Explosives - Rtx®-200
Applications | | RestekInstrumentation
Reliable analysis of nitroaromatic and aminotoluene-based explosives is critical for forensic investigations, environmental monitoring, and homeland security. Gas chromatography coupled with mass spectrometry (GC-MS) offers high sensitivity and selectivity for trace-level detection, enabling clear separation and identification of structurally similar compounds in complex matrices.
This application note presents a GC-MS method using an Rtx®-200 column for the simultaneous analysis of 23 explosives and related compounds. The primary goal is to achieve baseline separation of nitrotoluenes, dinitrotoluenes, trinitrotoluenes, aminonitrotoluenes, and related nitroaromatic compounds in a single run.
The optimized temperature program allows clear resolution of all 23 analytes within a 55-minute run. Key separations include: 2-nitrotoluene vs. 3-nitrotoluene vs. 4-nitrotoluene, and closely eluting isomers such as 2,3-dinitrotoluene vs. 2,6-dinitrotoluene. Trinitrotoluenes (TNT, 2,4,5-TNT) and higher molecular weight nitrobiphenyls also exhibit sharp peaks with minimal tailing. The use of Rtx-200 stationary phase provides enhanced polarity interactions, improving separation of amino derivatives from nitro-only compounds.
Advancements in column technology, such as shorter high-efficiency capillaries and hybrid stationary phases, could reduce analysis time and increase throughput. Coupling with tandem mass spectrometry (GC-MS/MS) will further enhance sensitivity and reduce background interference. Miniaturized GC-MS systems may enable on-site explosive screening in field deployments.
The described GC-MS method using Rtx®-200 demonstrates reliable separation and quantitation of a broad range of nitroaromatic explosives. Its robustness and sensitivity make it well suited for forensic, environmental, and security applications.
GC/MSD, GC columns, Consumables
IndustriesEnvironmental, Homeland Security
ManufacturerRestek
Summary
Significance of the Topic
Reliable analysis of nitroaromatic and aminotoluene-based explosives is critical for forensic investigations, environmental monitoring, and homeland security. Gas chromatography coupled with mass spectrometry (GC-MS) offers high sensitivity and selectivity for trace-level detection, enabling clear separation and identification of structurally similar compounds in complex matrices.
Objectives and Study Overview
This application note presents a GC-MS method using an Rtx®-200 column for the simultaneous analysis of 23 explosives and related compounds. The primary goal is to achieve baseline separation of nitrotoluenes, dinitrotoluenes, trinitrotoluenes, aminonitrotoluenes, and related nitroaromatic compounds in a single run.
Methodology and Instrumentation
- Column: Rtx®-200, 30 m length, 0.25 mm ID, 0.25 µm film thickness (cat. #15023)
- Injection: 1.0 µL splitless of 20 ng/µL standard solution
- Oven Program: 80 °C hold 2 min → ramp at 3 °C/min to 260 °C, hold 2 min
- Injector Temperature: 280 °C
- Carrier Gas: Helium at 20 cm/s (set at 80 °C)
- Detector: Mass spectrometer at 300 °C
- Splitless Hold Time: 0.6 min
Main Results and Discussion
The optimized temperature program allows clear resolution of all 23 analytes within a 55-minute run. Key separations include: 2-nitrotoluene vs. 3-nitrotoluene vs. 4-nitrotoluene, and closely eluting isomers such as 2,3-dinitrotoluene vs. 2,6-dinitrotoluene. Trinitrotoluenes (TNT, 2,4,5-TNT) and higher molecular weight nitrobiphenyls also exhibit sharp peaks with minimal tailing. The use of Rtx-200 stationary phase provides enhanced polarity interactions, improving separation of amino derivatives from nitro-only compounds.
Benefits and Practical Applications
- Forensic Screening: Rapid confirmation of explosive residues from post-blast samples.
- Environmental Testing: Trace monitoring of nitroaromatic pollutants in soil and water.
- Security Checks: Screening of suspicious materials at checkpoints.
- Quality Control: Verification of purity in synthesized energetic compounds.
