US EPA Method 608, 8081, OLM04.2 Organochlorine Pesticides on the 20 m x 0.18 mm I.D., 0.36 μm SLB-5ms

Applications | 2006 | MerckInstrumentation
GC, GC columns, Consumables
Industries
Environmental
Manufacturer
Merck

Summary

Significance of the Topic


Analysis of organochlorine pesticides is essential due to their persistence, bioaccumulation potential, and toxicity. Reliable detection ensures environmental safety, food quality, and regulatory compliance.

Objectives and Study Overview


  • Demonstrate rapid separation of 20 pesticides and 2 surrogates under US EPA Methods 8081 and 608.
  • Explore efficiency gains using a narrow-bore SLB-5ms column (20 m x 0.18 mm ID, 0.36 µm).
  • Reduce analysis time while maintaining baseline resolution.

Methodology and Instrumentation


  • Column: SLB-5ms, 20 m x 0.18 mm ID, 0.36 µm.
  • Oven program: 100 °C for 2 min, ramp at 15 °C/min to 325 °C, hold 3 min.
  • Injection: 1 µL splitless at 250 °C with 0.75 min splitless time.
  • Carrier gas: Helium, 0.5 mL/min constant flow.
  • Detection: Micro-electron capture detector (µ-ECD) at 325 °C.
  • Sample: 22-component chlorinated pesticide standard at 50 ppb in n-hexane.

Main Results and Discussion


  • All 22 compounds were baseline-resolved in under 20 min, demonstrating high column efficiency.
  • Stable performance at 325 °C minimized elution times, particularly for late-eluting surrogates like decachlorobiphenyl.
  • Narrow bore column increased peak capacity and reduced analysis time compared to wider columns.

Benefits and Practical Applications


  • Faster throughput enables high sample load in environmental and food testing labs.
  • Enhanced resolution ensures accurate quantitation of trace-level pesticides.
  • Method aligns with established US EPA protocols, facilitating regulatory acceptance.

Future Trends and Applications


  • Integration with mass spectrometry for improved selectivity and confirmatory analysis.
  • Development of even smaller diameter columns and higher temperature-resistant phases.
  • Automation and on-site field deployable GC systems for rapid pesticide monitoring.

Conclusion


The study confirms that a 0.18 mm ID SLB-5ms column under optimized GC-ECD conditions provides rapid, high-resolution separation of organochlorine pesticides, supporting efficient environmental and food safety analyses.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
US EPA Method 608, 8081, OLM04.2 Organochlorine Pesticides on the SPB-608
Application Report 388 Author: Katherine Stenerson Acquisition System: 6249 Notebook Reference: 1569-016 US EPA Method 608, 8081, OLM04.2 Organochlorine Pesticides on the SPB-608 This is the analysis of a 22-component standard containing 20 pesticides and 2 surrogate compounds commonly analyzed…
Key words
bhc, bhcendrin, endrinendosulfan, endosulfanchlordane, chlordaneheptachlor, heptachlorids, idsmethoxychlor, methoxychloraldehyde, aldehydeepoxide, epoxidedieldrin, dieldrinaldrin, aldrinketone, ketonehexane, hexanesulfate, sulfatewords
US EPA Method 608, 8081, OLM04.2 Organochlorine Pesticides on the SPB-608
Application Report 389 Author: Katherine Stenerson Acquisition System: 6249 Notebook Reference: 1569-015 US EPA Method 608, 8081, OLM04.2 Organochlorine Pesticides on the SLB-5ms This is the analysis of a 22-component standard containing 20 pesticides and 2 surrogate compounds commonly analyzed…
Key words
bhc, bhcendrin, endrinendosulfan, endosulfanchlordane, chlordaneheptachlor, heptachlorids, idsmethoxychlor, methoxychloraldehyde, aldehydeepoxide, epoxidedieldrin, dieldrinaldrin, aldrinketone, ketonehexane, hexanesulfate, sulfatewords
Fast GC - Increase GC Speed Without Sacrificing Resolution
Fast GC - Increase GC Speed Without Sacrificing Resolution
2015|Merck|Brochures and specifications
Fast GC Increase GC Speed Without Sacrificing Resolution Increase Productivity Utilize Resources More Effectively Achieve Quicker Turn-Around Times Analyze More Samples Fast GC | Increase GC Speed Without Sacrificing Resolution Overview Shorter analysis times allow increased sample throughput, which translates…
Key words
sacrificing, sacrificingfast, fastmin, mintrifluoromethylsulfonyl, trifluoromethylsulfonylincrease, increasebenzo, benzoimide, imidephthalate, phthalateresolution, resolutionspeed, speedfluoranthene, fluoranthenelimonene, limonenebhc, bhcbis, bisneral
US EPA Method 8081 Chlorinated Pesticides on Equity-1701
Application Report 27 Author: J. Walbridge/K. Kiefer US EPA Method 8081 Chlorinated Pesticides on Equity-1701 Raw Data File Name: 1701\211BC015.D US EPA method 8081 describes the GC/ECD analysis of organochlorine pesticides and requires the use of a primary and confirmation…
Key words
bhc, bhcendrin, endrinendosulfan, endosulfanchlordane, chlordaneheptachlor, heptachlorids, idsmethoxychlor, methoxychloraldehyde, aldehydeepoxide, epoxidedieldrin, dieldrinaldrin, aldrinketone, ketonesulfate, sulfatewords, wordskey
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike