How to Keep your Application Running - Routine Maintenance
Presentations | 2013 | Agilent TechnologiesInstrumentation
Reliable operation of gas chromatography (GC) instruments is critical for generating accurate analytical data in research, quality control, and industrial settings. Routine maintenance and optimization of sample handling, injection systems, inlet components, columns, and detectors minimize downtime, prevent data variability, and extend instrument lifetime.
This document outlines best practices for preventive maintenance of GC systems. It covers consistent sample preparation, auto‐injector setup, inlet sealing and liner selection, column installation and conditioning, detector upkeep, troubleshooting of common problems, and guidelines to optimize chromatographic performance.
Key methodological elements include:
Comparative studies of inlet liners demonstrated that deactivated, tapered, or direct-connect liners yield better response factors and lower relative standard deviations for active analytes such as 2,4-dinitrophenol. Routine replacement of septa and liners prevented leaks and baseline disturbances. Proper column installation and conditioning dramatically reduced abnormal bleed and background noise. Troubleshooting exercises linked poor peak shapes, resolution loss, and baseline drift to contaminated liners or misaligned fittings rather than column faults.
Implementing these maintenance strategies enhances chromatographic performance by:
Advances in GC technology are focusing on ultra-inert and high-temperature stationary phases (e.g., sol-gel coatings), custom-designed columns for specific analyses, enhanced software tools for predictive maintenance, and integration of real-time flow and leak monitoring. These developments will further improve system reliability and analytical throughput.
Consistent sample handling, correct auto-injector and inlet component selection, precise column installation, and regular preventative maintenance form the foundation of reliable GC operation. Keeping detailed maintenance logs and replacing wear items proactively maximizes instrument uptime and analytical confidence.
GC, Consumables
IndustriesManufacturerAgilent Technologies
Summary
Importance of the Topic
Reliable operation of gas chromatography (GC) instruments is critical for generating accurate analytical data in research, quality control, and industrial settings. Routine maintenance and optimization of sample handling, injection systems, inlet components, columns, and detectors minimize downtime, prevent data variability, and extend instrument lifetime.
Study Objectives and Overview
This document outlines best practices for preventive maintenance of GC systems. It covers consistent sample preparation, auto‐injector setup, inlet sealing and liner selection, column installation and conditioning, detector upkeep, troubleshooting of common problems, and guidelines to optimize chromatographic performance.
Methodology and Instrumentation
Key methodological elements include:
- Sample Preparation: Control temperature, vial seal integrity, pH, solvent purity, and light exposure to ensure reproducible injections.
- Auto‐Injector Practices: Use appropriate syringe volume (5 µL vs. 10 µL), perform solvent and sample washes, adjust plunger speed, and set viscosity delays to avoid carryover.
- Inlet Components: Replace septa (BTO or Advanced Green) and O-rings regularly; select septa with CenterGuide to reduce coring; choose liners based on volume, deactivation, and special features (glass wool, taper, Jennings cup, direct‐connect).
- Column Installation: Use proper ferrules (graphite or graphite/Vespel), measure and cut capillary columns cleanly, tighten connections without overtightening, perform leak checks, set linear velocity or flow rate by measuring retention time of non-retained peaks, and condition columns with a temperature hold to stabilize bleed.
- Detector Maintenance: Monitor detector temperatures and gas flows; perform regular preventative maintenance of FID, ECD, NPD, TCD, or MSD to sustain sensitivity and baseline stability.
Instrumentation Used
- Agilent 6890 GC system equipped with FID detector
- HP-5MS capillary column (30 m × 0.25 mm ID, 0.5 µm film)
- Agilent Certified Vials and septa (BTO and Advanced Green)
- 5 µL Agilent HP-Point syringe
- Agilent GC Pressure/Flow/Vapor Volume Calculator software
Main Results and Discussion
Comparative studies of inlet liners demonstrated that deactivated, tapered, or direct-connect liners yield better response factors and lower relative standard deviations for active analytes such as 2,4-dinitrophenol. Routine replacement of septa and liners prevented leaks and baseline disturbances. Proper column installation and conditioning dramatically reduced abnormal bleed and background noise. Troubleshooting exercises linked poor peak shapes, resolution loss, and baseline drift to contaminated liners or misaligned fittings rather than column faults.
Benefits and Practical Applications
Implementing these maintenance strategies enhances chromatographic performance by:
- Improving peak shape, reproducibility, and sensitivity
- Reducing unplanned downtime and repair costs
- Extending column and inlet component lifetimes
- Ensuring consistent data quality for QA/QC and research workflows
Future Trends and Applications
Advances in GC technology are focusing on ultra-inert and high-temperature stationary phases (e.g., sol-gel coatings), custom-designed columns for specific analyses, enhanced software tools for predictive maintenance, and integration of real-time flow and leak monitoring. These developments will further improve system reliability and analytical throughput.
Conclusion
Consistent sample handling, correct auto-injector and inlet component selection, precise column installation, and regular preventative maintenance form the foundation of reliable GC operation. Keeping detailed maintenance logs and replacing wear items proactively maximizes instrument uptime and analytical confidence.
References
- Dean Rood, A Practical Guide to the Care, Maintenance, and Troubleshooting of Capillary GC Systems, 3rd ed., Wiley-VCH, New York, 2001.
- Agilent Technologies, Improvements in the Agilent Cool On-Column-FID Injection of 11 Analyte Test Mix for USEPA Method 8270, Application Note 5988-3072EN.
- Agilent GC Pressure/Flow/Vapor Volume Calculator, online tool.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Tips on Making your GC System and Analysis more Robust
2015|Agilent Technologies|Presentations
Tips on Making your GC System and Analysis more Robust Agilent Restricted Primary Areas of Concern • Sample • Auto-Injector • Inlet • Column • Detector Agilent Restricted Sample Preparation and Care It is critical that the sample extract be…
Key words
restricted, restrictedagilent, agilenttitle, titlepresentation, presentationcolumn, columngroup, groupliner, linerinlet, inletliners, linersleak, leaksplit, splitsepta, septainjector, injectorbleed, bleedinstallation
Maintaining Your Agilent GC’s Split/Splitless Injection Port
|Agilent Technologies|Presentations
Maintaining Your Agilent GC’s GC s Split/Splitless Injection Port What’s going on in there anyway & how do I stay y in control? Group/Presentation Title Agilent Restricted Month ##, 200X Presentation Expectations •The #1 inlet used by far is the…
Key words
presentation, presentationtitle, titlerestricted, restrictedgroup, groupagilent, agilentsplit, splitliner, linerwool, woolsplitless, splitlessglass, glassinlet, inletliners, linersdiscrimination, discriminationseptum, septumvent
Around the GC System in 60 minutes
2018|Agilent Technologies|Presentations
Around the GC System in 60 minutes Title Slide with Image Arial: 28 pt Title Case Subtitle Arial: 21 pt sentence case September 13, 2018 Presenter’s Name Arial: 18 pt Presenter’s Title Arial: 16 pt Manually Formatted Date Arial: 16…
Key words
restricted, restrictedagilent, agilentliner, linercolumn, columnpress, pressferrules, ferrulesleak, leakinlet, inletseal, sealsepta, septalipid, lipidtitle, titlebleed, bleedgraphite, graphitetightening
GC Method Developement
2015|Agilent Technologies|Presentations
GC Method Developement What to Consider The Sample Method of injection Inlet Detector Carrier Gas Column COMPOUND REQUIREMENTS FOR GC Only 10-20% of all compounds are suitable for GC analysis The compounds must have: ✓ ✓ Sufficient volatility Thermal stability…
Key words
restricted, restrictedtitle, titlepresentation, presentationliner, linergroup, groupinlet, inletagilent, agilentsample, sampleliners, linerssplit, splitcolumn, columnglass, glasssplitless, splitlessstrong, strongdeveloping