High-Throughput GC-TOFMS Analysis Using EPA Method 8270D
Applications | 2008 | LECOInstrumentation
This application note addresses the need for fast, reliable analysis of semi-volatile organic compounds in environmental samples according to EPA Method 8270D. Rapid detection of alcohols, amines, aromatics, phenols, solvents and other analytes is critical for environmental monitoring, regulatory compliance and quality control. High-throughput methods enhance laboratory efficiency by reducing run times, minimizing maintenance downtime, and delivering timely, accurate results.
The goal of this study was to implement a high-speed GC-TOFMS workflow to separate and identify 90 target compounds from an EPA 8270D mid-level standard in under 12 minutes. Key objectives included meeting EPA tune criteria, achieving full chromatographic resolution, and demonstrating robust spectral deconvolution at trace concentration levels.
The optimized method achieved full separation and identification of 90 compounds in 11.2 minutes. A DFTPP tune check met all EPA 8270D abundance criteria under LECO’s ATP protocol. The total ion chromatogram showed well-resolved peaks spanning the entire run, and high spectral acquisition enabled the deconvolution of closely eluting analytes such as N-nitrosodimethylamine and pyridine separated by just 200 ms. Signal-to-noise ratios ranged from several thousand to over twenty-thousand, confirming robust detection at mid-level concentrations.
Further improvements in column technology and temperature ramping could shorten run times below ten minutes. Integration of machine-learning algorithms for real-time spectral deconvolution and non-target screening will expand method versatility. Cloud-based data management and overall workflow automation will support large-scale environmental monitoring and QA/QC testing.
This high-speed GC-TOFMS approach for EPA Method 8270D delivers rapid, accurate analysis of a broad range of semi-volatile organic compounds. It meets stringent tune requirements, resolves complex mixtures in under 12 minutes, and offers high data density for effective peak deconvolution—enabling environmental laboratories to enhance throughput and analytical confidence.
GC/MSD, GC/TOF
IndustriesEnvironmental
ManufacturerAgilent Technologies, LECO
Summary
Significance of the topic
This application note addresses the need for fast, reliable analysis of semi-volatile organic compounds in environmental samples according to EPA Method 8270D. Rapid detection of alcohols, amines, aromatics, phenols, solvents and other analytes is critical for environmental monitoring, regulatory compliance and quality control. High-throughput methods enhance laboratory efficiency by reducing run times, minimizing maintenance downtime, and delivering timely, accurate results.
Study Objectives and Overview
The goal of this study was to implement a high-speed GC-TOFMS workflow to separate and identify 90 target compounds from an EPA 8270D mid-level standard in under 12 minutes. Key objectives included meeting EPA tune criteria, achieving full chromatographic resolution, and demonstrating robust spectral deconvolution at trace concentration levels.
Methodology and Instrumentation
- Gas chromatograph: Agilent 6890 with Restek Rxi-5MS column (20 m × 0.18 mm × 0.18 µm)
- Oven program: 35 °C (0.2 min) to 320 °C at 30 °C/min, hold 2.5 min
- Injection: 1 µL split 10:1; inlet 250 °C; He carrier gas at 1.0 mL/min
- Mass spectrometer: LECO TruTOF™ HT time-of-flight MS; electron ionization at 70 eV; transfer line 280 °C; source 250 °C
- Acquisition: full mass range 30–500 u at 20 spectra/s
- Software: ChromaTOF with automated peak finding, deconvolution, and EPA tune reporting
Key Results and Discussion
The optimized method achieved full separation and identification of 90 compounds in 11.2 minutes. A DFTPP tune check met all EPA 8270D abundance criteria under LECO’s ATP protocol. The total ion chromatogram showed well-resolved peaks spanning the entire run, and high spectral acquisition enabled the deconvolution of closely eluting analytes such as N-nitrosodimethylamine and pyridine separated by just 200 ms. Signal-to-noise ratios ranged from several thousand to over twenty-thousand, confirming robust detection at mid-level concentrations.
Benefits and Practical Applications
- High throughput reduces analysis time by over 50% relative to conventional methods
- TOFMS detection provides trace-level sensitivity and full mass-range coverage
- Automated tune compliance streamlines quality control and regulatory reporting
- Advanced deconvolution improves confidence in identifying co-eluting compounds
Future Trends and Opportunities
Further improvements in column technology and temperature ramping could shorten run times below ten minutes. Integration of machine-learning algorithms for real-time spectral deconvolution and non-target screening will expand method versatility. Cloud-based data management and overall workflow automation will support large-scale environmental monitoring and QA/QC testing.
Conclusion
This high-speed GC-TOFMS approach for EPA Method 8270D delivers rapid, accurate analysis of a broad range of semi-volatile organic compounds. It meets stringent tune requirements, resolves complex mixtures in under 12 minutes, and offers high data density for effective peak deconvolution—enabling environmental laboratories to enhance throughput and analytical confidence.
Reference
- EPA Alternative Test Procedure letter 2-9-05 (ATP case numbers D04-002 and N04-002) approving DFTPP tune criteria for LECO GC-TOFMS instruments.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Optimized analysis of semi-volatile organic compounds (SVOC) in environmental samples in compliance with U.S. EPA Method 8270 utilizing GC-MS
2022|Thermo Fisher Scientific|Applications
Application note | 000761 Mass spectrometry Optimized analysis of semi-volatile organic compounds (SVOC) in environmental samples in compliance with U.S. EPA Method 8270 utilizing GC-MS Introduction Authors Semi-volatile organic compounds (SVOCs) are ubiquitous in our environment and raise Amit Gujar…
Key words
avgcalfact, avgcalfactphthalate, phthalatebenzo, benzofluoranthene, fluoranthenepass, passfactor, factorbenzyl, benzylresponse, responseanthracene, anthracenepyrene, pyreneminimum, minimumname, namefit, fitdiethylphthalate, diethylphthalatedioctyl
Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and the Agilent HydroInert Source by Gas Chromatography/Mass Spectrometry
2022|Agilent Technologies|Applications
Application Note Environmental Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and the Agilent HydroInert Source by Gas Chromatography/Mass Spectrometry Author Angela Smith Henry, PhD Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry (GC/MS) is integral to the analysis of…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluoranthenepass, passnitrobenzene, nitrobenzeneanthracene, anthracenelinear, lineararamite, aramitemethanesulfonate, methanesulfonatediallate, diallatepyrene, pyreneazobenzene, azobenzenebenzyl, benzylnitroso, nitrosohydroinert
EPA Method 8270 Semivolatile Organic Compounds Analysis on the Pegasus BT: A Benchtop GC-TOFMS
2016|Agilent Technologies|Applications
EPA Method 8270 Semivolatile Organic Compounds Analysis on the Pegasus® BT: A Benchtop GC-TOFMS LECO Corporation; Saint Joseph, Michigan USA Key Words: EPA Method 8270, GC-TOFMS, Sensitivity, Linear Dynamic Range, Robust, Reproducible Figure 1. Extracted ion chromatogram (XIC) of the…
Key words
analyte, analytepass, passsurrogate, surrogatephthalate, phthalatebenzo, benzoscience, sciencefluoranthene, fluoranthenelife, lifedelivering, deliveringchemical, chemicalpegasus, pegasusavg, avgsolutions, solutionsbenzyl, benzylright
EPA Method 8270 Semi Volatile Organic Compounds Analysis on a New Benchtop Time-of-Flight Mass Spectrometer
2017|Agilent Technologies|Posters
EPA Method 8270 Semi Volatile Organic Compounds Analysis on a New Benchtop Time-of-Flight Mass Spectrometer Jonathan D. Byer, Todd Richards, Lorne M. Fell, Joseph E. Binkley | LECO Corporation, Saint Joseph, MI USA Instrument Parameters and System Performance Evaluation GC-TOFMS…
Key words
analyte, analytepass, passsurrogate, surrogatephthalate, phthalatepegasus, pegasusbenzo, benzofluoranthene, fluorantheneavg, avghexachlorobenzene, hexachlorobenzenebenzyl, benzylleco, lecoanthracene, anthracenepyrene, pyrenecriteria, criterianitrosodi