How much sample can I load onto my SPE device? Is it linked to the concentration of my analyte of interest?

We, 2.7.2025
| Original article from: Phenomenex
How much sample can I load onto my SPE device? Is it linked to the concentration of my analyte of interest?
<p><strong>Phenomenex: </strong>How much sample can I load onto my SPE device? Is it linked to the concentration of my analyte of interest?</p>

Phenomenex: How much sample can I load onto my SPE device? Is it linked to the concentration of my analyte of interest?

When considering the mass of sorbent required to perform a solid phase extraction it is critical to consider the sample matrix (serum, plasma, urine, saliva, food, river water, soil etc). This information will allow you, with the use of a suitable look up table to determine the mass of SPE sorbent to use. The tables are produced to reflect the relative levels of interfering endogenous material present in the samples. 

The SPE device must have enough capacity to retain all analytes of interest together with any endogenous materials capable of retaining on the sorbent with the same type of interactions as the analyte. If there is insufficient sorbent, it is possible some analyte may pass unretained through the SPE device during sample loading, leading to lower recoveries and a loss of consistency of recovery between samples.

Sample Volume

Determines Bed Mass/ Solvent Volumes

Phenomenex: How much sample can I load onto my SPE device? Is it linked to the concentration of my analyte of interest?Phenomenex: How much sample can I load onto my SPE device? Is it linked to the concentration of my analyte of interest?

Generic Loading Capacities:

Polymeric: 10 - 15% of Total Bed Mass | Silica: 5% of Total Bed Mass

If we compare plasma and urine, we can load 500 µl of plasma onto a 60 mg polymeric sorbent such as Strata™-X. Using the same 60 mg SPE tube 2 mL of urine could be loaded, indicating that urine contains significantly less endogenous material than plasma.

For sample types not included in the table, the use of generic loading capacity, 5% of sorbent mass for silica materials, or 10-15% for polymeric materials are advised. The difference between silica and polymeric sorbent loading capacity is a function of available surface area, which is higher for polymeric materials and also ligand density – the entire surface of a polymeric material is essentially stationary phase, while on a silica material only the phase bonded to the silica takes part in the extraction mechanism for reversed phase or ion exchange extractions.

Phenomenex
LinkedIn Logo
 

Related content

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Quantitative Volatile PFAS Analysis in Textiles

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Agilent Technologies
Industries
Materials Testing

Effectiveness of the MonoTrap Collection Method for VOC Analysis in Exhaled Breath Using GC-MS

Applications
| 2026 | Shimadzu
Instrumentation
GC/MSD, Thermal desorption, GC/SQ
Manufacturer
Shimadzu
Industries
Clinical Research
 

Related articles

Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery
Article | Academy

Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery

Over-drying silica-based SPE cartridges during conditioning can reduce analyte retention and recovery. Learn best practices to avoid losses and ensure reproducible results.
Phenomenex
tag
share
more
All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?
Article | Academy

All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?

How can neutral analytes be efficiently concentrated in the presence of ionized acidic or basic contaminants? This SPE tip explains how mixed-mode cartridges improve selectivity and cleanup.
Phenomenex
tag
share
more
Troubleshooting SPE
Article | Academy

Troubleshooting SPE

A technical guide to troubleshooting low analyte recovery in SPE. Explains relative vs. absolute recovery and provides practical solutions for conditioning, loading, and sorbent selection.
Phenomenex
tag
share
more
 

Related content

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Quantitative Volatile PFAS Analysis in Textiles

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Agilent Technologies
Industries
Materials Testing

Effectiveness of the MonoTrap Collection Method for VOC Analysis in Exhaled Breath Using GC-MS

Applications
| 2026 | Shimadzu
Instrumentation
GC/MSD, Thermal desorption, GC/SQ
Manufacturer
Shimadzu
Industries
Clinical Research
 

Related articles

Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery
Article | Academy

Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery

Over-drying silica-based SPE cartridges during conditioning can reduce analyte retention and recovery. Learn best practices to avoid losses and ensure reproducible results.
Phenomenex
tag
share
more
All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?
Article | Academy

All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?

How can neutral analytes be efficiently concentrated in the presence of ionized acidic or basic contaminants? This SPE tip explains how mixed-mode cartridges improve selectivity and cleanup.
Phenomenex
tag
share
more
Troubleshooting SPE
Article | Academy

Troubleshooting SPE

A technical guide to troubleshooting low analyte recovery in SPE. Explains relative vs. absolute recovery and provides practical solutions for conditioning, loading, and sorbent selection.
Phenomenex
tag
share
more
 

Related content

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Quantitative Volatile PFAS Analysis in Textiles

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Agilent Technologies
Industries
Materials Testing

Effectiveness of the MonoTrap Collection Method for VOC Analysis in Exhaled Breath Using GC-MS

Applications
| 2026 | Shimadzu
Instrumentation
GC/MSD, Thermal desorption, GC/SQ
Manufacturer
Shimadzu
Industries
Clinical Research
 

Related articles

Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery
Article | Academy

Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery

Over-drying silica-based SPE cartridges during conditioning can reduce analyte retention and recovery. Learn best practices to avoid losses and ensure reproducible results.
Phenomenex
tag
share
more
All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?
Article | Academy

All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?

How can neutral analytes be efficiently concentrated in the presence of ionized acidic or basic contaminants? This SPE tip explains how mixed-mode cartridges improve selectivity and cleanup.
Phenomenex
tag
share
more
Troubleshooting SPE
Article | Academy

Troubleshooting SPE

A technical guide to troubleshooting low analyte recovery in SPE. Explains relative vs. absolute recovery and provides practical solutions for conditioning, loading, and sorbent selection.
Phenomenex
tag
share
more
 

Related content

High-Throughput BTEX Analysis in Nail Products by SPME and GC/TQ

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ, SPME
Manufacturer
Agilent Technologies
Industries
Materials Testing

Accurate multi-component blast furnace gas analysis maximizes iron production and minimizes coke consumption

Applications
| 2026 | Thermo Fisher Scientific
Instrumentation
GC/MSD
Manufacturer
Thermo Fisher Scientific
Industries
Energy & Chemicals

Gas Chromatograph Nexis GC-2060

Brochures and specifications
| 2026 | Shimadzu
Instrumentation
GC
Manufacturer
Shimadzu
Industries
Other

Quantitative Volatile PFAS Analysis in Textiles

Applications
| 2026 | Agilent Technologies
Instrumentation
GC/MSD, GC/MS/MS, GC/QQQ
Manufacturer
Agilent Technologies
Industries
Materials Testing

Effectiveness of the MonoTrap Collection Method for VOC Analysis in Exhaled Breath Using GC-MS

Applications
| 2026 | Shimadzu
Instrumentation
GC/MSD, Thermal desorption, GC/SQ
Manufacturer
Shimadzu
Industries
Clinical Research
 

Related articles

Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery
Article | Academy

Over-Drying of Silica-Based SPE Cartridges Can Lead to Poor Sample Recovery

Over-drying silica-based SPE cartridges during conditioning can reduce analyte retention and recovery. Learn best practices to avoid losses and ensure reproducible results.
Phenomenex
tag
share
more
All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?
Article | Academy

All My Analytes are Neutrals, but Contaminants are Acids or Amines. What is the Best Way to Concentrate and Clean Up my Analytes of Interest?

How can neutral analytes be efficiently concentrated in the presence of ionized acidic or basic contaminants? This SPE tip explains how mixed-mode cartridges improve selectivity and cleanup.
Phenomenex
tag
share
more
Troubleshooting SPE
Article | Academy

Troubleshooting SPE

A technical guide to troubleshooting low analyte recovery in SPE. Explains relative vs. absolute recovery and provides practical solutions for conditioning, loading, and sorbent selection.
Phenomenex
tag
share
more
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike