Explosives - Rtx®-200

Applications |  | RestekInstrumentation
GC/MSD, GC columns, Consumables
Industries
Environmental, Homeland Security
Manufacturer
Restek

Summary

Significance of the Topic


Reliable analysis of nitroaromatic and aminotoluene-based explosives is critical for forensic investigations, environmental monitoring, and homeland security. Gas chromatography coupled with mass spectrometry (GC-MS) offers high sensitivity and selectivity for trace-level detection, enabling clear separation and identification of structurally similar compounds in complex matrices.

Objectives and Study Overview


This application note presents a GC-MS method using an Rtx®-200 column for the simultaneous analysis of 23 explosives and related compounds. The primary goal is to achieve baseline separation of nitrotoluenes, dinitrotoluenes, trinitrotoluenes, aminonitrotoluenes, and related nitroaromatic compounds in a single run.

Methodology and Instrumentation


  • Column: Rtx®-200, 30 m length, 0.25 mm ID, 0.25 µm film thickness (cat. #15023)
  • Injection: 1.0 µL splitless of 20 ng/µL standard solution
  • Oven Program: 80 °C hold 2 min → ramp at 3 °C/min to 260 °C, hold 2 min
  • Injector Temperature: 280 °C
  • Carrier Gas: Helium at 20 cm/s (set at 80 °C)
  • Detector: Mass spectrometer at 300 °C
  • Splitless Hold Time: 0.6 min

Main Results and Discussion


The optimized temperature program allows clear resolution of all 23 analytes within a 55-minute run. Key separations include: 2-nitrotoluene vs. 3-nitrotoluene vs. 4-nitrotoluene, and closely eluting isomers such as 2,3-dinitrotoluene vs. 2,6-dinitrotoluene. Trinitrotoluenes (TNT, 2,4,5-TNT) and higher molecular weight nitrobiphenyls also exhibit sharp peaks with minimal tailing. The use of Rtx-200 stationary phase provides enhanced polarity interactions, improving separation of amino derivatives from nitro-only compounds.

Benefits and Practical Applications


  • Forensic Screening: Rapid confirmation of explosive residues from post-blast samples.
  • Environmental Testing: Trace monitoring of nitroaromatic pollutants in soil and water.
  • Security Checks: Screening of suspicious materials at checkpoints.
  • Quality Control: Verification of purity in synthesized energetic compounds.

Future Trends and Potential Uses


Advancements in column technology, such as shorter high-efficiency capillaries and hybrid stationary phases, could reduce analysis time and increase throughput. Coupling with tandem mass spectrometry (GC-MS/MS) will further enhance sensitivity and reduce background interference. Miniaturized GC-MS systems may enable on-site explosive screening in field deployments.

Conclusion


The described GC-MS method using Rtx®-200 demonstrates reliable separation and quantitation of a broad range of nitroaromatic explosives. Its robustness and sensitivity make it well suited for forensic, environmental, and security applications.

Used Instrumentation


  • Gas Chromatograph: Rtx®-200 30 m × 0.25 mm, 0.25 µm
  • Injection System: Splitless, 280 °C
  • Detector: Mass Spectrometer, 300 °C
  • Carrier Gas: Helium

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Explosives US EPA Method 8095 - Rtx®-TNT
Explosives US EPA Method 8095 Rtx®-TNT restek innovation! Rtx®-TNT Improved resolution of nitroaromatic compounds such as those listed in US EPA Method 8095. 9 7 15 11 16 14 8 10 17 6 13 1 5 12 2 34 min.…
Key words
tnt, tntrtx, rtxtrinitrotoluene, trinitrotoluenenitroglycerine, nitroglycerinenitroaromatic, nitroaromaticpetn, petntetryl, tetrylrdx, rdxexplosives, explosivesnitrobenzene, nitrobenzenerestek, restekinnovation, innovationepa, epalisted, listedthose
Explosives US EPA Method 8095 - Rtx®-TNT2
Explosives US EPA Method 8095 Rtx®-TNT2 restek innovation! Improved resolution of nitroaromatic compounds such as those listed in US EPA Method 8095. Rtx®-TNT2 15,13 9 7 16 11 12,14 8 10 6 17 5 1 min. 2 3 4 2…
Key words
trinitrotoluene, trinitrotoluenenitroglycerine, nitroglycerinenitroaromatic, nitroaromaticpetn, petntetryl, tetrylrdx, rdxexplosives, explosivesnitrobenzene, nitrobenzenerestek, restekinnovation, innovationepa, epalisted, listedthose, thoseimproved, improvedresolution
On-site Screening of Explosive Materials by Portable SPME-GC/MS and CME-GC/MS
APPLICATION Gas Chromatography / Mass Spectrometry Authors: Zachary Lawton Tiffany Arbon PerkinElmer, Inc. Shelton, CT  Thorsten Heinlein Abacus Analytical Systems Berlin, Germany On-site Screening of Explosive Materials by Portable SPME-GC/MS and CME-GC/MS Introduction Explosive threats are continually evolving as…
Key words
explosive, explosivecme, cmespme, spmemicroextraction, microextractioncustodion, custodionexplosives, explosivestnt, tnttrinitrotoluene, trinitrotoluenedinitrate, dinitratestabilizers, stabilizersthreats, threatsmaterials, materialsabacus, abacustagging, taggingglycol
Semivolatile Organics US EPA Method 8270 - Rtx®-5Sil MS
Semivolatile Organics US EPA Method 8270 Rtx®-5Sil MS Excellent benzo(b) & benzo(k) fluoranthene separation! 47 30m, 0.25mm ID, 0.5µm Rtx®-5Sil MS (cat# 12738) 80ng splitless injection of 8270 standard GC: Agilent 6890 Det.: MSD Oven temp.: 40°C (hold 2 min)…
Key words
fluoranthene, fluoranthenebenzo, benzohexachloropropene, hexachloropropenesilarylene, silaryleneisosafrole, isosafrolesafrole, safrolebutylphthalate, butylphthalatedimethylphthalate, dimethylphthalatephenacetin, phenacetindinoseb, dinosebpentachloronitrobenzene, pentachloronitrobenzeneisodrin, isodrinazobenzene, azobenzenebenzidine, benzidinedibenzofuran
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike