GCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Volatile Organics Confirmational Analysis EPA Method 502.2 Rtx®-1

Applications |  | RestekInstrumentation
GC, Purge and Trap, GC columns, Consumables
Industries
Environmental
Manufacturer
Thermo Fisher Scientific, Restek, Teledyne LABS

Summary

Importance of Monitoring Volatile Organic Compounds


Volatile organic compounds present a significant concern in environmental and water quality analysis due to their health and ecological impacts. Reliable quantification ensures compliance with regulatory standards and supports public health protection.

Objectives and Study Overview


This study follows EPA Method 502.2 to confirm the presence and concentration of 71 VOCs in water samples. The aim is to demonstrate the performance of a gas chromatograph system employing an Rtx 1 column combined with a tandem photoionization and Hall effect detector configuration.

Methodology and Instrumentation


  • Sample preparation 20 ppb aqueous standards prepared in 5 mL of ultrapure water
  • Purge and trap Tekmar LSC 3000 purger with Vocarb 3000 trap followed by thermal desorption
  • Gas chromatography Finnigan 9001 GC with an Rtx 1 column 75 m length 0.45 mm ID 2.55 micrometer film
  • Detection uGold tandem PID Hall 2000 detector array to enhance selectivity and sensitivity
  • Carrier gas Helium at constant flow of 9 mL per minute
  • Oven program 35 °C hold 6 min ramp to 115 °C at 11 °C per min hold 7 min ramp to 130 °C at 7 °C per min then to 220 °C at 9.2 °C per min hold 4 min

Main Results and Discussion


The method achieved baseline separation for 71 target VOCs including key chlorinated solvents benzene and trichloroethylene. Retention time precision remained within acceptable limits enabling reliable confirmation. The tandem detector system provided dual-mode response improving compound identification and quantification particularly for low-level analytes.

Benefits and Practical Applications


  • High sensitivity detection limits in low part per billion range
  • Broad analyte coverage from light chloromethanes to heavier aromatics
  • Enhanced selectivity dual PID and Hall detectors reduce false positives
  • Regulatory compliance suitable for routine water quality testing under EPA guidelines

Future Trends and Potential Applications


Advancements in detector technologies may further improve detection limits and spectral identification. Integration with mass spectrometry or high resolution detectors will expand applications in complex environmental matrices. Automated sample handling can drive higher throughput in industrial and regulatory laboratories.

Conclusion


The combined purge and trap GC method with tandem PID Hall detection offers a robust platform for comprehensive VOC analysis in water. Its performance under EPA Method 502.2 conditions supports accurate quantification across a wide spectrum of volatile organics enabling effective monitoring programs.

Instrumentation Summary


  • Column Rtx 1 75 m 0.45 mm ID 2.55 micrometer film
  • Purge and trap Tekmar LSC 3000 with Vocarb 3000 trap
  • GC Finnigan 9001
  • Detection uGold PID Hall 2000

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Volatile Organics US EPA Method 502.2 Rtx®-502.2
Volatile Organics US EPA Method 502.2 Rtx®-502.2
|Thermo Fisher Scientific|Applications
Volatile Organics US EPA Method 502.2 Rtx®-502.2 for more info A confirmational analysis can be found on www.restek.com/chromatograms Search for chromatogram number GC_EV00006 54/55 69 16 14 17 25 Rtx®-502.2 75m, 0.45mm ID, 2.55µm (cat.# 10986) 14 67 21 42/43…
Key words
butylbenzene, butylbenzenechloride, chloridexylene, xylenebenzene, benzenetert, tertelcd, elcdconfirmational, confirmationalpid, pidfluorobenzene, fluorobenzeneisopropyltoluene, isopropyltolueneallyl, allylinfo, infodibromomethane, dibromomethanedichlorodifluoromethane, dichlorodifluoromethanetrichlorofluoromethane
Volatile Organics US EPA Method 8021 Rtx®-VRX
Volatile Organics US EPA Method 8021 Rtx®-VRX
|Thermo Fisher Scientific|Applications
Volatile Organics US EPA Method 8021 Rtx®-VRX 39/40 PID • Good choice for waste water analysis. 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. 16. 17. 18. 19. 20. 21. 22. 23. 24.…
Key words
butylbenzene, butylbenzenetert, tertvrx, vrxpid, pidxylene, xylenefluorobenzene, fluorobenzenebenzene, benzenertx, rtxether, etherbutyl, butylelcd, elcdchloride, chloridesilcosteel, silcosteelmethyl, methylisopropyltoluene
Volatile Organics EPA Method 8021A/502.2 - Rtx®-502.2
Volatile Organics EPA Method 8021A/502.2 - Rtx®-502.2
|Thermo Fisher Scientific|Applications
Volatile Organics EPA Method 8021A/502.2 Rtx®-502.2 Confirmational column to the Rtx®-VGC. Rtx®-502.2 49 51 75m, 0.45mm ID, 2.55µm (cat.# 10986) 30 9 38 46,47 60 50 57 14,13 35 39,40 10 6 63 58 48 32 16 64 66 2…
Key words
organics, organicsconcentrator, concentratorpress, pressconnector, connectortight, tightconnection, connectionvolatile, volatiledirect, directcolumns, columnscolumn, columnfrom
Optimizing the Analysis of Volatile Organic Compounds
Technical Guide 1 Optimizing the Analysis of Volatile Organic Compounds Inside: EPA Method Definitions State GRO Methods Contract Laboratory Program (CLP) The Love Canal Scandal Purge and Trap Theory Sequences and Flow Paths of the Purge and Trap Unit Purge…
Key words
pack, packxylene, xylenertx, rtxtert, tertpid, pidpurge, purgebutylbenzene, butylbenzeneelcd, elcdeach, eachchloride, chloridetrap, trapbenzene, benzenedata, dataacetate, acetatevgc
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike