Phenols
Applications | 2011 | Agilent TechnologiesInstrumentation
The analysis of phenolic compounds is critical in environmental monitoring, industrial process control, and quality assurance. Phenols and their methylated derivatives exhibit toxicity and impact water and soil quality. Reliable chromatographic separation enables accurate quantification for regulatory compliance and research applications.
This application note demonstrates the separation of 14 structurally similar phenols using gas chromatography on an Agilent CP-Sil 43 CB column. The goal is to achieve baseline resolution within a 50-minute run time, illustrating the column’s selectivity for alkyl-substituted phenols.
A capillary gas chromatograph was configured with the following parameters:
The method resolved the following phenolic analytes:
The isothermal GC method delivered baseline separation of all 14 phenols in under 50 minutes. The CP-Sil 43 CB stationary phase provided distinct selectivity for methyl, ethyl, and isopropyl substituents on the aromatic ring. Peak shapes were sharp and symmetrical, indicating efficient mass transfer and minimal tailing. Carrier gas flow and split ratio produced reproducible retention times and consistent detector response.
Advances in column technology may further reduce analysis time and enhance peak capacity. Temperature programming or tandem column setups could extend the method to a broader range of phenolics. Coupling with mass spectrometry offers structural confirmation and lower detection limits. Automated sampling systems and high-throughput platforms will increase laboratory efficiency.
This application note confirms that the Agilent CP-Sil 43 CB column, operated under isothermal conditions at 100 °C, effectively separates a comprehensive set of 14 phenols in under 50 minutes. The method exhibits robustness and reproducibility suitable for environmental and industrial laboratories.
GC, GC columns, Consumables
IndustriesMaterials Testing
ManufacturerAgilent Technologies
Summary
Significance of the Topic
The analysis of phenolic compounds is critical in environmental monitoring, industrial process control, and quality assurance. Phenols and their methylated derivatives exhibit toxicity and impact water and soil quality. Reliable chromatographic separation enables accurate quantification for regulatory compliance and research applications.
Objectives and Overview of the Application Note
This application note demonstrates the separation of 14 structurally similar phenols using gas chromatography on an Agilent CP-Sil 43 CB column. The goal is to achieve baseline resolution within a 50-minute run time, illustrating the column’s selectivity for alkyl-substituted phenols.
Methodology and Instrumentation
A capillary gas chromatograph was configured with the following parameters:
- Column: Agilent CP-Sil 43 CB, 0.22 mm i.d. × 25 m, 0.2 µm film thickness (WCOT CP-Sil 43 CB, part no. CP7715)
- Temperature: isothermal at 100 °C
- Carrier gas: helium at 135 kPa (1.35 bar), linear velocity 28 cm/s
- Injection: split mode, 100 mL/min
- Detection: flame ionization detector (FID), sensitivity 4 × 10^–12 Afs
The method resolved the following phenolic analytes:
- Phenol
- 2,6-Xylenol
- o-Cresol
- p-Cresol
- m-Cresol
- o-Ethylphenol
- 2,4-Xylenol
- 2,5-Xylenol
- 2-Isopropylphenol
- 2,3-Xylenol
- 3,5-Xylenol
- p-Ethylphenol
- 3,4-Xylenol
- 2,3,5-Trimethylphenol
Main Results and Discussion
The isothermal GC method delivered baseline separation of all 14 phenols in under 50 minutes. The CP-Sil 43 CB stationary phase provided distinct selectivity for methyl, ethyl, and isopropyl substituents on the aromatic ring. Peak shapes were sharp and symmetrical, indicating efficient mass transfer and minimal tailing. Carrier gas flow and split ratio produced reproducible retention times and consistent detector response.
Benefits and Practical Applications
- Environmental analysis: monitoring phenolic pollutants in water, soil, and effluent streams.
- Industrial quality control: verification of phenolic content in chemical manufacturing, resin production, and polymer synthesis.
- Research applications: profiling methylated phenol isomers for degradation studies and reaction monitoring.
Future Trends and Opportunities
Advances in column technology may further reduce analysis time and enhance peak capacity. Temperature programming or tandem column setups could extend the method to a broader range of phenolics. Coupling with mass spectrometry offers structural confirmation and lower detection limits. Automated sampling systems and high-throughput platforms will increase laboratory efficiency.
Conclusion
This application note confirms that the Agilent CP-Sil 43 CB column, operated under isothermal conditions at 100 °C, effectively separates a comprehensive set of 14 phenols in under 50 minutes. The method exhibits robustness and reproducibility suitable for environmental and industrial laboratories.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Phenols
2011|Agilent Technologies|Applications
Phenols Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC separation of 12 phenols using an Agilent CP-Sil 19 CB column with FID is achieved in 15 minutes. Conditions Technique : GC Column : Agilent CP-Sil 19 CB, 0.32 mm…
Key words
phenols, phenolscourtesy, courtesyprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique, techniquesample, samplecarrier, carrieridentification, identificationsolvent, solventdetector, detectornote
PAHs and phenols
2011|Agilent Technologies|Applications
PAHs and phenols Application Note Environmental Authors Introduction Agilent Technologies, Inc. Gas chromatography with an Agilent CP-Wax 52 CB column separates 19 polyaromatic hydrocarbons and phenols in 22 minutes. Conditions Technique : GC-capillary Column : Agilent CP-Wax 52 CB, 0.25…
Key words
phenols, phenolscourtesy, courtesypahs, pahsprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak
Phenols and cresols - Separation of isomers and aromatic alcohols on a chiral capillary column
2011|Agilent Technologies|Applications
Phenols and cresols Separation of isomers and aromatic alcohols on a chiral capillary column Application Note Environmental Authors Introduction Agilent Technologies, Inc. The Agilent CP-Chirasil-Dex CB column is not only suited for the separation of optical isomers, but has also…
Key words
chirasil, chirasilcresols, cresolsdex, dexphenols, phenolschiral, chiralcolumn, columncourtesy, courtesyalcohols, alcoholsseparation, separationisomers, isomersaromatic, aromaticsuited, suitedprinted, printedinjector, injectorusa
Analysis of Free Volatile Phenols in Smoke-Impacted Wines by SPME
2021|Agilent Technologies|Applications
Application Note Food Testing and Agriculture Analysis of Free Volatile Phenols in Smoke-Impacted Wines by SPME Authors Jessica Westland and Vanessa Abercrombie Agilent Technologies, Inc. Abstract Ever since the 2003 wildfires in Australia and British Columbia, smoke impact has been…
Key words
smoke, smokecresol, cresolwine, wineguaiacol, guaiacolgrapes, grapesspme, spmephenols, phenolsnoir, noirpinot, pinotfranzia, franziaimpact, impactloq, loqvolatile, volatilefree, freetaint