Practical Steps in GC Troubleshooting Techniques, Tips, and Tricks

Presentations | 2017 | Agilent TechnologiesInstrumentation
GC
Industries
Manufacturer
Agilent Technologies

Summary

Significance of the topic


Gas chromatography (GC) remains a cornerstone technique in analytical laboratories for separating, identifying, and quantifying volatile compounds. Robust GC operation is critical across environmental monitoring, pharmaceuticals, food safety, and industrial quality control. Effective troubleshooting minimizes downtime, reduces resource waste, and ensures data integrity.

Objectives and overview of study/article


This application note by Mark Sinnott compiles practical guidance on diagnosing and resolving common GC issues. It outlines logical troubleshooting steps, catalogs typical chromatographic faults, presents diagnostic tests, and illustrates real-world examples.

Methodology and instrumentation


Addressed GC components include the injector, carrier gas flow system, column, detector, and electronics. Key diagnostic tools described are:
  • System change tracking and log book entries
  • Blank runs and injection of non-retained compounds
  • Baseline bleed profiling
  • Performance evaluation with test mixtures
  • Component isolation using jumper tubing and condensation tests

Main results and discussion


The note categorizes typical faults—peak tailing, ghost peaks, split peaks, missing peaks, retention shifts, loss of resolution, baseline disturbances, noisy or spiking baselines, and quantitation errors. For each category, it details probable causes (e.g., active sites in injector or column, leaks, contamination, overload, phase mismatch), offers diagnostic tips, and suggests corrective measures. Two case studies demonstrate systematic isolation of syringe blockage and column contamination, with step-by-step resolution.

Benefits and practical applications


Applying these structured troubleshooting strategies enables analysts to rapidly identify GC problems, restore optimal instrument performance, and maintain consistent analytical quality. Laboratories benefit from reduced downtime, more efficient maintenance routines, and greater confidence in data reproducibility.

Future trends and potential applications


Emerging predictive maintenance approaches, automated diagnostic software, and integration of real-time sensor data promise to further streamline GC troubleshooting. Artificial intelligence-driven analytics may soon enable proactive fault detection and rapid corrective action.

Conclusion


Effective GC troubleshooting relies on isolating system components, employing targeted diagnostic tests, and making incremental, logical adjustments. Adopting these practices ensures reliable chromatographic results and maximizes laboratory efficiency.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Troubleshooting GC Capillary Columns
Troubleshooting GC Capillary Columns
2011|Agilent Technologies|Presentations
Troubleshooting GC Capillary Columns Techniques, Tips, and Tricks Series 8 Simon Jones Application Engineer June 22, 2010 Page 1 What went wrong and how to fix it… Page 2  Common problems  Troubleshooting tools  Troubleshooting examples “Everything was…
Key words
injector, injectorcolumn, columnjumper, jumperdetector, detectortest, testtroubleshooting, troubleshootingcontamination, contaminationisolate, isolateelectronics, electronicstube, tubeblank, blankproblem, problemchange, changecontaminated, contaminatedpeaks
Practical Steps in GC Troubleshooting
Practical Steps in GC Troubleshooting
2017|Agilent Technologies|Presentations
Practical Steps in GC Troubleshooting Techniques, Tips, and Tricks Simon Jones Application Engineer GC Columns & Supplies Page 1 What went wrong and how to fix it… Common problems Troubleshooting tools Troubleshooting examples Page 2 “Everything was just fine and…
Key words
injector, injectorcolumn, columnjumper, jumperdetector, detectortest, testtroubleshooting, troubleshootingcontamination, contaminationisolate, isolatepeaks, peaksproblem, problemelectronics, electronicstube, tubeblank, blankchange, changecontaminated
A Tale of Two Peaks: Troubleshooting Poor Peak Shape
A Tale of Two Peaks: Troubleshooting Poor Peak Shape
2019|Agilent Technologies|Presentations
A Tale of Two Peaks: Troubleshooting Poor Peak Shape Alexander Ucci January 17, 2019 1 Agilent Restricted “Everything was Just Fine and then this Happened!” “How do I go about Troubleshooting?” Track events- log book • Changed column, liner, septum,…
Key words
restricted, restrictedagilent, agilentproblem, probleminjector, injectorwhat, whatelectronics, electronicsdetector, detectorcolumn, columnsyringe, syringedoes, doeslook, lookjumper, jumperoccurrence, occurrenceput, puttroubleshooting
General GC Column Troubleshooting
General GC Column Troubleshooting
2021|Agilent Technologies|Presentations
Get Your Peaks into Shape General GC Column Troubleshooting Alexander Ucci October 5, 2021 1 October 5, 2021 Get Your Peaks Into Shape DE44470.3469212963 “Everything Was Just Fine... and Then This Happened!” “How do I troubleshoot?” Track your actions/keep a…
Key words
get, getshape, shapepeaks, peaksyour, yourinjector, injectorcolumn, columnjumper, jumperdetector, detectorbaseline, baselinepeak, peakcontamination, contaminationtest, testtroubleshooting, troubleshootingtridecane, tridecanetailing
Other projects
LCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike