HIGH-SPEED FORMALDEHYDE ANALYSIS FOR THE PROCESS-LINE AND LABORATORY: SIFT-MS

Applications |  | AnatuneInstrumentation
HeadSpace, SIFT-MS
Industries
Environmental
Manufacturer
GERSTEL, Syft Technologies

Summary

Significance of the Topic


Formaldehyde is a high production volume chemical involved in resins adhesives and combustion byproducts affecting air quality and health. Its polar reactive nature makes trace analysis challenging for traditional chromatography based methods requiring derivatization and long sample preparation times. Rapid direct quantitation is needed for process monitoring environmental control and occupational safety.

Objectives and Study Overview


This work demonstrates high speed formaldehyde quantitation using selected ion flow tube mass spectrometry SIFTMS. The study examines real time direct air and headspace measurement scenarios compares performance against conventional chromatographic analysis and illustrates automated multiple headspace extraction workflows on polymer samples.

Methodology and Instrumentation


Soft chemical ionization using reagent ions coupled with quadrupole mass spectrometry provides selective reactions without a chromatographic stage. Real time sub part per billion detection is achieved in milliseconds across a wide linear range spanning six orders of magnitude.

Used Instrumentation


  • SIFTMS Voice200 and Voice200ultra instruments by Syft Technologies
  • GERSTEL Multipurpose Sampler with gas bag analysis accessory and headspace vial rack

Main Results and Discussion


  • Real time detection of 500 ppb formaldehyde achieved within seconds with repeatability below three percent
  • Automated analysis of shipping container emissions provided clear safe unsafe alerts for formaldehyde levels
  • Simultaneous quantitation of six saturated aldehydes from acetaldehyde to hexanal in under one minute
  • Rapid headspace screening of ground polymers revealed residual formaldehyde with sample throughput 25 times faster than HPLC
  • Multiple headspace extraction on polyoxymethylene yielded a total residual formaldehyde concentration of 13.8 ppm using sequential measurements

Benefits and Practical Applications


  • No derivatization or preconcentration required for direct gas or headspace analysis
  • Push button operation allows use by nonexpert personnel in field industrial and port environments
  • Wide dynamic range from ppt to ppm supports diverse sample types
  • Low operational consumable costs compared to conventional chromatography
  • Real time continuous monitoring supports process analytical technology and environmental applications

Future Trends and Potential Applications


Ongoing enhancements in reagent ion selection and data algorithms will improve selectivity in complex matrices. Integration into online monitoring networks and advanced process control platforms is anticipated. Automation and multi ion strategies will broaden use in environmental health safety and industrial quality control settings.

Conclusion


SIFTMS provides a breakthrough for rapid direct formaldehyde analysis overcoming limitations of traditional methods. Its speed sensitivity selectivity and ease of automation position it as a versatile tool for industrial environmental and research applications.

References


  • United States Environmental Protection Agency IRIS Toxicological Review of Formaldehyde
  • Langford V S et al Comparison of chromatographic methods and SIFTMS for formaldehyde analysis
  • Smith D and Spanel P Selected Ion Flow Tube Mass Spectrometry for on line trace gas analysis

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
HIGH-THROUGHPUT RESIDUAL SOLVENT AND RESIDUAL MONOMER ANALYSIS USING SELECTED ION FLOW TUBE MASS SPECTROMETRY
AS208 Wellbrook Court | Girton Road | Cambridge | CB3 0NA | | tel: +44 (0) 1223 279210 | fax: +44 (0) 1223 279253 | email: [email protected] | anatune.co.uk HIGH-THROUGHPUT RESIDUAL SOLVENT AND RESIDUAL MONOMER ANALYSIS USING SELECTED ION FLOW…
Key words
headspace, headspaceppmv, ppmvformaldehyde, formaldehydemonomer, monomerresidual, residualmhe, mhesift, siftanalysis, analysismarkedly, markedlypackaging, packagingmultiple, multipleconcentration, concentrationsolvents, solventsopportunities, opportunitiestechnique
HIGH-THROUGHPUT VOC AND INORGANIC GAS ANALYSIS: AUTOMATED SIFT-MS
AS216 Wellbrook Court | Girton Road | Cambridge | CB3 0NA | | tel: +44 (0) 1223 279210 | fax: +44 (0) 1223 279253 | email: [email protected] | anatune.co.uk HIGH-THROUGHPUT VOC AND INORGANIC GAS ANALYSIS: AUTOMATED SIFT-MS Mark J. Perkins,…
Key words
sift, siftmhe, mheheadspace, headspacesha, shaanalysis, analysisrapid, rapidpom, pomresidual, residualformaldehyde, formaldehydeoffers, offerseffective, effectiveautomated, automatedpolyoxymethylene, polyoxymethyleneprescreening, prescreeningrapidly
AUTOMATION APPLICATIONS - SYFT TECHNOLOGIES
AUTOMATION APPLICATIONS - SYFT TECHNOLOGIES
2017|Syft Technologies|Brochures and specifications
AUTOMATION APPLICATIONS SYFT TECHNOLOGIES AUTOMATION APPLICATIONS OF SIFT-MS Automation has been widely utilized by traditional chromatographic techniques for many years, enabling instrumentation to be productive 24/7 – a critical consideration because sample throughput is typically only 20 to 60 samples…
Key words
sift, siftsyft, syftheadspace, headspaceanalysis, analysisvocs, vocsautomation, automationmhe, mhereagent, reagenttrillion, trillionsha, shavery, veryapplications, applicationsgases, gasesresidual, residualrapid
Automated Selected Ion Flow Tube Mass Spectrometry (SIFT-MS)
Anatune Ltd Unit 4, Wellbrook Court, Girton Road, Cambridge, CB3 0NA, UK Tel: +44 (0) 1223279210 Fax: +44 (0) 1223279253 Email: [email protected] Internet: www.anatune.co.uk Copyright © 2017 Anatune Ltd. All Rights Reserved. Anatune is a trademark of Anatune Ltd. Automated…
Key words
headspace, headspaceppmv, ppmvsift, siftstatic, staticanalysis, analysisconcentration, concentrationintercept, interceptaqueous, aqueousslope, slopesyringe, syringemeasured, measuredgerstel, gersteltemperature, temperaturevocs, vocsworth
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike