Agilent 7200 Series Q-TOF GC/MS System - Data Sheet

Brochures and specifications | 2012 | Agilent TechnologiesInstrumentation
GC/MSD, GC/MS/MS, GC/HRMS, GC/Q-TOF
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the Topic

The Agilent 7200 Series Q-TOF GC/MS system offers high resolution and accurate mass measurements for trace-level identification and quantification of compounds in complex matrices. Its combination of quadrupole filtering, time-of-flight detection, and advanced acquisition software addresses critical demands in environmental analysis, food safety, forensics, and pharmaceutical research.

Objectives and Study Overview

This data sheet details the performance characteristics, operating modes, and practical capabilities of the Agilent 7200 Series Q-TOF GC/MS coupled with the 7890A gas chromatograph. Key goals include demonstrating sensitivity, mass accuracy, resolution, dynamic range, and workflow flexibility for routine and research applications.

Methodology and Instrumentation

This system integrates gas chromatographic separation with quadrupole time-of-flight mass spectrometry under multiple ionization modes. Electron ionization (EI) serves as the standard, with optional positive and negative chemical ionization (PCI, NCI) for enhanced molecular ion information. High-speed acquisition (up to 50 spectra per second) and MS/MS capability enable deconvolution of coeluting peaks and confident structural elucidation.

Used Instrumentation

  • Agilent 7200 Series Q-TOF GC/MS with proprietary hyperbolic gold-coated quadrupole and linear hexapole collision cell
  • Ion sources: non-coated inert EI source (dual filaments) plus optional PCI/NCI
  • Time-of-flight analyzer: two-stage corrected reflectron, 2 m flight path, microchannel plate detector
  • Agilent 7890A gas chromatograph with split/splitless, multimode inlet, PTV and headspace options
  • Autosamplers: 7693, 7683 ALS, CombiPAL, 7697A headspace
  • Software: Agilent MassHunter Acquisition, Qualitative, Quantitative and reporting modules


Main Results and Discussion

Performance tests demonstrate EI detection limits below 250 fg (OFN, m/z 271.9867) at 99% confidence and full-spectrum signal-to-noise ratios above 2000:1. TOF resolution exceeds 12 500 (FWHM) with mass accuracy better than 5 ppm. PCI detection limits approach 24 pg (Benzophenone, m/z 183.0804) with SNR above 1500:1. Quadrupole axis stability remains within ±0.10 Da over 24 hours, and dynamic range exceeds 105. Fast acquisition rates support quantification of narrow GC peaks and effective library matching.

Benefits and Practical Applications

  • Trace-level screening of environmental pollutants and pesticides
  • Forensic analysis of complex drug matrices with high selectivity
  • Food and flavor profiling using accurate mass fragmentation
  • Pharmaceutical impurity identification and QA/QC workflows
  • Non-targeted metabolomics and biomarker discovery


Future Trends and Applications

Advances are expected in automated data processing with machine learning, integration of high-throughput sampling technologies, miniaturized ion sources, and enhanced multi-omic workflows. Combining Q-TOF GC/MS with ambient ionization and real-time monitoring will expand applications in industrial process control and rapid field screening.

Conclusion

The Agilent 7200 Series Q-TOF GC/MS system delivers exceptional sensitivity, resolution, and mass accuracy for demanding analytical challenges. Its flexible ionization modes, fast acquisition rates, and robust performance support a wide range of routine and research applications, enabling confident identification of known and unknown compounds.

References

  • Agilent Technologies. Agilent 7200 Series Q-TOF GC/MS System Data Sheet. Publication 5991-0089EN, August 2012.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Agilent 7200B Series GC/Q-TOF System - Data Sheet
Agilent 7200B Series GC/Q-TOF System - Data Sheet
2014|Agilent Technologies|Brochures and specifications
Agilent 7200B Series GC/Q-TOF System Data Sheet The Agilent 7200B Series Q-TOF GC/MS System with Agilent MassHunter Software offers outstanding sensitivity, selectivity, and mass spectral information. High resolution and accurate mass allows identification of unknown compounds in the most complex…
Key words
mass, masstof, tofturbomolecular, turbomolecularpci, pcirough, roughquadrupole, quadrupolepneumatic, pneumaticion, ionsource, sourcecollision, collisioninseparable, inseparableresolution, resolutionpump, pumpfull, fullisolation
Agilent 7250 GC/Q-TOF - Data Sheet
Agilent 7250 GC/Q-TOF - Data Sheet
2017|Agilent Technologies|Brochures and specifications
Agilent 7250 GC/Q-TOF Data Sheet The Agilent 7250 GC/Q-TOF with Agilent MassHunter Software offers outstanding sensitivity, selectivity, and mass spectral information. Low energy Electron Ionization (EI) provides accessible molecular ion information for compound identification and structure elucidation workflows. High resolution…
Key words
tof, tofmass, massquadrupole, quadrupolespecification, specificationindependent, independentpneumatic, pneumaticion, ionoctofluoronapthalene, octofluoronapthalenecollision, collisionrecordable, recordableparameter, parameterenergy, energyrate, ratefwhh, fwhhscintillator
Agilent 7000B Triple Quadrupole GC/MS System
Agilent 7000B Triple Quadrupole GC/MS System
2012|Agilent Technologies|Brochures and specifications
Agilent 7000B Triple Quadrupole GC/MS System Data Sheet Agilent is leading the way by introducing IDL, Instrument Detection Limit, as a new metric for meaningful indication of system sensitivity. In MS/MS baselines with very little noise, the dubious selection of…
Key words
mrm, mrmpci, pciidl, idlfilaments, filamentsnci, ncielectronic, electronicmodes, modespneumatic, pneumaticmore, morecollision, collisiontriple, triplesheet, sheetaxis, axisquadrupole, quadrupolecustomary
Agilent 7000B Triple Quadrupole GC/MS System - Data Sheet
Agilent 7000B Triple Quadrupole GC/MS System - Data Sheet
2011|Agilent Technologies|Brochures and specifications
Agilent 7000B Triple Quadrupole GC/MS System Data Sheet Agilent is leading the way by introducing IDL, Instrument Detection Limit, as a new metric for meaningful indication of system sensitivity. In MS/MS baselines with very little noise, the dubious selection of…
Key words
mrm, mrmpci, pcifilaments, filamentsnci, ncielectronic, electronicmodes, modesidl, idlpneumatic, pneumaticmore, moretriple, triplecollision, collisionsheet, sheetaxis, axisquadrupole, quadrupolecustomary
Other projects
LCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike