Fast GC-MS Analysis of US EPA Method 8270D Semivolatiles Using Supelco SLB-5ms
Applications | 2006 | MerckInstrumentation
Semivolatile organic compounds regulated under US EPA Method 8270D are critical targets in environmental monitoring of solid waste and groundwater. Their diverse chemical properties and the requirement to complete analyses within a 12-hour window place high demands on chromatographic throughput and sensitivity. Implementing fast GC-MS workflows addresses these challenges by shortening run times while maintaining resolution and instrument performance.
This study aimed to apply fast gas chromatography principles to a standard 80-component EPA 8270D semivolatile mix and demonstrate complete separation in under 8.5 minutes. Key goals were to maintain chromatographic resolution, meet regulatory MS tune criteria, and ensure reliable quantitation using internal standards.
The method combined a high-efficiency column with rapid temperature programming and elevated carrier gas velocity.
The optimized method achieved baseline resolution of all 86 target compounds within an 8.5-minute window. Early-eluting analytes, including nitrosamines and phenols, were separated by 2–3 minutes, while high-molecular-weight polyaromatic hydrocarbons eluted between 6 and 8.5 minutes. Peak shapes remained sharp despite high ramp rates, and internal standards tracked expected retention behavior, confirming stable MS response and reproducible chromatography.
Advances in fast GC column technology and high-speed detectors will further shorten analysis times. Integration with automated sample preparation, two-dimensional GC, and high-resolution MS can extend this approach to emerging contaminants and complex matrices. Miniaturized and field-deployable fast GC-MS systems may enable on-site semivolatile monitoring.
This application demonstrates that fast GC-MS using a 20 m × 0.18 mm SLB-5ms column achieves full separation of EPA Method 8270D targets in under 8.5 minutes. The method meets regulatory performance criteria, boosts laboratory throughput, and offers a robust platform for routine semivolatile analysis.
GC/MSD, GC columns, Consumables
IndustriesEnvironmental
ManufacturerMerck
Summary
Significance of the Topic
Semivolatile organic compounds regulated under US EPA Method 8270D are critical targets in environmental monitoring of solid waste and groundwater. Their diverse chemical properties and the requirement to complete analyses within a 12-hour window place high demands on chromatographic throughput and sensitivity. Implementing fast GC-MS workflows addresses these challenges by shortening run times while maintaining resolution and instrument performance.
Objectives and Study Overview
This study aimed to apply fast gas chromatography principles to a standard 80-component EPA 8270D semivolatile mix and demonstrate complete separation in under 8.5 minutes. Key goals were to maintain chromatographic resolution, meet regulatory MS tune criteria, and ensure reliable quantitation using internal standards.
Methodology and Instrumentation Used
The method combined a high-efficiency column with rapid temperature programming and elevated carrier gas velocity.
- Column: Supelco SLB-5ms, 20 m × 0.18 mm I.D., 0.18 µm film thickness
- Oven program: 40 °C hold 0.7 min; ramp 55 °C/min to 240 °C; ramp 28 °C/min to 330 °C, hold 2 min
- Injection: 0.5 µL sample, split 10:1, injector at 250 °C
- Carrier gas: Helium, constant flow for 40 cm/s linear velocity
- Sample preparation: 80-component semivolatile standard at 50 ppm with six deuterated/internal standards at 40 ppm in methylene chloride
Instrumental Setup
- Gas chromatograph: Agilent 6890GC
- Mass spectrometer: Agilent 5973 MSD, transfer line at 330 °C, scan range m/z 40–450
- Injection liner: FAST FocusLiner™ with taper (2879501-U)
Main Results and Discussion
The optimized method achieved baseline resolution of all 86 target compounds within an 8.5-minute window. Early-eluting analytes, including nitrosamines and phenols, were separated by 2–3 minutes, while high-molecular-weight polyaromatic hydrocarbons eluted between 6 and 8.5 minutes. Peak shapes remained sharp despite high ramp rates, and internal standards tracked expected retention behavior, confirming stable MS response and reproducible chromatography.
Benefits and Practical Applications
- Run time reduced by more than 60% compared to conventional methods
- Enables compliance with EPA 8270D 12-hour analysis window
- Increases sample throughput in high-volume laboratories
- Reduces carrier gas consumption and energy use
- Maintains resolution and sensitivity for regulatory reporting
Future Trends and Potential Applications
Advances in fast GC column technology and high-speed detectors will further shorten analysis times. Integration with automated sample preparation, two-dimensional GC, and high-resolution MS can extend this approach to emerging contaminants and complex matrices. Miniaturized and field-deployable fast GC-MS systems may enable on-site semivolatile monitoring.
Conclusion
This application demonstrates that fast GC-MS using a 20 m × 0.18 mm SLB-5ms column achieves full separation of EPA Method 8270D targets in under 8.5 minutes. The method meets regulatory performance criteria, boosts laboratory throughput, and offers a robust platform for routine semivolatile analysis.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms
2006|Merck|Applications
Application Report 391 Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms US EPA Method 8270D specifies a procedure for the determination of semivolatiles in solid wastes and ground waters. The suitability of the SLB-5ms for the separation of 72…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenebutylbenzyl, butylbenzylremainder, remaindernitrosodiphenylamine, nitrosodiphenylaminecarbazole, carbazolebenzidine, benzidinedibenzofuran, dibenzofuranazobenzene, azobenzenepropylamine, propylamineisophorone, isophoronehexachlorocyclopentadiene
Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms
2005|Merck|Applications
Application Report 351 Author: Katherine Stenerson Acquisition System: 6890 GC / 5973 MSD Notebook Reference: 1509-098 Analysis of US EPA Method 8270D Semivolatiles Using SLB-5ms US EPA Method 8270D specifies a procedure for the determination of semivolatiles in solid wastes…
Key words
benzo, benzophthalate, phthalatefluoranthene, fluoranthenebenzidine, benzidineexcellent, excellentpentachlorophenol, pentachlorophenolanthracene, anthracenepyrene, pyrenebutylbenzyl, butylbenzylremainder, remaindermillions, millionsnitrosodiphenylamine, nitrosodiphenylaminecarbazole, carbazoledibenzofuran, dibenzofuranazobenzene
Analysis of US EPA Method 8270D Semivolatiles Using a 30 m x 0.25 mm I.D., 0.50 μm SLB-5ms
2006|Merck|Applications
Application Report 399 Analysis of US EPA Method 8270D Semivolatiles Using a 30 m x 0.25 mm I.D., 0.50 µm SLB-5ms US EPA Method 8270D specifies a procedure for the determination of semivolatiles in solid waste and ground water samples.…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenebutylbenzyl, butylbenzylnitrosodiphenylamine, nitrosodiphenylaminecarbazole, carbazoleazobenzene, azobenzenedibenzofuran, dibenzofuranbenzidine, benzidinepropylamine, propylaminehexachlorocyclopentadiene, hexachlorocyclopentadieneisophorone, isophoronenitroso
Analysis of US EPA Method 8270D Semivolatiles Using a 30 m x 0.25 mm I.D., 0.50 μm SLB-5ms
2006|Merck|Applications
Application Report 401 Analysis of US EPA Method 8270D Semivolatiles Using a 30 m x 0.25 mm I.D., 0.50 µm SLB-5ms US EPA Method 8270D specifies a procedure for the determination of semivolatiles in solid waste and ground water samples.…
Key words
phthalate, phthalatebenzo, benzofluoranthene, fluorantheneanthracene, anthracenepyrene, pyrenebutylbenzyl, butylbenzylnitrosodiphenylamine, nitrosodiphenylaminecarbazole, carbazoleazobenzene, azobenzenedibenzofuran, dibenzofuranbenzidine, benzidinepropylamine, propylaminehexachlorocyclopentadiene, hexachlorocyclopentadieneisophorone, isophoronenitroso