacturer
Bruker
Industries
Metabolomics

Is Your Next Sip Safe? A Simple, Rapid Method for Measuring volatile PFAS in Juices.

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, HeadSpace
Manufacturer
Shimadzu
Industries
Food & Agriculture
 

Related articles

Exploring PFAS in consumer goods using GCxGC-high-resolution mass spectrometry (David Alonso, MDCW 2026)
Presentation | Video

Exploring PFAS in consumer goods using GCxGC-high-resolution mass spectrometry (David Alonso, MDCW 2026)

GC×GC-HRTOFMS enables fast and reliable screening of PFAS and emerging contaminants in complex consumer goods, combining EI/PCI/NCI and mass defect workflows for confident identification.
The Multidimensional Chromatography Workshop
tag
share
more
On-Site Microfluidic Assay for the Detection of Multiple Drugs in Urine
Scientific article | Science and research

On-Site Microfluidic Assay for the Detection of Multiple Drugs in Urine

This study presents a fluorescence-based microfluidic assay for rapid, simultaneous detection of multiple drugs in urine with high sensitivity and specificity.
LabRulez
tag
share
more
A Rapid, Antigen-Independent Diagnostic Strategy for Chronic Chagas Disease Based on Serum ATR-FTIR Spectroscopy and Machine Learning
Scientific article | Science and research

A Rapid, Antigen-Independent Diagnostic Strategy for Chronic Chagas Disease Based on Serum ATR-FTIR Spectroscopy and Machine Learning

This study combines ATR-FTIR serum spectroscopy and machine learning to accurately diagnose chronic Chagas disease without antigen-based assays.
LabRulez
tag
share
more
Use of uncertainty in compliance
Article | Academy

Use of uncertainty in compliance

In this leaflet we present the Eurachem/CITAC guide on how to assess compliance with a specification or a regulation.
Eurachem
tag
share
more
 

Related content

Run Smarter, Go Greener: Smart Energy Saving In Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Gas Chromatograph Mass Spectrometer GCMS-TQ RX Series

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Shimadzu
Industries
Other

Hydrogen Carrier Gas Method Translation in Comprehensive Two- Dimensional Gas Chromatography for Sustainable Nontargeted Analysis

Scientific articles
| 2026 | William & Mary
Instrumentation
GCxGC, GC/MSD, GC/MS/MS, GC/TOF, GC/HRMS
Manufacturer
LECO
Industries
Other

Novel data processing workflow including in-silico derivatization for integrated CI & EI MS data acquired with the GC-ecTOFinstrument

Posters
| 2026 | Bruker (ASMS)
Instrumentation
GC/MS/MS, GC/MSD, GC/TOF, GC/HRMS
Manufacturer
Bruker
Industries
Metabolomics

Is Your Next Sip Safe? A Simple, Rapid Method for Measuring volatile PFAS in Juices.

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, HeadSpace
Manufacturer
Shimadzu
Industries
Food & Agriculture
 

Related articles

Exploring PFAS in consumer goods using GCxGC-high-resolution mass spectrometry (David Alonso, MDCW 2026)
Presentation | Video

Exploring PFAS in consumer goods using GCxGC-high-resolution mass spectrometry (David Alonso, MDCW 2026)

GC×GC-HRTOFMS enables fast and reliable screening of PFAS and emerging contaminants in complex consumer goods, combining EI/PCI/NCI and mass defect workflows for confident identification.
The Multidimensional Chromatography Workshop
tag
share
more
On-Site Microfluidic Assay for the Detection of Multiple Drugs in Urine
Scientific article | Science and research

On-Site Microfluidic Assay for the Detection of Multiple Drugs in Urine

This study presents a fluorescence-based microfluidic assay for rapid, simultaneous detection of multiple drugs in urine with high sensitivity and specificity.
LabRulez
tag
share
more
A Rapid, Antigen-Independent Diagnostic Strategy for Chronic Chagas Disease Based on Serum ATR-FTIR Spectroscopy and Machine Learning
Scientific article | Science and research

A Rapid, Antigen-Independent Diagnostic Strategy for Chronic Chagas Disease Based on Serum ATR-FTIR Spectroscopy and Machine Learning

This study combines ATR-FTIR serum spectroscopy and machine learning to accurately diagnose chronic Chagas disease without antigen-based assays.
LabRulez
tag
share
more
Use of uncertainty in compliance
Article | Academy

Use of uncertainty in compliance

In this leaflet we present the Eurachem/CITAC guide on how to assess compliance with a specification or a regulation.
Eurachem
tag
share
more
 

Related content

Run Smarter, Go Greener: Smart Energy Saving In Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Gas Chromatograph Mass Spectrometer GCMS-TQ RX Series

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Shimadzu
Industries
Other

Hydrogen Carrier Gas Method Translation in Comprehensive Two- Dimensional Gas Chromatography for Sustainable Nontargeted Analysis

Scientific articles
| 2026 | William & Mary
Instrumentation
GCxGC, GC/MSD, GC/MS/MS, GC/TOF, GC/HRMS
Manufacturer
LECO
Industries
Other

Novel data processing workflow including in-silico derivatization for integrated CI & EI MS data acquired with the GC-ecTOFinstrument

Posters
| 2026 | Bruker (ASMS)
Instrumentation
GC/MS/MS, GC/MSD, GC/TOF, GC/HRMS
Manufacturer
Bruker
Industries
Metabolomics

Is Your Next Sip Safe? A Simple, Rapid Method for Measuring volatile PFAS in Juices.

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, HeadSpace
Manufacturer
Shimadzu
Industries
Food & Agriculture
 

Related articles

Exploring PFAS in consumer goods using GCxGC-high-resolution mass spectrometry (David Alonso, MDCW 2026)
Presentation | Video

Exploring PFAS in consumer goods using GCxGC-high-resolution mass spectrometry (David Alonso, MDCW 2026)

GC×GC-HRTOFMS enables fast and reliable screening of PFAS and emerging contaminants in complex consumer goods, combining EI/PCI/NCI and mass defect workflows for confident identification.
The Multidimensional Chromatography Workshop
tag
share
more
On-Site Microfluidic Assay for the Detection of Multiple Drugs in Urine
Scientific article | Science and research

On-Site Microfluidic Assay for the Detection of Multiple Drugs in Urine

This study presents a fluorescence-based microfluidic assay for rapid, simultaneous detection of multiple drugs in urine with high sensitivity and specificity.
LabRulez
tag
share
more
A Rapid, Antigen-Independent Diagnostic Strategy for Chronic Chagas Disease Based on Serum ATR-FTIR Spectroscopy and Machine Learning
Scientific article | Science and research

A Rapid, Antigen-Independent Diagnostic Strategy for Chronic Chagas Disease Based on Serum ATR-FTIR Spectroscopy and Machine Learning

This study combines ATR-FTIR serum spectroscopy and machine learning to accurately diagnose chronic Chagas disease without antigen-based assays.
LabRulez
tag
share
more
Use of uncertainty in compliance
Article | Academy

Use of uncertainty in compliance

In this leaflet we present the Eurachem/CITAC guide on how to assess compliance with a specification or a regulation.
Eurachem
tag
share
more
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike