ANALYSIS OF A FORMALDEHYDE SCAVENGING COMPOUND USING SELECTED ION FLOW TUBE MASS SPECTROMETRY (SIFT-MS)
Applications | | AnatuneInstrumentation
Formaldehyde is a ubiquitous and toxic compound with recognized carcinogenicity. It is released from numerous indoor and industrial sources, including adhesives, fuels, and combustion processes. Reliable real-time measurement and effective removal of formaldehyde are critical to protecting human health and complying with regulatory limits.
This study evaluates a novel cage-based scavenger material designed to selectively capture formaldehyde from air streams, even in the presence of other volatile organic compounds (VOCs) and moisture. Selected Ion Flow Tube Mass Spectrometry (SIFT-MS) is employed to monitor formaldehyde uptake in real time, enabling determination of removal capacity and breakthrough behavior.
Samples of the cage material (10–100 mg) were packed into thermal desorption tubes and exposed to controlled gas mixtures in Tedlar bags. Dry standards of formaldehyde, mixtures of similar aldehydes (acetaldehyde, acrolein), non-polar VOCs (BTEX+), and “megamix” blends were used to test selectivity. Wet standards were prepared by adding water to evaluate performance under humid conditions. SIFT-MS continuously sampled at 20 mL/min over 45 minutes, with concentration profiles recorded to identify initial uptake and breakthrough points.
Syft Technologies Voice 200ultra SIFT-MS equipped with LabSyft™ software (v1.7.1) was used. The system generates reagent ions (H3O+, NO+, O2+, OH–, O2–, O–, NO2–, NO3–) via a microwave discharge source, selects them with a quadrupole filter, and quantifies product ions in a second mass analyzer. Helium was used as carrier gas with a high-precision inlet.
The cage material demonstrated complete selectivity for formaldehyde over similar aldehydes and a range of non-polar VOCs. Removal efficiency increased with sample mass, extending breakthrough times from minutes (10 mg) to over 45 minutes (100 mg). In humid tests, water alone sequestered some formaldehyde, but the cage material consistently achieved greater removal, reducing gas‐phase levels to near zero. Performance was unaffected by co-existing compounds or moisture, confirming the material’s robustness.
Advances may include scale-up of cage materials for continuous air purification systems, integration with online SIFT-MS analyzers for process feedback, and development of tailored scavengers for other problematic VOCs. Coupling real-time analytics with smart control algorithms could further enhance indoor air quality management.
The novel cage-based scavenger exhibits exceptional, selective formaldehyde removal in complex gas mixtures and humid environments. Combined with SIFT-MS analysis, this approach offers a powerful tool for continuous, real-time monitoring and control of formaldehyde exposure, supporting regulatory compliance and health protection.
SIFT-MS
IndustriesEnvironmental
ManufacturerSyft Technologies
Summary
Significance of the Topic
Formaldehyde is a ubiquitous and toxic compound with recognized carcinogenicity. It is released from numerous indoor and industrial sources, including adhesives, fuels, and combustion processes. Reliable real-time measurement and effective removal of formaldehyde are critical to protecting human health and complying with regulatory limits.
Objectives and Study Overview
This study evaluates a novel cage-based scavenger material designed to selectively capture formaldehyde from air streams, even in the presence of other volatile organic compounds (VOCs) and moisture. Selected Ion Flow Tube Mass Spectrometry (SIFT-MS) is employed to monitor formaldehyde uptake in real time, enabling determination of removal capacity and breakthrough behavior.
Methodology
Samples of the cage material (10–100 mg) were packed into thermal desorption tubes and exposed to controlled gas mixtures in Tedlar bags. Dry standards of formaldehyde, mixtures of similar aldehydes (acetaldehyde, acrolein), non-polar VOCs (BTEX+), and “megamix” blends were used to test selectivity. Wet standards were prepared by adding water to evaluate performance under humid conditions. SIFT-MS continuously sampled at 20 mL/min over 45 minutes, with concentration profiles recorded to identify initial uptake and breakthrough points.
Instrumentation Used
Syft Technologies Voice 200ultra SIFT-MS equipped with LabSyft™ software (v1.7.1) was used. The system generates reagent ions (H3O+, NO+, O2+, OH–, O2–, O–, NO2–, NO3–) via a microwave discharge source, selects them with a quadrupole filter, and quantifies product ions in a second mass analyzer. Helium was used as carrier gas with a high-precision inlet.
Main Results and Discussion
The cage material demonstrated complete selectivity for formaldehyde over similar aldehydes and a range of non-polar VOCs. Removal efficiency increased with sample mass, extending breakthrough times from minutes (10 mg) to over 45 minutes (100 mg). In humid tests, water alone sequestered some formaldehyde, but the cage material consistently achieved greater removal, reducing gas‐phase levels to near zero. Performance was unaffected by co-existing compounds or moisture, confirming the material’s robustness.