Future Trends and Potential Uses
Advancements in column technology, such as shorter high-efficiency capillaries and hybrid stationary phases, could reduce analysis time and increase throughput. Coupling with tandem mass spectrometry (GC-MS/MS) will further enhance sensitivity and reduce background interference. Miniaturized GC-MS systems may enable on-site explosive screening in field deployments.
Conclusion
The described GC-MS method using Rtx®-200 demonstrates reliable separation and quantitation of a broad range of nitroaromatic explosives. Its robustness and sensitivity make it well suited for forensic, environmental, and security applications.
Used Instrumentation
- Gas Chromatograph: Rtx®-200 30 m × 0.25 mm, 0.25 µm
- Injection System: Splitless, 280 °C
- Detector: Mass Spectrometer, 300 °C
- Carrier Gas: Helium
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Explosives US EPA Method 8095 - Rtx®-TNT
|Restek|Applications
Explosives US EPA Method 8095 Rtx®-TNT restek innovation! Rtx®-TNT Improved resolution of nitroaromatic compounds such as those listed in US EPA Method 8095. 9 7 15 11 16 14 8 10 17 6 13 1 5 12 2 34 min.…
Key words
tnt, tntrtx, rtxtrinitrotoluene, trinitrotoluenenitroglycerine, nitroglycerinenitroaromatic, nitroaromaticpetn, petntetryl, tetrylrdx, rdxexplosives, explosivesnitrobenzene, nitrobenzenerestek, restekinnovation, innovationepa, epalisted, listedthose
Explosives US EPA Method 8095 - Rtx®-TNT2
|Restek|Applications
Explosives US EPA Method 8095 Rtx®-TNT2 restek innovation! Improved resolution of nitroaromatic compounds such as those listed in US EPA Method 8095. Rtx®-TNT2 15,13 9 7 16 11 12,14 8 10 6 17 5 1 min. 2 3 4 2…
Key words
trinitrotoluene, trinitrotoluenenitroglycerine, nitroglycerinenitroaromatic, nitroaromaticpetn, petntetryl, tetrylrdx, rdxexplosives, explosivesnitrobenzene, nitrobenzenerestek, restekinnovation, innovationepa, epalisted, listedthose, thoseimproved, improvedresolution
On-site Screening of Explosive Materials by Portable SPME-GC/MS and CME-GC/MS
2019|PerkinElmer|Applications
APPLICATION Gas Chromatography / Mass Spectrometry Authors: Zachary Lawton Tiffany Arbon PerkinElmer, Inc. Shelton, CT Thorsten Heinlein Abacus Analytical Systems Berlin, Germany On-site Screening of Explosive Materials by Portable SPME-GC/MS and CME-GC/MS Introduction Explosive threats are continually evolving as…
Key words
explosive, explosivecme, cmespme, spmemicroextraction, microextractioncustodion, custodionexplosives, explosivestnt, tnttrinitrotoluene, trinitrotoluenedinitrate, dinitratestabilizers, stabilizersthreats, threatsmaterials, materialsabacus, abacustagging, taggingglycol
Semivolatile Organics US EPA Method 8270 - Rtx®-5Sil MS
|Restek|Applications
Semivolatile Organics US EPA Method 8270 Rtx®-5Sil MS Excellent benzo(b) & benzo(k) fluoranthene separation! 47 30m, 0.25mm ID, 0.5µm Rtx®-5Sil MS (cat# 12738) 80ng splitless injection of 8270 standard GC: Agilent 6890 Det.: MSD Oven temp.: 40°C (hold 2 min)…
Key words
fluoranthene, fluoranthenebenzo, benzohexachloropropene, hexachloropropenesilarylene, silaryleneisosafrole, isosafrolesafrole, safrolebutylphthalate, butylphthalatedimethylphthalate, dimethylphthalatephenacetin, phenacetindinoseb, dinosebpentachloronitrobenzene, pentachloronitrobenzeneisodrin, isodrinazobenzene, azobenzenebenzidine, benzidinedibenzofuran