US EPA Method 8081 Chlorinated Pesticides on Equity-1701

Applications | 2003 | MerckInstrumentation
GC, GC columns, Consumables
Industries
Environmental
Manufacturer
Merck

Summary

Significance of the Topic


The analysis of organochlorine pesticides is critical for environmental monitoring and regulatory compliance. US EPA Method 8081 is widely adopted for its sensitivity and specificity in detecting chlorinated pesticide residues in various matrices. Utilizing dual columns enhances result confidence and helps meet stringent quality control requirements.

Objectives and Study Overview


This application report evaluates the performance of an Equity-1701 capillary column for the separation of 20 target pesticides and two surrogate compounds as specified in EPA Method 8081. The study compares its elution profile to that of an Equity-5 column of similar dimensions and explores chromatographic conditions optimized for dual-column confirmation, focusing on the resolution of challenging analyte pairs.

Methodology


  • Analytes: 20 chlorinated pesticides plus two surrogate standards (50 ppb each).
  • Column dimensions: 30 m length, 0.25 mm ID, 0.25 µm film thickness.
  • Oven program: hold at 100 °C for 2 min; ramp to 160 °C at 15 °C/min; ramp to 280 °C at 5 °C/min; hold for 10 min.
  • Injection: 2 µL splitless (0.5 min) at 225 °C using a 4 mm ID double-taper liner.
  • Carrier gas: helium, constant flow at 30 cm/s at 100 °C.
  • Detector: electron capture detector at 310 °C.

Instrumentation


  • Gas chromatograph: GC 1777 system with dual injection ports.
  • Columns: Equity-1701 (confirmation) and Equity-5 (primary) of identical geometry.

Main Results and Discussion


All 22 compounds were fully resolved on the Equity-1701 column under the specified conditions, with a distinct elution order compared to the Equity-5 column, demonstrating its suitability as a confirmation column. Adjusting the carrier gas linear velocity in a dual-column configuration further improved the separation of closely eluting endosulfan sulfate and methoxychlor peaks, addressing a common analytical challenge.

Benefits and Practical Applications of the Method


  • Enhanced confidence through dual-column confirmation without changing run conditions.
  • Comprehensive separation of organochlorine pesticides suitable for QA/QC laboratories.
  • Streamlined workflow by using equivalent column geometries, reducing method development time.

Future Trends and Applications


Continued advances in column technology and detector sensitivity may allow faster run times and lower detection limits. Integration with mass spectrometry and automated sample handling can broaden applicability to complex environmental and food matrices. Custom dual-column configurations will likely become standard practice for regulatory compliance and research applications.

Conclusion


The Equity-1701 column effectively separates all EPA 8081 target pesticides and surrogates, providing a reliable confirmation option alongside the Equity-5 column. Its consistent performance and ease of implementation support robust monitoring of chlorinated pesticide residues.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
US EPA Method 8081 Chlorinated Pesticides on Equity-5
Application Report 26 Author: J. Walbridge/K. Kiefer Raw Data File Name: US EPA Method 8081 Chlorinated Pesticides on Equity-5 Acquisition System: GC1777 The potential health effects associated with many chlorinated pesticides has led to regulations regarding their use. US EPA…
Key words
bhc, bhcendrin, endrinendosulfan, endosulfanchlordane, chlordaneheptachlor, heptachlorids, idsmethoxychlor, methoxychloraldehyde, aldehydeepoxide, epoxidedieldrin, dieldrinaldrin, aldrinketone, ketonesulfate, sulfatewords, wordskey
Equity™ Capillary GC Columns
Equity™ Capillary GC Columns
2003|Merck|Brochures and specifications
SUPELCO EQUITY CAPILLARY GC COLUMNS Equity™ Capillary GC Columns The performance you demand... the service you deserve... from the company you trust. Updated with: ● NEW Equity-1701 columns ● New Applications 1 sigma-aldrich.com/supelco SUPELCO EQUITY CAPILLARY GC COLUMNS The performance…
Key words
equity, equitysupelco, supelcocapillary, capillarymin, mincolumns, columnscapseal, capsealexcellent, excellentbullet, bulletresponse, responsebleed, bleedshapes, shapesacetate, acetatemethyl, methylendosulfan, endosulfansupeltex
US EPA Method 608, 8081, OLM04.2 Organochlorine Pesticides on the SPB-608
Application Report 388 Author: Katherine Stenerson Acquisition System: 6249 Notebook Reference: 1569-016 US EPA Method 608, 8081, OLM04.2 Organochlorine Pesticides on the SPB-608 This is the analysis of a 22-component standard containing 20 pesticides and 2 surrogate compounds commonly analyzed…
Key words
bhc, bhcendrin, endrinendosulfan, endosulfanchlordane, chlordaneheptachlor, heptachlorids, idsmethoxychlor, methoxychloraldehyde, aldehydeepoxide, epoxidedieldrin, dieldrinaldrin, aldrinketone, ketonehexane, hexanesulfate, sulfatewords
US EPA Method 608, 8081, OLM04.2 Organochlorine Pesticides on the 20 m x 0.18 mm I.D., 0.36 μm SLB-5ms
Application Report 404 US EPA Method 608, 8081, OLM04.2 Organochlorine Pesticides on the 20 m x 0.18 mm I.D., 0.36 µm SLB-5ms Author: Katherine Stenerson Acquisition System: 9499 Notebook Reference: 1569-040 This is the analysis of a 22-component standard containing…
Key words
bhc, bhcendrin, endrinendosulfan, endosulfanchlordane, chlordaneheptachlor, heptachlorids, idsmethoxychlor, methoxychloraldehyde, aldehydeepoxide, epoxidedieldrin, dieldrinaldrin, aldrinketone, ketonesulfate, sulfatewords, wordskey
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike