Installation and Maintenance Instructions for 0.25mm and 0.32mm ID Fused Silica Capillary Columns
Guides | 1995 | MerckInstrumentation
High-performance gas chromatography relies on precise installation and maintenance of capillary columns to achieve reproducible results and extend column life. Proper handling of 0.25 mm and 0.32 mm ID fused silica columns reduces analysis variability and contamination risk.
This bulletin outlines step-by-step instructions for instrument preparation, column hanging, ferrule fitting, leak checking, gas flow setting, conditioning, and routine maintenance for fused silica capillary systems in gas chromatographs. The goal is to ensure optimal column performance and longevity.
Instrumentation and materials recommended:
Procedures include silanization of injector/detector liners, precise insertion depth marking, leak detection without liquid agents, and carrier gas flow optimization based on average linear velocity measurements.
Key findings and recommendations:
The use of guard columns and selection of appropriate solvents for splitless or on-column injections further protect analytical columns from contaminants.
Implementation of these guidelines leads to:
Emerging directions include automated ferrule fitting systems, advanced carrier gas purification technologies, and novel phases with enhanced thermal stability. Integration of real-time leak detection and flow control into GC software can further streamline column installation and monitoring.
By following systematic installation and maintenance practices, analysts can achieve reliable, high-quality separations with fused silica capillary columns. Regular conditioning, proper handling, and routine performance checks are essential for maximizing column performance and operational efficiency.
Grob K., High-Resolution Capillary and Capillary-Column Chromatography, HRC & CC 3 (1978) 173
GC columns, Consumables
IndustriesManufacturerMerck
Summary
Importance of the Topic
High-performance gas chromatography relies on precise installation and maintenance of capillary columns to achieve reproducible results and extend column life. Proper handling of 0.25 mm and 0.32 mm ID fused silica columns reduces analysis variability and contamination risk.
Objectives and Overview of the Study
This bulletin outlines step-by-step instructions for instrument preparation, column hanging, ferrule fitting, leak checking, gas flow setting, conditioning, and routine maintenance for fused silica capillary systems in gas chromatographs. The goal is to ensure optimal column performance and longevity.
Methodology and Instrumentation Used
Instrumentation and materials recommended:
- Gas chromatograph with injector and detector ports
- Pin Vise Drill Kit or Needle File for ferrule ID adjustments
- Supeltex M-2A and M-4 graphite ferrules
- Capillary Cleaving Tool for cutting column ends
- GOW-MAC Gas Leak Detectors for leak checking
- Carrier gas purifiers with stainless steel diaphragm regulators
Procedures include silanization of injector/detector liners, precise insertion depth marking, leak detection without liquid agents, and carrier gas flow optimization based on average linear velocity measurements.
Main Results and Discussion
Key findings and recommendations:
- Clean, deactivated injector and detector liners prevent adsorption and maintain peak shape.
- Correct ferrule sizing and square cutting of column ends avoid dead volume and jagged edges.
- Use thermal conductivity leak detectors to avoid liquid contamination.
- Set make-up gas, splitter vent flow, and column flow sequentially, adjusting for target split ratios.
- Condition columns by controlled oven programming and temperature cycling to stabilize baseline and minimize bleed.
- Reverse column orientation periodically and trim inlet loops to maintain consistent efficiency.
The use of guard columns and selection of appropriate solvents for splitless or on-column injections further protect analytical columns from contaminants.
Benefits and Practical Applications
Implementation of these guidelines leads to:
- Improved chromatographic resolution and reproducibility.
- Extended column lifetime by preventing oxidative damage and contamination buildup.
- Reduced downtime and maintenance costs through proactive conditioning and leak prevention.
- Enhanced sensitivity and peak symmetry in routine analyses.
Future Trends and Applications
Emerging directions include automated ferrule fitting systems, advanced carrier gas purification technologies, and novel phases with enhanced thermal stability. Integration of real-time leak detection and flow control into GC software can further streamline column installation and monitoring.
Conclusion
By following systematic installation and maintenance practices, analysts can achieve reliable, high-quality separations with fused silica capillary columns. Regular conditioning, proper handling, and routine performance checks are essential for maximizing column performance and operational efficiency.
Reference
Grob K., High-Resolution Capillary and Capillary-Column Chromatography, HRC & CC 3 (1978) 173
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Bulletin 896 Installation and Maintenance Instructions for 0.75mm Wide Bore Capillary Columns These instructions cover instrument preparation, ferrule, and column installation, leak checking, gas flow setting procedures, and maintenance requirements for 0.75mm wide bore capillary columns in packed column (equipped…
Key words
supelco, supelcocapillary, capillarycolumn, columnfused, fusedglasseal, glassealtubing, tubingaldrich, aldrichsilica, silicasigma, sigmagow, gowleak, leakbore, boremac, macferrule, ferrulecleaving
Installation and Maintenance Instructions for 0.53mm ID Fused Silica Capillary Columns
1995|Merck|Guides
Bulletin 897 Installation and Maintenance Instructions for 0.53mm ID Fused Silica Capillary Columns These instructions cover instrument preparation, column installation, leak checking, gas flow setting procedures, and maintenance requirements for 0.53mm ID capillary columns in packed column (equipped with capillary…
Key words
supelco, supelcocapillary, capillarycolumn, columnconversion, conversionglasseal, glassealyour, yourleak, leakferrule, ferrulenut, nutgow, gowcleaving, cleavingtubing, tubinggas, gasmac, macliners
Bulletin 853B Capillary GC Troubleshooting Guide: How to Locate Problems and Solve Them The real task in correcting a problem with your capillary GC system is identifying the cause of the problem without wasting time. The systematic approach to troubleshooting…
Key words
supelco, supelcoproblem, problemcolumn, columnsymptom, symptomremedy, remedynormal, normalcause, causepurifier, purifierseptum, septumgas, gastest, testpossible, possiblerecorder, recordercapillary, capillarydetector
Reduce Inlet Liner ID for Sharper Peaks by SPME/GC
1997|Merck|Technical notes
Application Note 136 Reduce Inlet Liner ID for Sharper Peaks by SPME/GC In rapid screening analyses for volatile organic compounds, it is critical that analyte band broadening be minimized. Extracting these analytes by solid phase microextraction and replacing the standard…
Key words
chromfax, chromfaxliners, linerseach, eachtechnical, technicalservice, serviceliterature, literaturespme, spmedepartment, departmentinlet, inletliner, lineranalytes, analytessupelco, supelcomicroextraction, microextractionbjc, bjcour