Carrier Gases in Capillary GC
Presentations | 2011 | Agilent TechnologiesInstrumentation
GC
IndustriesManufacturerAgilent Technologies
Summary
Importance of Carrier Gas Selection in Capillary GC
- Carrier gas type and flow conditions critically affect chromatographic efficiency, analysis time, and resolution.
- Inertness and purity of the gas ensure minimal interaction with analytes and stationary phase, avoiding baseline noise and contamination.
- Precise pressure and flow control are essential for reproducible retention times and peak shapes.
Objectives and Study Overview
- Compare common carrier gases (nitrogen, helium, hydrogen) in capillary gas chromatography.
- Examine the impact of gas properties on van Deemter behavior and optimal linear velocity.
- Evaluate effects of column dimensional tolerances on required head pressure and retention time stability.
- Demonstrate practical considerations for flow measurement, pressure programming, and retention time windows.
Methodology and Instrumentation Used
- Van Deemter analysis for packed and open-tubular columns to determine height equivalent to a theoretical plate (HETP) as a function of linear velocity for N2, He, H2.
- Comparison of constant pressure vs. constant flow modes, with calculation of optimal gas velocities (uopt) and optimal practical gas velocities (OPGV).
- Measurement tools: electronic pressure control (EPC), flow measuring inserts for FID, retention time of non-retained compounds (methane, butane, air, SF6, vinyl chloride).
- Evaluation of column dimensions (length and inner diameter tolerances) on head pressure using Agilent GC Pressure/Flow Calculator.
- Analysis of real samples (lavender oil spiked with camphor, linalool, linalyl acetate) under varied head pressures and temperature programs, detected by FID and MS.
Main Results and Discussion
- Optimal linear velocities: nitrogen 12–20 cm/s, helium 22–35 cm/s, hydrogen 35–60 cm/s; hydrogen provides highest efficiency and shortest run times.
- Pressure drop across columns increases sharply with decreasing inner diameter and increasing length; ±6 µm ID and ±0.5 m length tolerances can alter required head pressure by 20–35%.
- Constant pressure mode yields lower flow at higher oven temperatures, whereas constant flow mode requires increasing pressure with temperature.
- Variations in head pressure shift retention times and may reverse elution order for analytes with different vapor pressures and stationary phase interactions.
- Temperature programming and software lock steps impact retention time reproducibility and must be controlled in SOPs.
- Hydrogen use in GC–MS can lead to temporary hydrocarbon contamination due to its cleaning action; a 2–4 week conditioning period is typical.
Benefits and Practical Applications of the Method
- Hydrogen carrier gas offers faster separations and high throughput when safety measures are in place.
- Helium remains a safe compromise with good efficiency, widely used where hydrogen is restricted.
- Nitrogen may be cost-effective but leads to longer analysis times and reduced resolution.
- Understanding column dimension effects supports robust method transfer and routine QA/QC.
- Electronic pressure control and accurate flow measurement tools improve reproducibility across temperature programs.
Future Trends and Possibilities for Use
- Wider adoption of hydrogen with improved safety features and leak detection systems.
- Development of alternative carrier gases and gas blends to optimize selectivity and reduce cost.
- Advanced software for real-time flow and pressure modeling to compensate for column aging and dimensional drift.
- Integration of AI-driven method optimization for automated gas selection and parameter tuning.
Conclusion
Carrier gas selection, precise control of pressure and flow, and awareness of column dimensional tolerances are fundamental to high-performance capillary GC. Hydrogen offers the fastest separations but requires stringent safety controls; helium provides a reliable alternative, while nitrogen is limited by lower efficiency. Consistent methodology, including flow measurement, pressure programming, and retention time monitoring, ensures reproducible and robust analyses across laboratories.
Použitá instrumentace
- Agilent capillary GC systems with electronic pressure control (EPC).
- Detectors: FID, TCD, MS, ECD, NPD, PID, ELCD.
- FID flow measuring insert and red cap flow tools.
- Agilent GC Pressure/Flow Calculator software.
Reference
- Folk A. Carrier Gases in Capillary GC. Agilent Technologies Application Note, February 2011.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Secrets of GC Column Dimensions
2010|Agilent Technologies|Presentations
Secrets of GC Column Dimensions GC Columns and Consumables Mark Sinnott Application Engineer Folsom California March 12, 2010 The Secrets of GC Column Dimensions Agilent Restricted Page 1 Secrets of GC Column Dimensions •Do I have the right column phase?…
Key words
secrets, secretsrestricted, restricteddimensions, dimensionscolumn, columnagilent, agilentthickness, thicknessfilm, filmmegabore, megaborediameter, diameterlength, lengthincrease, increaseresolution, resolutionefficiency, efficiencycapacity, capacityopgv
Helium Conservation and Converting GC Methods to Nitrogen and Hydrogen Carrier Gas
2013|Agilent Technologies|Presentations
Addressing the World Helium Shortage For Gas Chromatography Helium Conservation and Converting GC Methods to Nitrogen and Hydrogen Carrier Gas James D. McCurry, Ph.D. Senior Scientist Agilent Technologies 1 July 30, 2013 Presentation Outline Carrier Gas Decision Tree • Decision…
Key words
helium, heliumhetp, hetpcarrier, carrierconservation, conservationgas, gasnitrogen, nitrogenrestricted, restrictedvelocity, velocityconsiderations, considerationsagilent, agilentepc, epcsafety, safetyhydrogen, hydrogenresolution, resolutioninlet
GC Method Developement
2015|Agilent Technologies|Presentations
GC Method Developement What to Consider The Sample Method of injection Inlet Detector Carrier Gas Column COMPOUND REQUIREMENTS FOR GC Only 10-20% of all compounds are suitable for GC analysis The compounds must have: ✓ ✓ Sufficient volatility Thermal stability…
Key words
restricted, restrictedtitle, titlepresentation, presentationliner, linergroup, groupinlet, inletagilent, agilentsample, sampleliners, linerssplit, splitcolumn, columnglass, glasssplitless, splitlessstrong, strongdeveloping
GC Method Development
2015|Agilent Technologies|Presentations
GC Method Development Simon Jones Chromatography Application Engineer Agilent Technologies, Inc. November 12, 2015 What to Consider The Sample Method of injection Inlet Detector Carrier Gas Column COMPOUND REQUIREMENTS FOR GC Only 10-20% of all compounds are suitable for GC…
Key words
restricted, restrictedtitle, titlepresentation, presentationsplit, splitsample, sampleliner, linergroup, groupagilent, agilentinlet, inletsplitless, splitlesscolumn, columnliners, linerspulsed, pulsedglass, glasslow