Analysis of Semivolatile Organic Compounds with Agilent Sintered Frit Liner by Gas Chromatography/Mass Spectrometry
Applications | 2019 | Agilent TechnologiesInstrumentation
Gas chromatography–mass spectrometry (GC–MS) is a cornerstone technique for quantifying semivolatile organic compounds (SVOCs) in complex environmental samples such as soil, wastewater, and solid waste extracts. The choice of GC inlet liner greatly influences system inertness, peak integrity, maintenance intervals, and overall laboratory throughput. The Agilent Ultra Inert splitless single-taper liner with a sintered frit packing is designed to minimize active sites, prevent glass wool breakage, and extend operational lifetime under challenging matrix loads.
This study evaluated the performance of the Agilent sintered frit liner against EPA Method 8270D/E requirements for SVOC analysis. Key goals included:
An Agilent 7890B GC with a single MS flowpath and Inert EI source was coupled to an Agilent 5977 MSD. A 30 m × 0.25 mm × 0.25 µm DB-8270D Ultra Inert column was used with a 9 mm drawout plate. The inlet liner was an Agilent Ultra Inert splitless single-taper with sintered frit. Sample introduction was via a 10 µL PTFE-tip syringe, and automated injections (1 µL) were performed with an Agilent 7650A autosampler. Carrier gas helium was held at 1.30 mL/min. The temperature program ramped from 40 °C to 320 °C in stepped increments to achieve a ~24 min run time.
• System inertness: Measured 4,4′-DDT breakdown averaged 0.9% (limit < 20%), and ion-ratio tailing factors for pentachlorophenol and benzidine remained < 1.6 (limit ≤ 2.0).
• Isomer resolution: All critical pairs (benzo(b/k)fluoranthene, benz[a]anthracene/chrysene, phenanthrene/anthracene, 1-/2-naphthalenamine) were baseline or near‐baseline resolved.
• Calibration: Linear, weighted (1/x) regression yielded R² > 0.995 for active compounds. Average response-factor RSD across 0.1–100 µg/mL was 10.3% (requirement ≤ 20%).
• CCV performance: Verification every 12 h showed < 10% of compounds drifted beyond ±20%, even after >200 injections, well below the EPA 8270E threshold of 20%.
• Internal standards: Normalized areas for six deuterated ISTDs remained within ±2× over 260 injections, indicating stable MS sensitivity.
• Liner lifetime: In a sequence of 260 soil extract injections, each sintered frit liner sustained acceptable performance for an average of 23 injections before reaching cleanup thresholds and requiring replacement. After each liner change, all suitability checks passed immediately.
As environmental monitoring demands evolve, sintered frit liners may be paired with high-resolution and tandem MS systems to improve sensitivity and selectivity. Automated liner exchange and in-line matrix removal could further extend lifetime and throughput. Research into alternative inert packings and miniaturized inlet designs may optimize performance for emerging contaminants.
The Agilent Ultra Inert splitless single-taper liner with sintered frit demonstrates excellent inertness, reproducibility, and extended lifetime under EPA 8270E conditions. The liner supports stable calibration, minimal peak tailing, and reliable system suitability, making it a valuable component for routine SVOC analysis in complex environmental matrices.
GC/MSD, GC/SQ, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Significance of the Topic
Gas chromatography–mass spectrometry (GC–MS) is a cornerstone technique for quantifying semivolatile organic compounds (SVOCs) in complex environmental samples such as soil, wastewater, and solid waste extracts. The choice of GC inlet liner greatly influences system inertness, peak integrity, maintenance intervals, and overall laboratory throughput. The Agilent Ultra Inert splitless single-taper liner with a sintered frit packing is designed to minimize active sites, prevent glass wool breakage, and extend operational lifetime under challenging matrix loads.
Objectives and Study Overview
This study evaluated the performance of the Agilent sintered frit liner against EPA Method 8270D/E requirements for SVOC analysis. Key goals included:
- Verifying system inertness and reproducibility with tuning standards (DFTPP, 4,4′-DDT, pentachlorophenol, benzidine).
- Assessing chromatographic resolution of critical isomer pairs (e.g., benzo(b)fluoranthene/benzo(k)fluoranthene).
- Characterizing calibration linearity, response-factor stability, and internal standard behavior over 0.1–100 µg/mL.
- Determining liner lifetime by repeated soil extract injections until performance limits were reached.
Methodology and Instrumentation
An Agilent 7890B GC with a single MS flowpath and Inert EI source was coupled to an Agilent 5977 MSD. A 30 m × 0.25 mm × 0.25 µm DB-8270D Ultra Inert column was used with a 9 mm drawout plate. The inlet liner was an Agilent Ultra Inert splitless single-taper with sintered frit. Sample introduction was via a 10 µL PTFE-tip syringe, and automated injections (1 µL) were performed with an Agilent 7650A autosampler. Carrier gas helium was held at 1.30 mL/min. The temperature program ramped from 40 °C to 320 °C in stepped increments to achieve a ~24 min run time.
