GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Analysis of polydimethylsiloxanes by GC/TOFMS

Applications | 2008 | JEOLInstrumentation
GC/MSD, GC/TOF
Industries
Energy & Chemicals
Manufacturer
JEOL

Summary

Importance of the Topic


Polydimethylsiloxanes are ubiquitous in materials science, pharmaceuticals and polymer research due to their thermal stability and chemical inertness. Accurate mass calibration is essential in mass spectrometry to ensure precise molecular identification, especially when using time-of-flight (TOF) analyzers. By studying common siloxane compounds, this work addresses the need for reliable internal standards in high-resolution GC/TOFMS workflows.

Study Objectives and Overview


This application study evaluates three polydimethylsiloxane species for their fragmentation behavior and potential as internal mass calibration references. Sample 1 (octamethylcyclotetrasiloxane, MW 296) and Sample 2 (decamethylcyclopentasiloxane, MW 370) represent low-molecular-weight cyclic siloxanes. Sample 3 (octa(dimethylsiloxy)silsesquioxane, MW 1,016) is a higher-mass cage structure requiring GC inlet introduction. The goal is to characterize fragment ions under electron ionization (EI) and field ionization (FI) using the JMS-T100GC AccuTOF GC instrument.

Methodology and Instrumentation


Sample Preparation and Introduction
  • Samples 1 and 2 analyzed neat via the dedicated reservoir inlet for mass reference compounds.
  • Sample 3 prepared at 2 mg/mL in acetone and injected by conventional GC split mode.
GC Conditions
  • Column: ZB-5ms, 30 m × 0.25 mm, 0.25 µm film.
  • Injector: 320 °C, 1 mL/min constant flow, split 50:1.
  • Oven Program: 50 °C (1 min) ramp 20 °C/min to 320 °C, hold 5.5 min.
Mass Spectrometry Conditions
  • Instrument: JMS-T100GC “AccuTOF GC”.
  • Ionization: EI at 70 eV, 300 µA; FI at –10 kV, 7 mA pulse.
  • Temperatures: ion source 280 °C, GC-ITF 280 °C, reservoir 100 °C.
  • Acquisition: m/z 35–1,400, 0.4 s spectral interval.

Main Results and Discussion


Samples 1 and 2 produced dominant [M–CH3]+ ions, with calculated exact masses at m/z 248.9891, 265.0204 and 281.0517 for progressive alkyl losses. No intact molecular ions were observed in EI. Similar fragmentation patterns appeared under FI, but molecular ion signals remained below detection limits. Sample 3 yielded characteristic [M–H]+ at m/z 1,014.9595 and [M–CH3]+ at m/z 1,000.9438 in both EI and FI. In EI, additional low-mass fragments (e.g., m/z 73.0474) were detected; FI spectra showed only the two major ions. The consistency of these fragment ions under different ionization methods validates their use as internal mass markers.

Benefits and Practical Applications


Using polydimethylsiloxane fragments as internal references offers:
  • Improved mass axis calibration accuracy in GC/TOFMS analyses.
  • Enhanced reproducibility in routine quality control and polymer characterization.
  • Compatibility with both EI and FI ionization techniques.

Future Trends and Potential Applications


Future developments may include exploration of novel siloxane derivatives with tailored fragmentation, expansion to liquid chromatography-TOF platforms and optimization of ambient ionization interfaces to capture intact molecular ions. Integration into automated calibration routines could further streamline high-throughput workflows.

Conclusion


This study demonstrates that selected polydimethylsiloxanes produce stable, high-intensity fragment ions suitable for internal mass calibration in GC/TOFMS. Their predictable behavior under EI and FI enhances confidence in accurate mass measurements across a range of analytical applications.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
AccuTOF GC series Applications Notebook
AccuTOF GC series Applications Notebook
2016|Agilent Technologies|Applications
s High Performance Gas Chromatograph Time-of-Flight Mass Spectrometer AccuTOF GC series Applications Notebook Edition April 2016 AccuTOF GC series Applications Notebook Edition April 2016 Table of Contents Introduction ................................................................................................................. 1 New Gas Chromatography / High Resolution Time-of-Flight Mass Spectrometer JMS-T200GC…
Key words
accutof, accutofjeol, jeolmass, masstofms, tofmsspectrometry, spectrometryion, ionsheet, sheettips, tipsbusiness, businessgroup, groupionization, ionizationgcx, gcxセンター, センターdata, datafield
Analysis of advanced materials by FD/FI using “AccuTOF GC” - analysis of polymer building blocks with isocyanate and pilazole groups
日本電子株式会社 Mass Spectrometry Application Group Sheet JEOL MS Data She et 応用研究センター Mass Spectrometry Business Unit 〒 196-8558 東京都昭島市武蔵野 3-1-2 TelLtd. : (042) 542-2242, Fax : (042) 542-3132 JEOL MS Tips No.089 No. 040 Analysis of advanced materials by FD/FI…
Key words
moi, moiabundance, abundancerelative, relativeaoi, aoibei, beijeol, jeolblocks, blocksbuilding, buildingpilazole, pilazolenco, ncofunctional, functionaltunneling, tunnelingaccutof, accutofセンター, センターisocyanate
GC/TOFMS analysis of high boiling point compounds - cholesterol and Irganox® 1010
U U 日本電子株式会社 Mass Spectrometry Application Group JEOL MS Data Sheet 応用研究センター MS Tips Mass Spectrometry Business Unit 〒 196-8558 東京都昭島市武蔵野 3-1-2 TelLtd. : (042) 542-2242, Fax : (042) 542-3132 JEOL No.123 No. 040 GC/TOFMS analysis of high boiling point…
Key words
jeol, jeolintensity, intensitycholesterol, cholesterolpeaks, peaksセンター, センターretention, retentionrelative, relativemass, massspectrometry, spectrometrysheet, sheettofms, tofmsbusiness, businesstips, tipstime, timeboiling
Analysis of photo polymerization initiator in UV light curingadhesives by GC/TOFMS
JEOL MS Data Sheet Mass Spectrometry Application Group 日本電子株式会社 MS Tips JEOL Ltd. Tel : (042) 542-2242, Fax : (042) 542-3132 応用研究セン Mass Spectrometry Business Unit ター 〒 196-8558 東京都昭島市武蔵野 3-1-2 No.108 No. 040 Analysis of photo polymerization initiator in…
Key words
calcurated, calcuratedmormula, mormulainitiator, initiatorjeol, jeolcuring, curingpolymerization, polymerizationphoto, photomda, mdaestimated, estimatederror, errorunreacted, unreactedlight, lightadhesives, adhesivesobserved, observedadhesive
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike