JetClean Self cleaning ion source
Presentations | 2016 | Agilent TechnologiesInstrumentation
The reliability and uptime of gas chromatography–mass spectrometry (GC/MS) systems are often limited by the need for regular manual cleaning of the ion source. Frequent source maintenance reduces productivity, increases downtime and requires experienced staff. The JetClean self-cleaning ion source offers an automated alternative that can dramatically extend maintenance intervals and improve laboratory efficiency.
This application note introduces the Agilent JetClean self-cleaning quadrupole ion source for single and triple quadrupole GC/MS instruments. It describes the operating principle, software integration, and performance validation in various analytical contexts, including environmental analysis, food safety, material science and forensic applications.
The JetClean system introduces controlled hydrogen pulses at low flow rates into the ion source via a mass flow controller (MFC). Cleaning cycles can run in two modes:
Key components and software:
Cleaning efficacy was demonstrated by treating a rhodamine-contaminated lens: hydrogen-based JetClean removed heavy residues that helium bake-out alone could not. In a validated PAH study (62 compounds, 1–10,000 pg/µL), calibration linearity remained R² > 0.998 and limits of detection averaged 1.02 ± 0.82 pg/µL over two years without manual cleaning. A pyrolyzer laboratory reported stable isotope ratio measurements with only biweekly verification. In Japanese laboratories, standard systems required monthly manual cleaning, whereas JetClean-equipped systems operated 23 months without source disassembly. Additional demonstration studies included pesticide residue analysis in fruits, vegetables, cereals and honey (1–250 ppb, ~160 analytes), detection of sub-pg levels of PH₃, H₂S, AsH₃ and COS in olefin gases, and consistent quantitation of 16 phthalate esters.
Future developments may integrate predictive maintenance algorithms using instrument log data to optimize cleaning schedules. Expansion of self-cleaning chemistries beyond hydrogen, remote monitoring through cloud-based platforms and further integration with high-throughput workflows are anticipated. Adoption in regulated laboratories will accelerate as robustness data accumulate, and coupling with emerging MS technologies may extend maintenance-free operation to high-resolution systems.
The Agilent JetClean self-cleaning ion source transforms GC/MS maintenance by automating source cleaning with controlled hydrogen pulses. It delivers consistent analytical performance over months to years, cuts manual cleaning by over 80 %, and enhances productivity across diverse applications. Its simplicity, upgradeability and proven reliability position it as a valuable solution for modern analytical laboratories.
GC/MSD
IndustriesManufacturerAgilent Technologies
Summary
Importance of the Topic
The reliability and uptime of gas chromatography–mass spectrometry (GC/MS) systems are often limited by the need for regular manual cleaning of the ion source. Frequent source maintenance reduces productivity, increases downtime and requires experienced staff. The JetClean self-cleaning ion source offers an automated alternative that can dramatically extend maintenance intervals and improve laboratory efficiency.
Objectives and Study Overview
This application note introduces the Agilent JetClean self-cleaning quadrupole ion source for single and triple quadrupole GC/MS instruments. It describes the operating principle, software integration, and performance validation in various analytical contexts, including environmental analysis, food safety, material science and forensic applications.
Methodology and Instrumentation
The JetClean system introduces controlled hydrogen pulses at low flow rates into the ion source via a mass flow controller (MFC). Cleaning cycles can run in two modes:
- Clean Only (batch mode): periodic cleaning with hydrogen between analyses.
- Acquire and Clean (concurrent mode): alternating analysis with helium and cleaning with hydrogen during a single sequence.
Key components and software:
- GC/MS system with JetClean ion source option or upgrade.
- Mass flow controller for precise H₂ and He flow regulation.
- MassHunter software for easy setup, method storage and logging of cleaning events.
- Optional pyrolyzer interface for polymer and material analysis.
Main Results and Discussion
Cleaning efficacy was demonstrated by treating a rhodamine-contaminated lens: hydrogen-based JetClean removed heavy residues that helium bake-out alone could not. In a validated PAH study (62 compounds, 1–10,000 pg/µL), calibration linearity remained R² > 0.998 and limits of detection averaged 1.02 ± 0.82 pg/µL over two years without manual cleaning. A pyrolyzer laboratory reported stable isotope ratio measurements with only biweekly verification. In Japanese laboratories, standard systems required monthly manual cleaning, whereas JetClean-equipped systems operated 23 months without source disassembly. Additional demonstration studies included pesticide residue analysis in fruits, vegetables, cereals and honey (1–250 ppb, ~160 analytes), detection of sub-pg levels of PH₃, H₂S, AsH₃ and COS in olefin gases, and consistent quantitation of 16 phthalate esters.
Benefits and Practical Applications
- Substantial reduction (>80 %) in manual cleaning frequency, boosting instrument uptime.
- Consistent, long-term response without drift or carryover, enhancing data quality.
- Lower labor requirements—less experienced operators can maintain the system.
- Simple software-driven setup and method portability between instruments.
- Broad applicability across environmental monitoring, food safety, materials characterization and forensic analysis.
- Field upgradeable on existing Agilent single- and triple-quad GC/MS platforms.
Future Trends and Opportunities
Future developments may integrate predictive maintenance algorithms using instrument log data to optimize cleaning schedules. Expansion of self-cleaning chemistries beyond hydrogen, remote monitoring through cloud-based platforms and further integration with high-throughput workflows are anticipated. Adoption in regulated laboratories will accelerate as robustness data accumulate, and coupling with emerging MS technologies may extend maintenance-free operation to high-resolution systems.
Conclusion
The Agilent JetClean self-cleaning ion source transforms GC/MS maintenance by automating source cleaning with controlled hydrogen pulses. It delivers consistent analytical performance over months to years, cuts manual cleaning by over 80 %, and enhances productivity across diverse applications. Its simplicity, upgradeability and proven reliability position it as a valuable solution for modern analytical laboratories.
References
- Elizabeth Almasi. JetClean Self-cleaning Ion Source Application Note. Agilent Technologies; June 2016.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Agilent JetClean self-cleaning ion source
2019|Agilent Technologies|Brochures and specifications
Less Cleaning, More GC/MS Results Agilent JetClean self-cleaning ion source Keep Your Ion Source Clean— Automatically Matrix deposits inevitably build up during routine analysis. In the past, you would have to remove the ion source, scrub the lens and other…
Key words
jetclean, jetcleancleaning, cleaningcleanings, cleaningsper, persource, sourceyearly, yearlyrevenue, revenueyear, yearself, selfwage, wagesavings, savingsbenzo, benzowages, wagesion, ionoctober
Maintaining Sensitivity and Reproducibility with the Agilent JetClean Self-Cleaning Ion Source for Pesticides in Food and Feed
2016|Agilent Technologies|Applications
Maintaining Sensitivity and Reproducibility with the Agilent JetClean Self-Cleaning Ion Source for Pesticides in Food and Feed Application Note Author Abstract Jessica Westland Approximately 200 various pesticides were analyzed in organic honey extract on Agilent Technologies, Inc. the Agilent 7010A…
Key words
jetclean, jetcleanagilent, agilenthoney, honeymin, mintime, timewithout, withoutcleaning, cleaningself, selfmasshunter, masshunterendrin, endrinpuu, puusource, sourcetransitions, transitionsuse, usepesticide
Increasing Productivity in Pesticide Residue Analysis Using JetClean with GC/MS and StreamSelect with LC/MS Systems
2019|Agilent Technologies|Posters
Increasing Productivity in Pesticide Residue Analysis Using JetClean with GC/MS and StreamSelect with LC/MS Systems Introduction Today’s high throughput food laboratories require reliable and timely pesticide residue analysis of varied food samples. To gain the highest productivity of the analytical…
Key words
jetclean, jetcleanstreamselect, streamselectprotectants, protectantssource, sourcebinary, binarycleaning, cleaningtimeline, timelinesystems, systemsquaternary, quaternarymanual, manuallens, lensdeposits, depositsspinach, spinachlaprw, laprwacquire
Maintaining Sensitivity and Reproducibility with the JetCleanSelf-Cleaning Ion Source for Pesticides in Food and Feed
2016|Agilent Technologies|Posters
Maintaining Sensitivity and Reproducibility with the JetClean Self-Cleaning Ion Source for Pesticides in Food and Feed Jessica Westland1; Vivian Chen2 and Elizabeth Almasi3; 1Agilent Technologies, Wilmington, DE 19808, USA 2Agilent Technologies, Shanghai Company Limited; 200131,CHINA; 3Agilent ASMS 2016 TP 216…
Key words
jetclean, jetcleansource, sourceendosulfan, endosulfanepoxide, epoxidemethyl, methylheptachlor, heptachlorendrin, endrinstd, stdacquire, acquireirgarol, irgaroltune, tunevivian, viviantemperature, temperatureelizabeth, elizabethdbcp