Installation and Maintenance Instructions for 0.75mm Wide Bore Capillary Columns

Guides | 1995 | MerckInstrumentation
GC columns, Consumables
Industries
Manufacturer
Merck

Summary

Importance of the Topic


Accurate installation and upkeep of wide-bore capillary columns are critical for achieving consistent separations, minimizing downtime, and prolonging column life in gas chromatography. Proper techniques reduce dead volume, prevent contamination, and ensure reliable quantitative and qualitative analyses across diverse applications.

Objectives and Overview of the Article


This bulletin outlines step-by-step procedures for installing, conditioning, and maintaining 0.75 mm inner-diameter fused silica capillary columns in both packed-column GC systems (using conversion hardware) and dedicated capillary GC instruments. It covers instrument preparation, leak checking, carrier gas handling, column installation, flow optimization, and troubleshooting guidelines.

Methodology


  • Prepare injection port and detector liners by thorough cleaning and silanization to reduce adsorption.
  • Remove oxygen and moisture from carrier gas using purifiers and stainless-steel diaphragm regulators.
  • Purging: flow carrier gas through the column for 30–60 minutes before heating to avoid phase oxidation.
  • Conditioning: program temperature cycles up to application limits, using rates below 25 °C/min to stabilize baseline and prevent thermal shock.
  • Flow setting (packed-column system): adjust carrier gas to ~5 mL/min helium at detector; set makeup gas to achieve total detector flow of 30–60 mL/min.
  • Flow setting (capillary system): sequentially set makeup gas, split-vent flow (≈240–250 mL/min for ~50:1 split), and column flow (≈5 mL/min) using methane retention time as a reference.
  • Test injections: use diluted test mix in packed-column mode and neat mix in capillary mode to verify efficiencies and peak ratios.

Instrumentation Used


  • Gas chromatograph with packed-to-capillary conversion kit or capillary inlet system.
  • Silonized injector and detector liners.
  • Carrier gas purifiers and stainless-steel diaphragm regulators.
  • GOW-MAC thermal-conductivity gas leak detector.
  • Capillary cleaving tools, pin vise drill kit, ferrules, and butt connectors for 0.75 mm tubing.

Main Results and Discussion


By following the recommended cutting, ferrule installation, and tightening steps, users achieve leak-free connections with minimal dead volume, evident in sharp, symmetric peaks. Proper purge and conditioning protocols stabilize baselines rapidly under temperature programming. Regular test mix injections provide early detection of performance drift, attributing issues to inlet contamination, dead volume, or phase bleed.

Benefits and Practical Applications of the Method


  • Enhanced column lifetime through contamination control and oxygen/moisture exclusion.
  • Improved chromatographic resolution by minimizing dead volume and thermal shocks.
  • Consistent test mix performance enables routine quality checks in QA/QC laboratories.
  • Flexibility to adapt wide-bore columns in both packed and capillary GC systems with standardized hardware.

Future Trends and Possibilities of Application


As GC technology advances, integration of automated leak detection, real-time flow monitoring, and inert flow path materials will further simplify wide-bore column handling. Novel column coatings and high-capacity purifiers may extend operating temperature ranges and reduce conditioning time. The methodology can be extended to multidimensional GC and rapid screening protocols in environmental, petrochemical, and forensic analyses.

Conclusion


Adherence to meticulous installation, leak checking, and conditioning procedures ensures optimal performance of 0.75 mm wide-bore capillary columns. Routine maintenance, including test injections and periodic reversal of flow direction, prevents contamination build-up and phase gradients, resulting in reliable separations and prolonged column service life.

Reference


No formal literature references were provided in the original text.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Installation and Maintenance Instructions for 0.53mm ID Fused Silica Capillary Columns
Bulletin 897 Installation and Maintenance Instructions for 0.53mm ID Fused Silica Capillary Columns These instructions cover instrument preparation, column installation, leak checking, gas flow setting procedures, and maintenance requirements for 0.53mm ID capillary columns in packed column (equipped with capillary…
Key words
supelco, supelcocapillary, capillarycolumn, columnconversion, conversionglasseal, glassealyour, yourleak, leakferrule, ferrulenut, nutgow, gowcleaving, cleavingtubing, tubinggas, gasmac, macliners
Installation and Maintenance Instructions for 0.25mm and 0.32mm ID Fused Silica Capillary Columns
Bulletin 895 Installation and Maintenance Instructions for 0.25mm and 0.32mm ID Fused Silica Capillary Columns These instructions cover instrument preparation, column hanging, ferrule and column installation, leak checking, gas flow setting procedures, and maintenance requirements for 0.25mm and 0.32mm ID…
Key words
column, columncapillary, capillarysupelco, supelcochromfax, chromfaxglasseal, glassealfused, fusedgow, gowmac, macsilica, silicaleak, leakgas, gasnonbonded, nonbondedcleaving, cleavingtubing, tubingconnectors
Capillary GC Troubleshooting Guide: How to Locate Problems and Solve Them
Bulletin 853B Capillary GC Troubleshooting Guide: How to Locate Problems and Solve Them The real task in correcting a problem with your capillary GC system is identifying the cause of the problem without wasting time. The systematic approach to troubleshooting…
Key words
supelco, supelcoproblem, problemcolumn, columnsymptom, symptomremedy, remedynormal, normalcause, causepurifier, purifierseptum, septumgas, gastest, testpossible, possiblerecorder, recordercapillary, capillarydetector
The Supelco Guide to Leak-Free Connections: Ferrules and Fittings for Packed and Capillary GC
Bulletin 741D The Supelco Guide to Leak-Free Connections: Ferrules and Fittings for Packed and Capillary GC The ideal GC ferrule provides a leak-tight seal, accommodates column OD variations, seals with minimum torque, and does not stick to the column or…
Key words
ferrule, ferruleferrules, ferrulesswagelok, swagelokcapseal, capsealbullet, bulletglasrench, glasrenchsupeltex, supeltexnpt, nptcolumn, columnsupelco, supelcoleak, leakgraphite, graphiteunion, uniondescription, descriptionglass
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike