Around the GC System in 60 minutes
Presentations | 2018 | Agilent TechnologiesInstrumentation
Gas chromatography is fundamental for organic analysis across environmental, food, clinical and petrochemical laboratories. Reliable GC performance depends on careful maintenance, high-purity carrier gas, proper sample preparation and optimized consumables. Anticipating contamination, retention shifts and peak-shape issues maximizes uptime, sensitivity and reproducibility for routine QA/QC and advanced research.
This presentation outlines strategies to plan ahead, prevent common GC and GC/MS problems and improve overall system robustness. Key goals include reducing contamination, extending column lifetime, enhancing peak quality and streamlining sample preparation workflows.
Recommended practices and instruments cover:
Clean carrier gas and proper traps eliminated background and irreversible column damage. High-quality syringes and septa reduced coring and bleed contaminants. EMR-Lipid improved reproducibility (RSD <3%) by removing lipids more effectively than simple protein precipitation. QuEChERS delivered rapid, multi-residue extraction with minimal solvent. Ultra Inert liners and columns produced symmetrical peaks for acids, bases and polar analytes while self-tightening nuts maintained leak-free connections over hundreds of runs.
Emerging opportunities include more advanced sorbent chemistries for selective cleanup, automated in-line monitoring of gas purity and flow, miniaturized consumables for microGC, and integration of smart diagnostics and predictive maintenance through machine learning.
Systematic planning, proper instrument configuration and proactive sample cleanup are key to maximizing GC performance and extending column and detector lifetime. Adopting these best practices empowers laboratories to achieve reliable, high-quality results with reduced operational costs.
GC
IndustriesManufacturerAgilent Technologies
Summary
Importance of Topic
Gas chromatography is fundamental for organic analysis across environmental, food, clinical and petrochemical laboratories. Reliable GC performance depends on careful maintenance, high-purity carrier gas, proper sample preparation and optimized consumables. Anticipating contamination, retention shifts and peak-shape issues maximizes uptime, sensitivity and reproducibility for routine QA/QC and advanced research.
Objectives and Overview
This presentation outlines strategies to plan ahead, prevent common GC and GC/MS problems and improve overall system robustness. Key goals include reducing contamination, extending column lifetime, enhancing peak quality and streamlining sample preparation workflows.
Methodology and Instrumentation
Recommended practices and instruments cover:
- Carrier gas: Ultra-high purity (≥99.9995%) with inline oxygen and moisture traps (e.g. Agilent GasClean).
- Sample handling: Fixed and removable needle syringes, high-quality vials and caps to prevent septum coring and siloxane bleed.
- Sample cleanup: Filtration, SPE cartridges, QuEChERS kits and EMR-Lipid cartridges for protein/lipid removal to reduce matrix effects.
- Inlet maintenance: Selection of deactivated liners (single or glass-wool taper), centerguide septa and gold or Ultimetal ferrules; use of self-tightening column nuts and regular leak checks.
- Column selection: Phase polarity, dimensions and bleed characteristics; use of Ultra Inert phases for challenging compounds and integrated DuraGuard guard columns.
- Detectors and accessories: FID, TCD, ECD and MSD; digital flow meters (ADM) with built-in leak-check and record features.
Main Results and Discussion
Clean carrier gas and proper traps eliminated background and irreversible column damage. High-quality syringes and septa reduced coring and bleed contaminants. EMR-Lipid improved reproducibility (RSD <3%) by removing lipids more effectively than simple protein precipitation. QuEChERS delivered rapid, multi-residue extraction with minimal solvent. Ultra Inert liners and columns produced symmetrical peaks for acids, bases and polar analytes while self-tightening nuts maintained leak-free connections over hundreds of runs.
Benefits and Practical Applications
- Enhanced sensitivity and lower detection limits from reduced ion suppression.
- Less downtime and maintenance cost by preventing matrix buildup.
- Improved reproducibility and chromatographic resolution.
- Faster sample throughput via streamlined cleanup protocols.
Future Trends and Applications
Emerging opportunities include more advanced sorbent chemistries for selective cleanup, automated in-line monitoring of gas purity and flow, miniaturized consumables for microGC, and integration of smart diagnostics and predictive maintenance through machine learning.
Conclusion
Systematic planning, proper instrument configuration and proactive sample cleanup are key to maximizing GC performance and extending column and detector lifetime. Adopting these best practices empowers laboratories to achieve reliable, high-quality results with reduced operational costs.
Used Instrumentation
- Agilent GC and GC/MS systems
- Ultra-high-purity gas filters (GasClean O2/H2O traps)
- Fixed and removable needle syringes, centerguide septa
- EMR-Lipid cartridges and QuEChERS kits
- Deactivated inlet liners (single taper, wool, Ultra Inert)
- Gold and Ultimetal ferrules, self-tightening column nuts
- J&W capillary columns (DB-5ms UI, DB-624 UI, WAX UI, DuraGuard)
- FID, TCD, ECD, MSD detectors
- ADM digital flow meters with leak-check and record functions
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
How to Keep and Good Thing Going: Preventative and Routine Maintenance for Your GC System
2024|Agilent Technologies|Presentations
How to Keep and Good Thing Going: Preventative and Routine Maintenance for Your GC System Alexander Ucci Online Application Engineer December 5, 2023 1 December 4, 2023 When Should I Change That? DE08858184 The Million Dollar Question – How Long…
Key words
change, changeshould, shouldwhen, whennut, nutcolumn, columnsepta, septaagilent, agilentferrule, ferruleyour, youruse, useseal, sealsmart, smartferrules, ferrulescleanup, cleanuptricks
A Holiday Gift Buying Guide for your GC: Latest Supply Innovations
2019|Agilent Technologies|Presentations
A Holiday Gift Buying Guide for your GC: Latest Supply Innovations Alexander Ucci Online Application Engineer November 12, 2019 1 November 12, 2019 Latest GC Supply Innovations Agilent Restricted Exclusive Offer for Attending Agilent Chemistries Webinars Receive a 25% discount…
Key words
innovations, innovationsrestricted, restrictedlatest, latestsupply, supplyagilent, agilentwax, waxsepta, septainert, inertcolumn, columnsmart, smartliner, linercleanup, cleanupliners, linersultra, ultracrimpers
When Should I Change That? Ensuring Proper Instrument Performance for GC
2021|Agilent Technologies|Presentations
When Should I Change That? Ensuring Proper Instrument Performance for GC Alexander Ucci Online Application Engineer December 14, 2021 1 December 13, 2021 When Should I Change That? DE08858184 The Million Dollar Question – How Long Will My Consumables Last?…
Key words
change, changeshould, shouldwhen, whencolumn, columnsepta, septaseal, sealsmart, smartyour, youruse, usecleanup, cleanupagilent, agilentpolyimide, polyimidegraphite, graphiteliner, linerdamage
A Beginner’s Guide to your GC Columns: Installation, Care and Maintenance
2020|Agilent Technologies|Presentations
A Beginner’s Guide to your GC Columns: Installation, Care, and Maintenance Alexander Ucci Application Engineer May 21, 2020 1 May 21, 2020 A Beginner’s Guide to your GC Columns: Installation, Care, and Maintenance DE.3875347222 Agenda • Unboxing / “getting to…
Key words
happy, happycolumn, columnrestricted, restrictedkeeping, keepingyour, youragilent, agilentleak, leakpolyimide, polyimideinstallation, installationsmart, smartdamage, damageferrule, ferruleintuvo, intuvoferrules, ferrulesdetector