Benefits and Practical Applications
- Real-time monitoring and control of formaldehyde emissions without derivatization or chromatography.
- High sensitivity (pptv) and selectivity for workplace air monitoring, environmental surveys, and industrial process control.
- Quantitative assessment of scavenger capacity and breakthrough for optimized filter design and replacement scheduling.
Future Trends and Potential Applications
Advances may include scale-up of cage materials for continuous air purification systems, integration with online SIFT-MS analyzers for process feedback, and development of tailored scavengers for other problematic VOCs. Coupling real-time analytics with smart control algorithms could further enhance indoor air quality management.
Conclusion
The novel cage-based scavenger exhibits exceptional, selective formaldehyde removal in complex gas mixtures and humid environments. Combined with SIFT-MS analysis, this approach offers a powerful tool for continuous, real-time monitoring and control of formaldehyde exposure, supporting regulatory compliance and health protection.
References
- US Environmental Protection Agency (EPA). IRIS Toxicological Review of Formaldehyde. 2019.
- Perkins M., Anatune Ltd. Analysis of a Formaldehyde Scavenging Compound Using Selected Ion Flow Tube Mass Spectrometry. Application Note, 2019.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Automated Selected Ion Flow Tube Mass Spectrometry (SIFT-MS)
2017|GERSTEL|Technical notes
Anatune Ltd Unit 4, Wellbrook Court, Girton Road, Cambridge, CB3 0NA, UK Tel: +44 (0) 1223279210 Fax: +44 (0) 1223279253 Email: [email protected] Internet: www.anatune.co.uk Copyright © 2017 Anatune Ltd. All Rights Reserved. Anatune is a trademark of Anatune Ltd. Automated…
Key words
headspace, headspaceppmv, ppmvsift, siftstatic, staticanalysis, analysisconcentration, concentrationintercept, interceptaqueous, aqueousslope, slopesyringe, syringemeasured, measuredgerstel, gersteltemperature, temperatureworth, worthvocs
Preliminary Investigation of a Range of Selected Ion Flow Tube Mass Spectrometry (SIFT-MS) Applications
2015|GERSTEL|Technical notes
Anatune Ltd Unit 4, Wellbrook Court, Girton Road, Cambridge, CB3 0NA, UK Tel: +44 (0) 1223279210 Fax: +44 (0) 1223279253 Email: [email protected] Internet: www.anatune.co.uk Copyright © 2018 Anatune Ltd. All Rights Reserved. Anatune is a trademark of Anatune Ltd. Preliminary…
Key words
sift, siftconcentration, concentrationpreliminary, preliminaryreagent, reagentppbv, ppbvmeasurements, measurementsair, airxylenes, xylenesrelease, releaseselected, selectedvehicle, vehicleanalysed, analysedvial, vialtube, tubereaction
Analysis of Vehicle Interior Air Quality (VIAQ) using SIFT-MS
2017|Syft Technologies|Applications
Anatune Ltd Unit 4, Wellbrook Court, Girton Road, Cambridge, CB3 0NA, UK Tel: +44 (0) 1223279210 Fax: +44 (0) 1223279253 Email: [email protected] Internet: www.anatune.co.uk Copyright © 2017 Anatune Ltd. All Rights Reserved. Anatune is a trademark of Anatune Ltd. Analysis…
Key words
sift, siftacetaldehyde, acetaldehydeacrolein, acroleinlimit, limitviaq, viaqinterior, interiorformaldehyde, formaldehydeacetone, acetoneset, setbutanal, butanalpropanal, propanalpentanal, pentanalethylbenzene, ethylbenzenestyrene, styrenetube
HIGH-SPEED FORMALDEHYDE ANALYSIS FOR THE PROCESS-LINE AND LABORATORY: SIFT-MS
|GERSTEL|Applications
AS201 Wellbrook Court | Girton Road | Cambridge | CB3 0NA | | tel: +44 (0) 1223 279210 | fax: +44 (0) 1223 279253 [email protected] | anatune.co.uk HIGH-SPEED FORMALDEHYDE ANALYSIS FOR THE PROCESS-LINE AND LABORATORY: SIFT-MS V.S. Langford,1 M. Perkins,2…
Key words
formaldehyde, formaldehydesift, siftheadspace, headspacemonomer, monomerconcentration, concentrationfrom, frommhe, mhemass, masstube, tubetechnique, techniquedifferent, differentresidual, residualhigh, highgas, gasconcentrations