Key Results and Discussion
• System inertness: Measured 4,4′-DDT breakdown averaged 0.9% (limit < 20%), and ion-ratio tailing factors for pentachlorophenol and benzidine remained < 1.6 (limit ≤ 2.0).
• Isomer resolution: All critical pairs (benzo(b/k)fluoranthene, benz[a]anthracene/chrysene, phenanthrene/anthracene, 1-/2-naphthalenamine) were baseline or near‐baseline resolved.
• Calibration: Linear, weighted (1/x) regression yielded R² > 0.995 for active compounds. Average response-factor RSD across 0.1–100 µg/mL was 10.3% (requirement ≤ 20%).
• CCV performance: Verification every 12 h showed < 10% of compounds drifted beyond ±20%, even after >200 injections, well below the EPA 8270E threshold of 20%.
• Internal standards: Normalized areas for six deuterated ISTDs remained within ±2× over 260 injections, indicating stable MS sensitivity.
• Liner lifetime: In a sequence of 260 soil extract injections, each sintered frit liner sustained acceptable performance for an average of 23 injections before reaching cleanup thresholds and requiring replacement. After each liner change, all suitability checks passed immediately.
Benefits and Practical Applications
- Extended maintenance intervals and reduced downtime in environmental laboratories.
- Consistent quantitation across liner changes without recalibration.
- Robust performance in high‐throughput SVOC analysis for regulatory compliance (EPA 8270D/E).
- Lower risk of column and ion source contamination from matrix carryover.
Future Trends and Applications
As environmental monitoring demands evolve, sintered frit liners may be paired with high-resolution and tandem MS systems to improve sensitivity and selectivity. Automated liner exchange and in-line matrix removal could further extend lifetime and throughput. Research into alternative inert packings and miniaturized inlet designs may optimize performance for emerging contaminants.
Conclusion
The Agilent Ultra Inert splitless single-taper liner with sintered frit demonstrates excellent inertness, reproducibility, and extended lifetime under EPA 8270E conditions. The liner supports stable calibration, minimal peak tailing, and reliable system suitability, making it a valuable component for routine SVOC analysis in complex environmental matrices.
References
- US EPA Method 8270D: Semivolatile Organic Compounds by GC–MS, Revision 4, February 2007.
- US EPA Method 8270E: Semivolatile Organic Compounds by GC–MS, Revision 4, June 2018.
- Padilla-Sánchez J. A., Plaza-Bolaños P., Frenich A. G., "Advanced GC–LRMS Strategies for Environmental Contaminants," in Comprehensive Analytical Chemistry, Ferrer I., Thurman E., Eds., Elsevier, 2013.
- Agilent Technologies, "Analysis of SVOCs Using the Intuvo 9000 GC," Application Note 5991-7256EN, 2016.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Comparison of Fritted and Wool Liners for Analysis of Semivolatile Organic Compounds by Gas Chromatography/Mass Spectrometry
2020|Agilent Technologies|Applications
Application Note Environmental Comparison of Fritted and Wool Liners for Analysis of Semivolatile Organic Compounds by Gas Chromatography/Mass Spectrometry Author Angela Smith Henry, PhD Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry (GC/MS) is commonly used in the analysis of semivolatile…
Key words
wool, woolliner, linerfrit, fritliners, linersglass, glassbreakdown, breakdownddt, ddtccv, ccvagilent, agilenttrim, trimfailing, failingtaper, taperinitial, initialcalibration, calibrationfritted
Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph
2018|Agilent Technologies|Applications
Application Note Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph Author Matthew Giardina, Ph.D., Applications Chemist, Gas Phase Separations Division, Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry is a workhorse for the analysis of semivolatile organic…
Key words
intuvo, intuvochip, chipreplacement, replacementguard, guardpass, passccv, ccvbenzidine, benzidinebenzo, benzomatrix, matrixresponse, responsepentachlorophenol, pentachlorophenolcalibration, calibrationafter, afternumber, numberinjections
Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and the Agilent HydroInert Source by Gas Chromatography/Mass Spectrometry
2022|Agilent Technologies|Applications
Application Note Environmental Analysis of Semivolatile Organic Compounds Using Hydrogen Carrier Gas and the Agilent HydroInert Source by Gas Chromatography/Mass Spectrometry Author Angela Smith Henry, PhD Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry (GC/MS) is integral to the analysis of…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluoranthenepass, passnitrobenzene, nitrobenzeneanthracene, anthracenearamite, aramitelinear, linearmethanesulfonate, methanesulfonatediallate, diallatepyrene, pyreneazobenzene, azobenzenebenzyl, benzylnitroso, nitrosohydroinert
Selecting a GC Liner to Decrease Column Trim Frequency
2022|Agilent Technologies|Technical notes
Application Brief Environmental Selecting a GC Liner to Decrease Column Trim Frequency Author Abstract Angela Smith Henry, PhD Agilent Technologies, Inc. GC inlet liner selection is an important part of reduced column maintenance. A liner with a frit or wool…
Key words
liner, linerfritted, frittedddt, ddtbreakdown, breakdowntaper, taperpercent, percentwool, woolcolumn, columnmatrix, matrixtrim, trimliners, linerstrims, trimsinitial, initialsingle, singlefrit