Automated Selected Ion Flow Tube Mass Spectrometry (SIFT-MS)

Technical notes | 2017 | AnatuneInstrumentation
SIFT-MS
Industries
Manufacturer
GERSTEL, Syft Technologies

Summary

Importance of the Topic


Automated Selected Ion Flow Tube Mass Spectrometry (SIFT-MS) addresses the need for real-time, high-sensitivity volatile organic compound (VOC) analysis without labor-intensive sample preparation. By combining soft chemical ionisation with automated headspace sampling, this approach accelerates laboratory throughput and enhances quality-control precision across environmental monitoring, polymer analytics, food and flavour studies, and water testing.

Objectives and Overview of the Study


This application note demonstrates the integration of a GERSTEL MultiPurpose Sampler with Syft Technologies’ Voice 200ultra SIFT-MS instrument. Key aims were to establish method repeatability, evaluate static headspace and multiple headspace extraction workflows, develop protocols for residual monomer and aqueous VOC analysis, and compare performance to conventional GC-based techniques.

Methodology


  • Headspace sampling employs slow, controlled syringe injections (25 mL/min make-up flow) to deliver gas-phase analytes into the flow tube under helium or nitrogen carrier gas.
  • Reagent ions (H3O+, NO+, O2+, OH–, etc.) are generated via microwave discharge and mass-filtered, reacting softly with VOCs to form product ions.
  • Analyte concentrations are derived from product-to-reagent ion ratios using known rate constants, enabling simultaneous multi-compound quantitation without chromatography.
  • Multiple experiments covered ambient static headspace, elevated-temperature headspace, residual monomer quantitation, multiple headspace extraction (MHE), method development (incubation time optimization), and aqueous BTEX and alcohol analyses.

Instrumentation


  • Voice 200ultra SIFT-MS with LabSyft software.
  • GERSTEL MPS xt Dual Head autosampler with 20 mL headspace tray, heated syringes (50–150 °C), agitator and purge tools.
  • Helium carrier gas; inlet maintained at 150 °C.

Key Results and Discussion


  • Repeatability: 50 injections from a Tedlar bag containing flavour and breath VOCs yielded RSDs of 0.7–5.6 % and 42 s cycle times.
  • Static Headspace Aldehyde Analysis: C1–C6 n-aldehyde and formaldehyde spikes measured every 30 s demonstrated rapid profiling of low-molecular polar VOCs.
  • Residual Monomer: Linear quantitation of formaldehyde release from polyoxymethylene (0–150 mg) with R² > 0.998; polymer samples run in under 6 min each, including incubation overlap.
  • Multiple Headspace Extraction (MHE): Exponential decay of THF release from polybutylene terephthalate over 10 cycles enabled total VOC determination via standard MHE equations.
  • Method Development: Styrene release from polystyrene plateaued at ≈12 min incubation, guiding optimum headspace parameters in 25 min for 10 samples versus 4.5 h by GC-MS.
  • Aqueous VOCs: BTEX and chloroform in water (0.1–20 nL/mL) and alcohols (ethanol, pentanol) at 40–60 °C exhibited excellent linearity (R² > 0.995) with 50 s acquisitions.

Benefits and Practical Applications of the Method


  • High throughput: Automated sequencing and PrepAhead overlap reduce instrument idle time.
  • No chromatography: Simplifies workflows and avoids carryover or matrix interferences.
  • Broad analyte scope: Simultaneous detection of polar and non-polar VOCs at ppt–ppb levels.
  • Flexible formats: Tedlar bags, vials, and aqueous solutions.
  • Quality by Design: Rapid residual monomer screening suits pharmaceutical, food-contact, and industrial QA/QC.

Future Trends and Possibilities for Use


As SIFT-MS automation matures, integration with robotic sample handling and in-line process monitoring will expand. Enhanced reagent-ion libraries and machine-learning calibration may further improve specificity. Potential applications include real-time environmental surveillance, on-site breath diagnostics, and continuous emissions monitoring.

Conclusion


The coupling of GERSTEL’s MPS with SIFT-MS yields a versatile automated platform for rapid, quantitative VOC analysis spanning headspace, polymer, and aqueous matrices. The approach surpasses traditional GC methods in speed and simplicity while maintaining robust analytical performance.

References


  • Agilent Technologies. "Multiple Headspace Extraction – Capillary Gas Chromatography for Quantitative Determination of Volatiles in Solid Matrices," Application Note, May 1996.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
SPECIATION OF DICHLOROMETHANE AND CHLOROFORM USING NEGATIVE ION SIFT-MS
AS233 Wellbrook Court | Girton Road | Cambridge | CB3 0NA | | tel: +44 (0) 1223 279210 | fax: +44 (0) 1223 279253 | email: [email protected] | anatune.co.uk SPECIATION OF DICHLOROMETHANE AND CHLOROFORM USING NEGATIVE ION SIFT-MS Mark Perkins,…
Key words
chlorofor, chloroforsift, siftchloroform, chloroformppmv, ppmvheadspace, headspacedichloromethane, dichloromethanebranching, branchinglinearity, linearityconcentratio, concentratiorounding, roundingsuites, suitesrepeatability, repeatabilityreaction, reactionion, ionconstants
HIGH-SPEED FORMALDEHYDE ANALYSIS FOR THE PROCESS-LINE AND LABORATORY: SIFT-MS
AS201 Wellbrook Court | Girton Road | Cambridge | CB3 0NA | | tel: +44 (0) 1223 279210 | fax: +44 (0) 1223 279253 [email protected] | anatune.co.uk HIGH-SPEED FORMALDEHYDE ANALYSIS FOR THE PROCESS-LINE AND LABORATORY: SIFT-MS V.S. Langford,1 M. Perkins,2…
Key words
formaldehyde, formaldehydesift, siftheadspace, headspacemonomer, monomerconcentration, concentrationfrom, frommhe, mhemass, masstube, tubetechnique, techniquedifferent, differentresidual, residualgas, gashigh, highinjection
HIGH-THROUGHPUT RESIDUAL SOLVENT AND RESIDUAL MONOMER ANALYSIS USING SELECTED ION FLOW TUBE MASS SPECTROMETRY
AS208 Wellbrook Court | Girton Road | Cambridge | CB3 0NA | | tel: +44 (0) 1223 279210 | fax: +44 (0) 1223 279253 | email: [email protected] | anatune.co.uk HIGH-THROUGHPUT RESIDUAL SOLVENT AND RESIDUAL MONOMER ANALYSIS USING SELECTED ION FLOW…
Key words
headspace, headspaceppmv, ppmvformaldehyde, formaldehydemonomer, monomerresidual, residualmhe, mhesift, siftanalysis, analysispackaging, packagingmarkedly, markedlymultiple, multipleconcentration, concentrationsolvents, solventsopportunities, opportunitiestechnique
ANALYSIS OF A FORMALDEHYDE SCAVENGING COMPOUND USING SELECTED ION FLOW TUBE MASS SPECTROMETRY (SIFT-MS)
AS223 Wellbrook Court | Girton Road | Cambridge | CB3 0NA | | tel: +44 (0) 1223 279210 | fax: +44 (0) 1223 279253 | email: [email protected] | anatune.co.uk ANALYSIS OF A FORMALDEHYDE SCAVENGING COMPOUND USING SELECTED ION FLOW TUBE…
Key words
formaldehyde, formaldehydeppmv, ppmvcage, cageactual, actuallowest, lowesttrimethylbenzenes, trimethylbenzenessift, siftpoint, pointend, endmaterial, materialmeasured, measuredconcentration, concentrationxylenes, xylenesacrolein, acroleinisooctane
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike