Chlorinated pesticides - Analysis of EPA 625 pesticides with direct injection
Applications | 2011 | Agilent TechnologiesInstrumentation
This application note addresses the reliable analysis of trace-level chlorinated pesticides using direct on-column injection in gas chromatography. Chlorinated pesticides remain environmental contaminants of high concern due to their persistence, bioaccumulation and toxicological effects. Accurate, reproducible analysis at picogram levels is essential for regulatory compliance, environmental monitoring and public health protection.
The main goal of the study was to demonstrate a robust GC method for simultaneous determination of 16 EPA-625 priority chlorinated pesticides. The note provides an optimized protocol for direct injection, minimizing analyte decomposition and adsorption often encountered with hot-split injection techniques.
Two microliters of pesticide standard solution (200 pg/μL) were injected directly onto the column. The temperature program began at 120 °C (1 min), ramped at 9 °C/min to 300 °C, and held for the remainder of analysis. This approach ensured low elution temperatures for early-eluting components and resolved closely eluting isomers without thermal degradation.
The method achieved clear separation of 16 target analytes, including α-, β-, γ- and δ-benzene hexachloride, heptachlor, aldrin, dieldrin, endrin, DDT and its metabolites, and endosulfan isomers. Baseline resolution was observed for critical pairs such as γ-BHC versus heptachlor. The direct injection approach reduced sample handling and improved precision across replicate runs.
Advances in detector technology, such as tandem mass spectrometry coupling, could further improve selectivity and quantitation limits. Automated on-column injection systems and micro-fabricated GC columns promise higher throughput and lower solvent consumption. Integration with sample-preconcentration techniques may enable multi-class contaminant screening in a single run.
The direct on-column GC-FID method on a wide-bore VF-5ms column offers a reliable, high-throughput solution for analysis of EPA-625 chlorinated pesticides. Its robustness and sensitivity make it well suited for environmental compliance and academic research focused on persistent organic pollutants.
GC, GC columns, Consumables
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Significance of the Topic
This application note addresses the reliable analysis of trace-level chlorinated pesticides using direct on-column injection in gas chromatography. Chlorinated pesticides remain environmental contaminants of high concern due to their persistence, bioaccumulation and toxicological effects. Accurate, reproducible analysis at picogram levels is essential for regulatory compliance, environmental monitoring and public health protection.
Objectives and Overview of the Study
The main goal of the study was to demonstrate a robust GC method for simultaneous determination of 16 EPA-625 priority chlorinated pesticides. The note provides an optimized protocol for direct injection, minimizing analyte decomposition and adsorption often encountered with hot-split injection techniques.
Instrumentation Used
- Gas chromatograph equipped with on-column injector.
- Agilent VF-5ms fused-silica capillary column (0.53 mm id × 30 m, 0.5 μm film thickness).
- Flame ionization detector (FID) operated at 300 °C.
- Helium carrier gas at a linear velocity of 50 cm/s.
Methodology
Two microliters of pesticide standard solution (200 pg/μL) were injected directly onto the column. The temperature program began at 120 °C (1 min), ramped at 9 °C/min to 300 °C, and held for the remainder of analysis. This approach ensured low elution temperatures for early-eluting components and resolved closely eluting isomers without thermal degradation.
Main Results and Discussion
The method achieved clear separation of 16 target analytes, including α-, β-, γ- and δ-benzene hexachloride, heptachlor, aldrin, dieldrin, endrin, DDT and its metabolites, and endosulfan isomers. Baseline resolution was observed for critical pairs such as γ-BHC versus heptachlor. The direct injection approach reduced sample handling and improved precision across replicate runs.
Benefits and Practical Applications
- Enhanced sensitivity for trace-level pesticide monitoring in water, soil and biological samples.
- Reduced risk of thermal decomposition, ensuring accurate quantification of labile analytes.
- Streamlined workflow suitable for routine environmental and quality-control laboratories.
Future Trends and Potential Applications
Advances in detector technology, such as tandem mass spectrometry coupling, could further improve selectivity and quantitation limits. Automated on-column injection systems and micro-fabricated GC columns promise higher throughput and lower solvent consumption. Integration with sample-preconcentration techniques may enable multi-class contaminant screening in a single run.
Conclusion
The direct on-column GC-FID method on a wide-bore VF-5ms column offers a reliable, high-throughput solution for analysis of EPA-625 chlorinated pesticides. Its robustness and sensitivity make it well suited for environmental compliance and academic research focused on persistent organic pollutants.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Halogenated pesticides
2011|Agilent Technologies|Applications
Halogenated pesticides Application Note Environmental Authors Introduction Agilent Technologies, Inc. GC Analysis of halogenated pesticides uses direct injection and the stabilized 50% phenyl PDMS of an Agilent VF-17ms column. Conditions Technique : GC Column : Agilent VF-17ms, 0.53 mm x…
Key words
halogenated, halogenatedcourtesy, courtesypesticides, pesticidesprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak
Bulletin 841C Chlorinated Pesticides Analysis by Capillary GC: Alternative Approaches We manufacture and test SPB-608 capillary columns specifically for analyses of chlorinated pesticides. 30m x 0.25mm ID columns can be used in capillary GC systems; 15m x 0.53mm ID columns…
Key words
bhc, bhcendosulfan, endosulfanendrin, endrinheptachlor, heptachlorpesticides, pesticidesepoxide, epoxidedieldrin, dieldrinaldrin, aldrinaldehyde, aldehydesupelco, supelcosulfate, sulfateextraction, extractionchlordane, chlordaneenvi, envimethoxychlor
Organochlorine Pesticides Analysis in Water by GC/ECD - Consumable workflow ordering guide
2021|Agilent Technologies|Brochures and specifications
Organochlorine Pesticides Analysis in Water by GC/ECD Consumable workflow ordering guide Confidently Identify and Quantify Organochlorine Pesticides Organochlorine pesticides are common target analytes because they persist in soil or sediment, which affects water source quality. They are often measured following…
Key words
organochlorine, organochlorinepesticides, pesticidespesticide, pesticidesurrogate, surrogatestandard, standardmix, mixsupplies, suppliestoxaphene, toxaphenepath, pathmin, minbhc, bhcbreakdown, breakdownchlordane, chlordaneendrin, endrinagilent
Organochlorine pesticides
2011|Agilent Technologies|Applications
Organochlorine pesticides Application Note Environmental Authors Introduction Agilent Technologies, Inc. Sixteen organochlorine pesticides are analyzed by GC using the stabilized 50% phenyl PDMS phase of Agilent VF-17ms in less than 27 minutes. Conditions Technique : GC Column : Agilent VF-17ms,…
Key words
organochlorine, organochlorinecourtesy, courtesypesticides, pesticidesprinted, printedinjector, injectorusa, usaauthors, authorsenvironmental, environmentalsize, sizetechnique, techniquecarrier, carrieridentification, identificationdetector, detectornote, notepeak